Difference between revisions of "HM 2014"
Line 1: | Line 1: | ||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
=Primary and secondary number of electrons: references and simulations 02/26/15= | =Primary and secondary number of electrons: references and simulations 02/26/15= | ||
Revision as of 18:34, 5 March 2015
Primary and secondary number of electrons: references and simulations 02/26/15
practice writing it as though you are writing a thesis.
outline the points you want to make, then construct a first paragraph with those points
- Alpha energy and previously measured ionization.
- G4 and Srim/Trim Performed Simulations
try this
The ionization produced within the GEM detector's Ar/Co2 drift region due to the passage of alpha particles emitted by an embedded U-233 source is calculated below. First, the alpha emission rates were quantified according to their energy. Then a simulation was benchmarked with published data to determine the amount of primary ionization produced from a single alpha particle of a given energy. The ionization that takes place when the above primary electrons are accelerated in an external elective field was also simulated. The amplification of this ionization signal by the GEM foil preamplifier gain and the signal processing electronics was included to predict the ADC's response to an ionization event cause by the alpha particles from the embedded U-233 source. Finally, the impact of the shutter on the signal was predicted using a simulation. Now insert a paragraph for each of the sentences (after the first sentence) above
- Alpha emission rates and their energies
A U-233 is a source of alpha particles was installed inside GEM detector. U-233 coated on the cathode of the GEM detector which separated with a drift area of 1cm. U-233 is an alpha emitter, the figure below shows the energy spectrum and the rate for each alpha energy,
insert table of alpha rates for each energy.
Alpha particles of an energy of 4.82 MeV has the highest rate of 85 percent compared to the other alpha energy's rate.
- Simulation benchmarks with published data to determine the amount of primary ionization produced from a single alpha particle of a given energy.
When an alpha particle travels in a pure argon, it liberates up to 30,000 electrons through primary and secondary ionization. <ref> Fabio, S. (2014). Basic processes in gaseous counters. In Gaseous Radiation Detectors: Fundamentals and Applications. Cambridge: University Printing House </ref> with without any electric field effect that may decrease the probability of electron-ion reattachment. On the other hand, Saito <ref> Saito, K., & Sasaki, S. (2003). Simultaneous Measurements of Absolute Numbers of Electrons and Scintillation Photons Produced by 5.49 MeV Alpha Particles in Rare Gases. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 50(6), 2452-2460 </ref> measured the number of primary and secondary electrons for a 5.49 MeV alpha particle, when about 4.7 kV/cm drift electric field collected the free electrons in the drift area to a collector, the number of collected electrons reached to 200k electron. Saito's measurements shows that the collector almost counts for all electrons, and a 1 kV/cm electric field decreases the probability of any electron-ion reattachment to increase the number of the collected electrons 8 times compared to that without the electric field.
Simulations of GEANT4 <ref> Agostinelli, S. (2003). Geant4—a simulation toolkit. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 506(3), 250–303 </ref> and Srim/Trim <ref>Ziegler, J. (2010). SRIM - The stopping and range of ions in matter (2010). Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 268(11-12), 1818-1823 </ref> were used to estimate the number of primary and secondary electrons for an alpha particle, both of the tools are able to calculate the number of primary and secondary electrons in a specific gaseous medium with a specific physical conditions for pressure, temperature, and density. The model in each tool was tested by estimating the range in Ar and CO2 gases, the results are shown below,
According to the figures above G4 succeeded to simulate the alpha range accurately in Ar and CO2 gas, but Srim/Trim model estimated the range with 50 percent less than that measured by Hanke and Bichsel <ref> Hanke, C., & Bichsel, H. (1970). Precision energy loss measurements for natural alpha particles in argon. Kbh.: Det Kongelige Danske Videnskabernes Selskab </ref>, also Srim/Trim estimated the same range of 4 cm for different alpha's energies.
G4 simulated the number of primary and secondary electrons in 1 cm of Ar/CO2 90/10 gas mixture, and the figures below show the results:
File:Excitation ionization ratio saito 2003.pdf
G4 estimated 2.7k secondary electrons when a 5.49 MeV alpha penetrated a 1cm of the same medium without any applied electric field, the estimation is close to Saito's measurements, which could be interpreted that G4 did not consider the reattachment of the electrons as they were collected in the drift area to end up counting for all the free electron in the gas. On the other hand, Scrim and Trim simulation for alpha particles ionizing a 1cm of pure Ar gas estimated about 1 MeV deposited energy in the medium within that distance, it made the final number of electrons around 37.5k electron (considering w = 26.7 eV/ip), which the same estimation mentioned by Sauli.
