Difference between revisions of "Cf252 runs with HPGe"
Jump to navigation
Jump to search
Line 18: | Line 18: | ||
|- | |- | ||
|} | |} | ||
+ | |||
+ | |||
+ | '''Total n flux: 13885 n/cm^2''' |
Revision as of 22:13, 30 December 2016
I made a page discussing neutron damage to the HPGe detector as a function of the total amount of fast neutrons per cm^2 irradiated on the detector face. 5*10^6 n/cm^2 is a factor of 10 below the point where damage was first observed in an experiment.
12/29/16
Run # | RunTime [sec] | distance[cm] | n flux (Cf252 rates) |
Integral n flux |
configuration | notes |
---|---|---|---|---|---|---|
5609 | 4060 | 30 | 6850 | 6850 | Cf252 in. Coincidence between HPGe and timing PMT | A ToF spectrum exists but is very hectic. Problem may be that the TDC stop is determined by a widened timing PMT pulse in coinc. with a HPGe pulse that has high timing jitter. |
none | 1250 | 26 | 2800 | 9655 | Not taking data during this time. | |
5611 | 2509 | 30 | 4230 | 13885 | Cf252 in. Coincidence between HPGe and timing PMT. TDC stop was changed to always be 1200 ns after PMT timing pulse, otherwise same setup as 5609. | ToF looks good. Neutron rate approx. 3 Hz for detector 90. |
5613 | 10025 | 15 | 0 | 13885 | Na-22 source. Coincidence between HPGe and timing PMT | Calibration |
5614 | 654 | 15 | 0 | 13885 | Co-60 source. Coincidence between HPGe and timing PMT | Calibration |
Total n flux: 13885 n/cm^2