Difference between revisions of "Cf252 runs with HPGe"
		
		
		
		
		
		Jump to navigation
		Jump to search
		
				
		
		
	
| Line 11: | Line 11: | ||
|-  | |-  | ||
| 5611 || 2509 ||  30  ||  4230 ||  13885 || Cf252 in. Coincidence between HPGe and timing PMT. TDC stop always 1200 ns after PMT timing pulse, otherwise same setup as 5609.  || ToF looks good. Neutron rate approx. 3 Hz for detector 90.     | | 5611 || 2509 ||  30  ||  4230 ||  13885 || Cf252 in. Coincidence between HPGe and timing PMT. TDC stop always 1200 ns after PMT timing pulse, otherwise same setup as 5609.  || ToF looks good. Neutron rate approx. 3 Hz for detector 90.     | ||
| + | |-  | ||
| + | | 5613 || 10025  || 15  ||  0    ||   13885  || Na-22 source.  Coincidence between HPGe and timing PMT ||   | ||
|-  | |-  | ||
|}  | |}  | ||
Revision as of 06:58, 30 December 2016
12/29/16
| Run # | RunTime [sec] | distance[cm] | n flux  (Cf252 rates)  | 
Integral  n flux  | 
configuration | notes | 
|---|---|---|---|---|---|---|
| 5609 | 4060 | 30 | 6850 | 6850 | Cf252 in. Coincidence between HPGe and timing PMT | A ToF spectrum exists but is very hectic. Problem may be that the TDC stop is determined by a widened timing PMT pulse in coinc with a HPGe pulse that has high timing jitter. | 
| none | 1250 | 26 | 2800 | 9655 | Not taking data during this time. | |
| 5611 | 2509 | 30 | 4230 | 13885 | Cf252 in. Coincidence between HPGe and timing PMT. TDC stop always 1200 ns after PMT timing pulse, otherwise same setup as 5609. | ToF looks good. Neutron rate approx. 3 Hz for detector 90. | 
| 5613 | 10025 | 15 | 0 | 13885 | Na-22 source. Coincidence between HPGe and timing PMT | |