- The amplification of this ionization signal by the GEM foil preamplifier gain and the signal processing electronics was included to predict the ADC's response to an ionization event cause by the alpha particles from the embedded U-233 source
Garfield is used to simulate the electron multiplication in Ar/CO2 gas in GEM based detectors. Garfield simulated a triple GEM detector gain of 200 eV electron that passed through a 1cm drift area of an electric field of 1-4kV/cm. The readout collected 8 +_1 electrons for each incident electron by he end of the drift area, such a result explained increasing the number of alpha's electrons by ionization to 200k electron <\Saito>, as the free electrons travelled through the drift area, a multiplication of order of 8+_1 electrons was generated by accelerating the incident electrons by the 4.7 kV/cm drift electric field, by dividing 200 k electron by 8, the result was 25k electron is the number for each alpha particle before multiplication, and it is within the same order as one predicted by Scrim/Trim and Sauli.
Garfield was also used for simulating the electron multiplication by triple GEM detector. By the end of the drift area, the electron will have 3 stages of multiplication by GEM preamplifiers, the voltage on the potential divider ciccuit is 2800 kV, which will provide 300-350 V on each GEM preamplifier. Garfield succeeded to simulate the gain for 2.8 kV triple GEM in Ar/CO2 90/10 gas mixture under STP conditions of an average value of 7.88 +_ 2.8 x 10^3. So, the charge collected by the charge collector for 4.82 MeV alpha particles is expected to be:
The figure below shows the signal processing electronics configuration,
A signal of an amplitude 11.8 mV passed through the loop to output a 20.8 mV amplitude signal as shown by a 50 ohm terminated oscilloscope, using the figure below, the amount charge that ADC collected is,
- 100 keV Electrons and 1keV photons
G4 is used to evaluate the number of primary and secondary electrons produced in a 1 cm of pure Ar gas, the result is summarized in the following table:
Particle | Primary | Secondary |
1 keV X-ray | 0.69 | 72 |
100keV electron | 906 ip/cm | 3455 ip/cm |
which make the simulation close up to 10% for the primary and secondary electrons.
Particle | Primary | Secondary |
1 keV X-ray | 1 | 50 |
100keV electron | 1000 ip/cm | 3000 ip/cm |
5 MeV alpha particle | 10^4 | 3X10^4 |
Fabio Sauli, Gaseous Radiation Detectors: Fundamentals and Applications. [1] page 21.
02/17/15 QDC & PS-ADC measurements
each run lasted for 5 min. unless is mentioned differently
cath (kV) | V_drift | V_GEM (kV) | open | closed | notes |
3.87 | 1000 | 2.87 | 8569 | 6570 | |
3.57 | 700 | 2.87 | 8572 | 8571 | QDC does not show a observed difference between open and closed shutter |
3.57 | 700 | 2.87 | 8573 | 8574 | |
3.87 | 1000 | 2.87 | 8576 | 6575 | |
3.87 | 1000 | 2.87 | 8578 | 6577 | |
3.87 | 1000 | 2.87 | 8580 | 6579 | |
3.87 | 1000 | 2.87 | 8582 | 6581 |
- QDC Calibration
A triangle test pulse was ejected into the QDC for calibration, details of the pulse amplitude (mV) and width (us) in the table below
amplitude (mV) +_ 0.10 | width (us) +_0.10 | charge (nC) | channel number | run number |
2.04 | 1.56 | 0.03 +_ 0.13 | 1513.2 +_ 1.8 | 8591 |
7.00 | 2.32 | 0.16 +_ 0.37 | 2161.7 +_ 1.5 | 8589 |
10.00 | 2.46 | 0.25 +_ 0.51 | 2767.5 +_ 1.5 | 8592 |
13.00 | 2.64 | 0.34 +_ 0.66 | 3169.7 +_ 1.2 | 8593 |
17.50 | 2.65 | 0.46 +_ 0.89 | 3618.2 +_ 0.9 | 8585 |
- QDC Calibration
Asqaure test pulse was ejected into the QDC for calibration, details of the pulse amplitude (mV) and width (us) in the table below
amplitude (mV) +_ 0.10 | charge (nC) | channel number | run number |
7.00 | 1.14 | 407.3+_ 0.5 | 8617 |
8.00 | 1.30 | 1077.0 +_ 0.002 | 8618 |
9.00 | 1.47 | 2210.3 +_ 0.4 | 8619 |
10.00 | 1.63 | 3422.0 +_ 0.4 | 8620 |
pulse width was 8.16 us.
