R3833 Efficiency1: Difference between revisions

From New IAC Wiki
Jump to navigation Jump to search
Foretony (talk | contribs)
Created page with " My first attempt at an efficiency measurement compares wire hits when you require that the first (wire 1) and last (wire 6) wires in a single super layer are hit. First I looke…"
 
Foretony (talk | contribs)
No edit summary
Line 1: Line 1:
My first attempt at an efficiency measurement compares wire hits when you require that the first (wire 1) and last (wire 6) wires in a single super layer are hit.
My first attempt at an efficiency measurement compares wire hits when you require that the first (wire 1) and last (wire 6) wires in a single super layer are hit.


First I looked at which wires were hit when I require that there is a hit on wire 1 and wire 2.
First I looked at which wires were hit when I require that there is a hit on wire 1 and wire 2.
HV was G:F:S = 536:-750:1500 Volts





Revision as of 00:34, 7 March 2013

My first attempt at an efficiency measurement compares wire hits when you require that the first (wire 1) and last (wire 6) wires in a single super layer are hit.

First I looked at which wires were hit when I require that there is a hit on wire 1 and wire 2.

HV was G:F:S = 536:-750:1500 Volts


Most of the hits are in row 9 (I was only looking at 16 sense wires)


If I histogram all wire hits requiring a hit in both wire 1, wire 6 and row 9 I see the following

Now I divide by the number of hits seen by both wire 1 and wire 6

CLAS12_R1_S1_QA