CLAS12 R1 S1 QA
HV burn in
Begin burn in after at least 3 volume changes.
File:High Voltage Region 1 Sector 1 data.xls
Step 1: Set Voltage to 50 volts on all channels. Probe HVB to be sure that all sense, field, and guard wires have a voltage. (You should set current threshold to 10 microAmps on the HV supply)
Step 2: While monitoring the current, ramp the HV to the following
G:F:S= 525,-700,700
Step 3: Begin to take data recording the time and sense wire current as you ramp up the sense wire HV to 1400 volts while restricting the current draw to less than 10 micraAmps.
keep current under 10 micro amps while moving sense voltage up to 1400
Measured Sense wire output
Sense Wire Scope pictures on DAQ machine
Cosmic Runs
3811
Cosmic rays were used to measure drift chamber time spectra with respect to scintillators placed above and below the drift chambers (more details are at the wiki location CLAS12R1_CosmicTestStand)
3833
3834
Cosmic measurements
CLAS6 Drift Chamber NIM paper
Run # | Date | Description |
3806 | 1/10/13 | 3 hour cosmic run, 16 channels with leading edge discriminators (-80 mV, 100 ns pulse width) |
3809 | 1/10/13 | cosmic run on 1/10. Signal cable 4 (ch. 48-63). 16 channels with leading edge discriminators (-80 mV, 100 ns pulse width). |
3810 | 1/10/13 | cosmic run from 1/10 - 1/14. Signal cable 4 (ch. 48-63). 16 channels with leading edge discriminators (-80 mV, 100 ns pulse width). Scintillator paddle F1/F2 on table above Sector 1, paddle I1/I2 below. |
3811 | 1/10/13 | |
3831 | 2/1/13 | Changed disciminator from 80 mV to 60 mV, We improved the window, installed HV & STB boards, purged and started taking cosmic data again, DC is sandwhiched between 2 scints with PMTs aligned (PMT on only one end of Scint is turned on) |
3833 | 2/4/13 | Changed disciminator from 60 mV to 30 mV |
3834 | 2/11/13 | Changed disciminator from 30 mV to 60 mV |
3835 | 2/19/13 | Changed disciminator from 60 mV to 20 mV |
Using ADB crate
TDC-CH# | WIRE | |
104 | PMT I2 | |
106 | PMT F2 | |
107 | PMT F1 | |
00 | U1 | L2W1 |
01 | U2 | L4W1 |
02 | U3 | L6W1 |
03 | U4 | L1W1 |
04 | U5 | L3W1 |
05 | U6 | L5W1 |
06 | U7 | L2W2 |
07 | U8 | L4W2 |
08 | U9 | L6W2 |
09 | U10 | L1W2 |
10 | U11 | L3W2 |
11 | U12 | L5W3 |
12 | U13 | L2W3 |
13 | U14 | L4W3 |
14 | U15 | L6W3 |
15 | U16 | L1W3 |
16 | U17 | L3W3 |
17 | U18 | L5W3 |
18? | U19 | L2W4 |
19 | U20 | L4W4 |
20 | U21 | L6W4 |
21 | U22 | L1W4 |
22 | U23 | L3W4 |
23 | U24 | L5W4 |
24 | U25 | L2W5 |
25 | U26 | L4W5 |
26 | U27 | L6W5 |
27 | U28 | L1W5 |
28 | U29 | L3W5 |
29 | U30 | L5W5 |
30 | U31 | L2W6 |
31 | U32 | L4W6 |
32 | U33 | L6W6 |
33 | U34 | L1W6 |
34 | U35 | L3W6 |
35 | U36 | L5W6 |
36 | U37 | L2W7 |
37 | U38 | L4W7 |