The after signal processing of an input signal of 11.2 mV,the QDC's input signal is 20.8 mV according to what observed on the oscilloscope.
Roy's measurements is as the following
Shutter position | Alpha particles /min. | Beta particles /min. |
Open | 6879 | 900 |
Close | 1 | 38 |
Depending on G4 simulation, each alpha particle has 270k secondary electrons which appear as a result of ionization. their total charge is
why is the charge multiplied by 2?
you need to document the amplification of this charge to describe how it is measured by ADC.
Compared to the measured charge spectrum, the figure below average spectrum of five different measurements after subtracting the pedestal.
1/15/15
- Rate calculation (Shutter Open)
- SNR -vs- V
01/11/15 2.97kV GEM
cathode drift potential and count rate runs for 2.97kV GEM
cathode drift 1-1.1k potential and count rate runs for 2.87kV GEM
01/08/14 Peak shift with changing the GEM preamp. Voltage
The ratio of signal to noise for each voltage is show below:
Even if the error bars looks small in the first figure that shows the sig./noi. and voltage of the GEM, it is noticed the error bars insect for the measurements of open shutter and the open-closed. So the error bars of the first figure can't tell if the increment of 100V for the GEM preamp's is really causing a change in the collected charge spectra.
12/31/14 Peak shift with changing the GEM preamp. Voltage
The following figures show the effect of opening and closing the shutter as the GEM voltage is 2.67, 2.87 and 2.97 kV.
The ratio of signal to noise for each voltage is show below:
You need to add error bars and make measurements at additional voltages ( measure for every 5 volts if error bars can tell the difference between each voltage)
It is noticed that increasing the voltage decreased the signal to noise ratio unless it is within the error bars, also increasing the voltage of the GEM from 2.87 to 2.97 kV does not affect the noise signal ratio, but it increases the charge detected which is expected as main characteristic for the GEM preamp. (results are same using both definitions of signal to noise ratio).
V_GEM (kV) | open | closed | notes |
2.87 | 8479,8481, 8408 | 8480,8482,8409 | |
2.97 | 8484, 8486,8415 | 8483,8485,8416 |
12/30/14 Determining the noise level using pulse shape and peak sensing discrimination simultaneously
Using an AND-gate for a peak sensing discriminator and a leading edge one is used to determine the relationship between the discrimination level and the the signal rate.
Psensing spectra with an AND-gate output as a gate for the module
- Using LED only
The LED disc. is used instead of the PS disc. to measure the Psensing charge spec., the following figure shows the result.
the signal is amplified so it is higher than the noise level, but still the LED disc. is unable to distinguish between open shutter and closed one.
the noise level is measured and it is 78 mV. (run 8445)
The counts near channel 100 suggest that the leading edge discriminator is not rejecting noise that we called a pedestal event.
12/27/14 Peak shift with changing the GEM preamp. Voltage
GEM3k peak sensing charge spectra
GEM3k peak sensing charge spectra
- GEM 2.87kV and cath. 3.47kV
- GEM 2.67 and cath. 3.27kV
- GEM 2.97 and cath. 3.57kV
12/24/14 Peak shift with changing the GEM preamp. Voltage
The following figure shows the peak shift in the Peak sensing charge spectrum as Drfit voltage is 700 V and the GEM voltage is changed as shown in the figure:
The figure above shows the peak shift as the GEM preamp. voltage is 2.77kV and 2.9kV. the higher voltage for the GEM preamp., the higher charge collected from the detector. Also the noise peak rate at channel is less at a higher GEM voltage.
12/24/14 GEM_V = 2.82 kV
cathode drift potential and count rate runs for 2.82kV GEM
12/18/14 Comparing increasing the GEM voltage 30 V to 2.9 KV
- GEM_V= 2.87 kV
- GEM_V= 2.9 kV
The results shows the integral under the charge spectra:
The maximum rate for the first peak in case of open shutter and in case of the subtraction is shown below:
- Notes
Increasing the the GEM voltage
- Increased the charge as the shutter is open as the drift voltage is in between 600-800V
- Increased the rate of the first peak rate, the rate at 800V and 900 V is not the same for all runs, the latest runs have higher rate, and the error bars are bigger for 800 and 900V.
