Difference between revisions of "PAA ResultsFromTheIAC"

From New IAC Wiki
Jump to navigation Jump to search
 
(12 intermediate revisions by the same user not shown)
Line 1: Line 1:
 +
==Nickel Normaization==
 +
 +
[[Ni-08-22-13]]
 +
 +
== Flux Rates on Jack==
 +
 +
===Jack Photon Flux===
 +
 +
[[JackPhotonFlux_06-04-2018]]
 +
 +
===Jack Neutron Flux===
 +
 +
[[NAA_SolarPanel]]
 +
 
==Scotch Tape -vs- Kapton tape==
 
==Scotch Tape -vs- Kapton tape==
  
Line 37: Line 51:
 
   
 
   
 
[[File:IrradiatedKapton-vsScotchTapeRates_9-22-16.png | 400 px]]
 
[[File:IrradiatedKapton-vsScotchTapeRates_9-22-16.png | 400 px]]
 +
 +
==HEU disks==
 +
 +
A 20 MeV electron was used to make a bremsstrahlung beam that impinged HEU disks
 +
 +
 +
Below is an uncalibrated spectrum using Thorium rods on Detector D.  Lines should appear at 238.8,338.8, 511, 583.4, 911.2, and 2614.6 keV
 +
 +
The HEU samples were measured on two different HpGe detectors that has similar efficiencies.  Get B was used for the Unirradiate sample and Detector D for the Irradiated HEU disk.
 +
 +
 +
=== Det D calibrations===
 +
{| border="3"  cellpadding="5" cellspacing="0"
 +
|  Channel|| Energy (keV)
 +
|-
 +
|236 || 238.8
 +
|-
 +
|334 || 338.8
 +
|-
 +
|505 || 511
 +
|-
 +
|576 || 583
 +
|-
 +
|901 || 911.2
 +
|-
 +
|2586 || 2614.6
 +
|}
 +
 +
slope=1.0108 +/- 0.000328475
 +
y-intercept=0.6533 +/- 0.385247
 +
 +
MPA->Draw("0.6533+1.0108*evt.Chan>>(8000,0,8000)");
 +
 +
 +
===HEU comparison Before and After irradiation===
 +
 +
Uniradiated HEU sample was measured for 2671.8 seconds
 +
 +
MPA->Draw("-3.12645+1.0003*evt.Chan>>NHEUback(2000,0,8000)","0.0003742795");
 +
 +
Iradiated HEU sample was measured overnight for 50554.469 seconds (14 hours)
 +
 +
MPA->Draw("0.6533+1.0108*evt.Chan>>NIradHEU(2000,0,8000)","0.00001978064");
  
 
[[PAA_Research]]
 
[[PAA_Research]]

Latest revision as of 23:09, 14 June 2018

Nickel Normaization

Ni-08-22-13

Flux Rates on Jack

Jack Photon Flux

JackPhotonFlux_06-04-2018

Jack Neutron Flux

NAA_SolarPanel

Scotch Tape -vs- Kapton tape

A 20 MeV electron was used to make a bremsstrahlung beam that impinged a sample of regular Scotch Tape and Kapton tape. Both tapes had similar areas of 2.5 x 3 cm. The Kapton had twice as much mass as the Scotch (0.1247 g -vs 0.0699 g).

The scotch taper became brittle after 4 hours of irradiation on the 25 N machine.

Below is an uncalibrated spectrum using Thorium rods. Lines should appear at 238.8,338.8, 511, 583.4, 911.2, and 2614.6 keV


Channel Energy (keV)
242 238.8
342 338.8
514 511
586 583
914 911.2
2617 2614.6

slope=1.0003 +/- 0.000156023 y-intercept=-3.12645 +/- 0.185346

MPA->Draw("-3.12645+1.0003*evt.Chan>>(8000,0,8000)");

Scotch tape run was 6663.4 seconds long (MPA->Draw("-3.12645+1.0003*evt.Chan>>NScotchTape(8000,0,8000)","0.00015");)

Kapton Tape run was 488.44 seconds long (MPA->Draw("-3.12645+1.0003*evt.Chan>>NKaptonTape(8000,0,8000)","0.0020473344");)


IrradiatedKapton-vsScotchTapeRates 9-22-16.png

HEU disks

A 20 MeV electron was used to make a bremsstrahlung beam that impinged HEU disks


Below is an uncalibrated spectrum using Thorium rods on Detector D. Lines should appear at 238.8,338.8, 511, 583.4, 911.2, and 2614.6 keV

The HEU samples were measured on two different HpGe detectors that has similar efficiencies. Get B was used for the Unirradiate sample and Detector D for the Irradiated HEU disk.


Det D calibrations

Channel Energy (keV)
236 238.8
334 338.8
505 511
576 583
901 911.2
2586 2614.6

slope=1.0108 +/- 0.000328475 y-intercept=0.6533 +/- 0.385247

MPA->Draw("0.6533+1.0108*evt.Chan>>(8000,0,8000)");


HEU comparison Before and After irradiation

Uniradiated HEU sample was measured for 2671.8 seconds

MPA->Draw("-3.12645+1.0003*evt.Chan>>NHEUback(2000,0,8000)","0.0003742795");

Iradiated HEU sample was measured overnight for 50554.469 seconds (14 hours)

MPA->Draw("0.6533+1.0108*evt.Chan>>NIradHEU(2000,0,8000)","0.00001978064");

PAA_Research