38 | U39 | L6W7 |
39 | U40 | L1W7 |
40 | U41 | L3W7 |
41 | U42 | L5W7 |
42 | U43 | L2W8 |
43 | U44 | L4W8 |
44 | U45 | L6W8 |
45 | U46 | L1W8 |
46 | U47 | L3W8 |
47 | U48 | L5W8 |
48 | U49 | L2W9 |
49 | U50 | L4W9 |
50 | U51 | L6W9 |
51 | U52 | L1W9 |
52 | U53 | L3W9 |
53 | U54 | L5W9 |
54 | U55 | L2W10 |
55 | U56 | L4W10 |
56 | U57 | L6W10 |
57 | U58 | L1W10 |
58 | U59 | L3W10 |
59 | U60 | L5W10 |
60 | U61 | L2W11 |
61 | U62 | L4W11 |
62 | U63 | L6W11 |
63 | U64 | L1W11 |
64 | U65 | L3W11 |
65 | U66 | L5W11 |
66 | U67 | L2W12 |
67 | U68 | L4W12 |
68 | U69 | L6W12 |
69 | U70 | L1W12 |
70 | U71 | L3W12 |
71 | U72 | L5W12 |
72 | U73 | L2W13 |
73 | U74 | L4W13 |
74 | U75 | L6W13 |
75 | U76 | L1W13 |
76 | U77 | L3W13 |
77 | U78 | L5W13 |
78 | U79 | L2W14 |
79 | U80 | L4W14 |
80 | U81 | L6W14 |
81 | U82 | L1W14 |
82 | U83 | L3W14 |
83 | U84 | L5W14 |
84 | U85 | L2W15 |
85 | U86 | L4W15 |
86 | U87 | L6W15 |
87 | U88 | L1W15 |
88 | U89 | L3W15 |
89 | U90 | L5W15 |
90 | U91 | L2W16 |
91 | U92 | L4W16 |
92 | U93 | L6W16 |
93 | U94 | L1W16 |
94 | U95 | L3W16 |
95 | U96 | L5W16 |
Runlist Disc=100 mv
Run # | Date | Description |
4008 | 4/11/13 | set ADB threshold to 100 , installed all 6 STB connectors, key was on the left as you face the ADC crate. Two top PMTs and one bottom PMT were in coincidence. The rate was about 0.5 Hz. at 9:41 on 4/12/13 I observed no voltage coming out of the discriminator monitor, the fuses were NOT blown. Hits seemed to disappear in the data. I recycled the ADB crate power and saw 100 mV on the discriminator monitor. Start a new run 40009 |
4009 | 4/12/13 | ADC threshold at 100 mV |
4014 | 4/16/13 | Removed cable from J10 connector on ADB crate and becan a cosmic run, the ROC & EB1 segfaulted after 57702 events |
4082 | 4/17/13 | first run with fully installed ADC crate & known map; ROC crashed after 5522 event around 20:53 |
4099 | 4/17/13 | ROC & EB crash after 4702 events, restarted all processes prior to taking this data |
4127 | 4/24/13 | 100 mv Threshold, S:F:G=1500:-750:550 ROC crashed because there was not enough memory for all the data, increased memory in the readoutlist R4127Histos |
4128 | 4/24/13 | run ended on 4/29 at 10:43,100 mv Threshold, S:F:G=1500:-750:550 re-centered scintillator based on occupancy plot from 4127, the event rate dropped after 2 hrs (event 5000) , TDC chan 32 -> 63 dropped out between event number 5000 and 360000, why did they start working? |
4129 | 4/29/13 | 100 mv Threshold, S:F:G=1500:-750:550 , lost hits after 50000 events |
4130 | 4/30/13 | 100 mv Threshold, S:F:G=1500:-750:550 , only 3, 16 channel ADB inputs are working (TDC channels 16->31, 64->95) |