- There is not any difference in maximum rate of the first peak for open shutter and the subtraction, so the shutter is completely stopping the charge flow to the anode as it is closed, even after the GEM's voltage increment.
- I will upload more graphs that have the average for each figure for the two values of the GEM voltages.
- 2.87 kV and 2.9 kV
The figure above shows that increasing the GEM voltage with 30 V does not change the amount of charge except for the drift voltage in the range of 500 to 700 V (increase) and 900 to 1kV (decrease). The first peak rate in increases when the GEM voltage increases 30 V as the drfit voltage in the range of 300-900 V as shown below:
12/15/14
Data analysis,
File:12 2 3 4 count rate cathV.xls
12/11/14
you forgot to put possible reason for fluctuation in legend. Listing the possible reasons below is not informative
Figure 1.) Open-Closed data , from 10/25/14 -> 12/9/14, color code different days, legend indicates date and possible reason for fluctuations.
Figure 2.) Open data , from 10/25/14 -> 12/9/14, color code different days, legend indicates date and possible reason for fluctuations.
Figure 3.) Closed data , from 10/25/14 -> 12/9/14, color code different days, legend indicates date and possible reason for fluctuations.
- Fluctuations
- Changing the bottle twice.
- Weather temperature.
- Ar/CO2 gas' temperature was lower than the lab's temperature for the last bottle, a change is observed before and after the 11/20/14.
Figure 4.) Open-Closed data , from 10/9-present, color code different days, legend indicates date and possible reason for fluctuations.
Figure 5.) Open data , from 10/9-present, color code different days, legend indicates date and possible reason for fluctuations.
Figure 6.) Closed data , from 10/9-present, color code different days, legend indicates date and possible reason for fluctuations.
- The effect of the drift field
File:Laura De Nardo, Marchiori Elena, PhD thesis,Università degli studi di Padova, 2013/2014
There is an effect for the drift potential on the count rate as its value is less than 0.4 kV, one of the reasons that the drift force line ends on the surface of the GEM electrode, on the other hand, when the drift more than 0.4 kV, the drift force lines converge inside the hole.
Another reason is in case of the drift potential is more than 0.4 kV, the probability of electron-ion recombination before the GEM electrode is less compared to the case when the drift potential is less than 0.4 kV.
If the electric field is higher than 0.8 kV, the electric filed lines end on the surface, and the ones in the holes do not converge as much as in the case of of V_drift =< 0.8 kV, which cause an electron defacusing effect.
File:Bachmann GEM charge transcfer properties driftE.pdf
12/10/14
Three figures
Figure 1.) Open-Closed data , from 10/25/14 -> 12/9/14, color code different days, legend indicates date and possible reason for fluctuations.
Figure 2.) Open data , from 10/25/14 -> 12/9/14, color code different days, legend indicates date and possible reason for fluctuations.
Figure 3.) Closed data , from 10/25/14 -> 12/9/14, color code different days, legend indicates date and possible reason for fluctuations.
Figure 4.) Open-Closed data , from 10/9-present, color code different days, legend indicates date and possible reason for fluctuations.
Figure 5.) Open data , from 10/9-present, color code different days, legend indicates date and possible reason for fluctuations.
Figure 6.) Closed data , from 10/9-present, color code different days, legend indicates date and possible reason for fluctuations.
Figure 7.) Overlay average of 10/25/14-> 12/9/10 measurements and average of 10/9-present, legend indicated GEM voltage,
12/09/14
- Reference for number pf counts vs the cathode voltage
12/08/14 2.9kV GEM
cathode drift potential and count rate runs for 2.9kV GEM
12/08/14
The figure shows the drift voltage vs the count rate in another weekend.
The graph is different from the graph taken by the previous weekend 11/29/14 as the drift potential is less than 350 V. as shown below
Investigate ionized electron transmission by looking at previous published studies of Cathode voltage differences.
12/05/14
- Beta Transmission ratio percentage in FR4
The figure below shows the number of beta transmitted through 1 mm FR4, the collected data considers only the incident beta only without counting for primary electrons as result of the ionization through the shutter and without electron ionization in the gas in the drift volume.
- Cathode vs Count Rate
The figure shows the count rate as the shutter is open (green), closed (red), and their subtraction (blue) as the drift voltage changes. The measurements were taken in the weekend.
The figures above show the count rate in three different working days, the color refernce is as the figure above.
File:12 2 3 4 count rate cathV.xls
The average of previous of the data for the three figures above is calculated with the error bars.