4131 | 5/1/13 | 100 mv Threshold, S:F:G=1500:-750:550 , Moved DC output connectors so the channels centered on the scintillator are on the working ADB inputs (ribbon cable 3 is on ADB input 2, cable 4 is on input 5, and cable 5 is on ADB input 6); Only ADB input 6 shows hits |
4132 | 5/2/13 | 100 mv Threshold, S:F:G=1500:-750:550 , Moved DC output connectors so ribbon cable 2 is on ADB input 1, cable 5 is on input 4, and cable 6 is on ADB input 6, all others are unplugged |
4134 | 5/3/13 | 100 mv Threshold, S:F:G=1500:-750:550 , Moved DC output connectors so ribbon cable 2 is on ADB input 3, cable 5 is on input 5, and cable 6 is on ADB input 6, all others are unplugged |
4135 | 5/4/13 | 100 mv Threshold, S:F:G=1500:-750:550 , replaced preamp power supply, CLAS12_r4135 |
4136 | 5/6/13 | 100 mv Threshold, S:F:G=1500:-750:550 , CLAS12_r4136 |
4138 | 5/8/13 | 100 mv Threshold, S:F:G=1500:-750:550 , CLAS12_r4138,internet failed and lost connection to rocdaq1 after 479752 events on 5/13/13 at 6:11 am |
Runlist Disc=30 mv
4139 | 5/13/13 | 30 mv Threshold, S:F:G=1500:-750:550 CLAS12_r4139, run started at 9:05:57 on 5/13 and ended at 12:52:44 on 5/20 |
Runlist Disc=58 mv
4140 | 5/20/13 | 58 mv Threshold, S:F:G=1280:-750:550 CLAS12_r4140, run started at 13:57:34 on 5/20 and ended at 16:43:39 on 5/20 |
4141 | 5/20/13 | 58 mv Threshold, S:F:G=1500:-750:550 CLAS12_r4141, run started at 16:47:48 on 5/20 and ended , TDC channels 62->76 were failing after event 80000 and then seemed to turn on and off for a while |
4142 | 5/20/13 | 58 mv Threshold, S:F:G=1500:-750:550 , run started at , TDC channels 62->76 were failing they seemed to turn on and off for a while |
4143 | 5/27/13 | 58 mv Threshold, S:F:G=1500:-750:550 , swapped last two inputs to ADB board for channels 64->96. Check if hits start failing on ch 80->96 |
4145 | 5/28/13 | 58 mv Threshold, S:F:G=1500:-750:550 , swapped middle inputs between two ADC boards (ch 17->32 echanged with ch 64->80), all channels were working??? lets switch it back, perhaps connector had bad connection? |
4146 | 5/29/13 | 58 mv Threshold, S:F:G=1500:-750:550 , everything connected back to order used in run 4142, run started at 08:12:35 on 5/29/13 |
Readout Map
Wire | PostAmpOut (CH 1-96) | TDC Channel | |
L | W | ||
5 | 1 | 1 | 1 |
3 | 1 | 2 | 2 |
1 | 1 | 3 | 3 |
6 | 1 | 4 | 4 |
4 | 1 | 5 | 5 |
2 | 1 | 6 | 6 |
5 | 2 | 7 | 7 |
3 | 2 | 8 | 8 |
1 | 2 | 9 | 9 |
6 | 2 | 10 | 10 |
4 | 2 | 11 | 11 |
2 | 2 | 12 | 12 |
5 | 3 | 13 | 13 |
3 | 3 | 14 | 14 |
1 | 3 | 15 | 15 |
6 | 3 | 16 | 16 |
4 | 3 | 17 | 17 |
2 | 3 | 18 | 18 |
5 | 4 | 19 | 19 |
3 | 4 | 20 | 20 |
1 | 4 | 21 | 21 |
6 | 4 | 22 | 22 |
4 | 4 | 23 | 23 |
2 | 4 | 24 | 24 |
5 | 5 | 25 | 25 |
3 | 5 | 26 | 26 |
1 | 5 | 27 | 27 |
6 | 5 | 28 | 28 |
4 | 5 | 29 | 29 |
2 | 5 | 30 | 30 |
5 | 6 | 31 | 31 |
3 | 6 | 32 | 32 |
1 | 6 | 33 | 33 |
6 | 6 | 34 | 34 |
4 | 6 | 35 | 35 |
2 | 6 | 36 | 36 |
5 | 7 | 37 | 37 |
3 | 7 | 38 | 38 |
1 | 7 | 39 | 39 |
6 | 7 | 40 | 40 |
4 | 7 | 41 | 41 |
2 | 7 | 42 | 42 |
5 | 8 | 43 | 43 |
3 | 8 | 44 | 44 |
1 | 8 | 45 | 45 |
6 | 8 | 46 | 46 |
4 | 8 | 47 | 47 |
2 | 8 | 48 | 48 |
5 | 9 | 49 | 49 |
3 | 9 | 50 | 50 |
1 | 9 | 51 | 51 |
6 | 9 | 52 | 52 |
4 | 9 | 53 | 53 |
2 | 9 | 54 | 54 |
5 | 10 | 55 | 55 |
3 | 10 | 56 | 56 |
1 | 10 | 57 | 57 |
6 | 10 | 58 | 58 |
4 | 10 | 59 | 59 |
2 | 10 | 60 | 60 |
5 | 11 | 61 | 61 |
3 | 11 | 62 | 62 |
1 | 11 | 63 | 63 |
6 | 11 | 64 | 64 |
4 | 11 | 65 | 65 |
2 | 11 | 66 | 66 |
5 | 12 | 67 | 67 |
3 | 12 | 68 | 68 |
1 | 12 | 69 | 69 |
6 | 12 | 70 | 70 |
4 | 12 | 71 | 71 |
2 | 12 | 72 | 72 |
5 | 13 | 73 | 73 |
3 | 13 | 74 | 74 |
1 | 13 | 75 | 75 |
6 | 13 | 76 | 76 |
4 | 13 | 77 | 77 |
2 | 13 | 78 | 78 |
5 | 14 | 79 | 79 |
3 | 14 | 80 | 80 |
1 | 14 | 81 | 81 |
6 | 14 | 82 | 82 |
4 | 14 | 83 | 83 |
2 | 14 | 84 | 84 |
5 | 15 | 85 | 85 |
3 | 15 | 86 | 86 |
1 | 15 | 87 | 87 |
6 | 15 | 88 | 88 |
4 | 15 | 89 | 89 |
2 | 15 | 90 | 90 |
5 | 16 | 91 | 91 |
3 | 16 | 92 | 92 |
1 | 16 | 93 | 93 |
6 | 16 | 94 | 94 |
4 | 16 | 95 | 95 |
2 | 16 | 96 | 96 |
Measurements with source
4/20/12 1hour counts:
SL2 HVB1 Column 18 Row 5
-With source 14459 pulses/1502 sec
4/23/12
SL1 HVB7 column 203 row 4
-With source 50681 pulses/3760 sec.
4/25/12
1 hour counts:
SL2 HVB4 Column 118 row 5
- Without source: 33145 pulses/3610 Sec.
- With source(CS-137):38262 pulses/3600 Sec.
- Without source: 37548 pulses/3715 Sec.
Preamp voltage:5V
Background measurements
HV off 4/5/12 Background counts:
27 pulses/61806 Sec.
SL2 Column 132 row 2
Preamp voltage 6.3V
Discriminator threshold 46mV
Preamp on= 6.3
4/9/12: Back ground count measurements
4531 counts/69564 seconds
2044 counts/24620 seconds
4/10/12 Background counts:
3 pulses/ 74500 sec.
SL1 HVB7 Column 196 Row 2
Preamp voltage 6.3 V
Discriminator threshold 46mV
4/16/12
Background counts:
280 pulses/245850 sec
SL1 HVB7 column 226 row 2
preamp voltage 5V
Discriminator threshold 46mV
4/17/12 Background counts:
6 pulses/62407 sec
SL2 HVB7 column 226 row 8
preamp voltage 5V
discriminator threshold 46mV
4/20/12
Background counts:
63 pulses/ 71724 Sec.
SL1 HVB7 column 226 row 8
Preamp voltage 5V
Discriminator threshold 46mV
4/23/12
Background counts:
41078 pulse/ 245104 sec.
SL1 HVB7 Column 203 Row 4
Preamp voltage 5V
Discriminator threshold 46mV
Cosmic measurements
4/24/12 Cosmic count:
SL1 HVB1 Column 3 Row 10
14534 pulses/11547 sec.
Preamp voltage 5V
4/26/12
Overnight Cosmic pulse count:
SL2 HVB4 Column 118 Row 5
53122 pulses/49793 Sec.
Preamp voltage 5V