Notes:
- As mentioned below, as the drift potential is between 400-900V, the same count rate is measured in average for the the three cases.
- The count rate increase between 193V and 200V was not observed.
- As the voltage is 900V, in all days, a high noise is observed for shutter open. It confirms what we discussed before; it indicates a need for more amplification and a higher discrimination level if we are interested in determining the full charge spectrum from U-233.
- Another set of measurements will be taken this weekend to reproduce the graph for 11/29.
12/01/14
The figure shows the drift voltage effect on the number of counts for shutter open (green), closed (red), and their subtraction (blue)
I noticed:
- There is not any effect for increasing the cathode voltage when the drift voltage is 400V or higher.
- The count rate is doubled between as the drift voltage changes from 193V to 200V.
- Compared to the previous graph (measurements before 11/20/14) a shift in the voltage for the area that number of count rate dramatically increased, it changed from around 600V to 200V. Also the count rate decreased with about 10Hz (only for the subtraction graph in blue) after changing the old bottle.
- The data are measured in the weekend, and it would be measured again through the normal working days to compare the results.
11/29/14 Electron range in FR4
The calculations considers the excitation energy for FR4 equals to 107.1 eV.[3], and composite is from the site [4]
11/25
11/03-19 Ionization calculations
File:Total ionization 3particles.xls
11/20/14
A difference in the number is noticed after changing to the new bottle!!!!!
10/27/14
Plot
V -vs- RThe
V -vs- R graph seems to be jumping around a lot.
This may mean we need more measurements in order to determine if there is a smooth dependence or not
- Ar ionization cross section
The energy is in the unit of MeV/amu which is relative to the atomic mass unit of hte target. In our case we are interested in Ar which has an amu = 40.
10/24/14
Table of Histograms with NO SOURCE, shutter open and closed, and changing V cathode
(750 Volts)
Plot the rate of all three particles from source onto one plot
File:U-33 SourceParticlePercentage.xmgrace.txt
Electron energy threshold to penetrate 1mm thick FR4, and the energy for full penetration.
Plot transmission by taking ratio of (Number particle through shutter)/(Number of particles hitting shutter)
Plot Energy loss ( 100 *
)10/23/14
Table of Histograms with NO SOURCE, shutter open and closed, and changing V cathode
10/22/14
Long run peak sensing histograms
20min run peak sensing histograms
20 min. GEM-2.87 kV CATH-3.435 kV 10/17/14
you need to label the axis and color code the statistics box so I know which distribution corresponds to which configuration.
10/16/14
you need to label these histograms with more detail that will identify the run number and the run conditions. Name each histogram according to the run number then in a figure caption identify the run conditions.
you need to show more than just 4 runs, try to take 10, why is the pedestal peak different in the four that you show.
You need to have units of Hz for the count rate on the y-axis.
I should not be asking you for the above things at this stage of your career, you should be doing hem automatically
I am not giving any formal results yet since the voltage of the GEM may increase.
The point is to able to understand what is plotted and have a reference to fall back on
Formal results will be even more detailed
- Last update for the today's measurements
- Reproduciblity is achieved when for the peak sensing spectra for shutter open and shutter close without source.
- Increasing the GEm amplification is needed for the case of the shutter open with the source on.
Analyze all runs with same conditions of HV, Gas, … to determine the integral number of counts in the ADC(PADC) histogram that are above the pedestal.
10/14-15/14
Shutter | run numbers | average number of counts above the pedestal |
Open | 7797, 7802, 7808, 7816 | |
Closed | 7798, 7803, 7809, 7817 |
Fix the percentage plot for the number of betas emitted by the source.
The repcentages do not add to 100% according to two different references.[5][6] they add to about 20%, the first referece bentioned that the average energy for the emitted particles from U-233 are
Ebeta (eV) : 5.043230e+3 (5.363190e+2)
Egamma (eV) : 1.110210e+3 (1.076780e+2)
Ealpha (eV) : 4.888350e+6 (2.896760e+4)
Fix the alpha log plot for cross-section, change units on the beta plot for cross section so they are in barns.
Add plot for expected voltage on oscilloscope.
Alpha | |||
Beta | [7] | ||
Gamma | 300 px |
10/13/14
stripchart
- Relationship between barns and g/cm^2
paragraph describing percentage plots for alpha, beta, and gamma with references.
long run | time | shutter | source | Notes |
7832 | 6h | open | off | 8dB |
7834 | 35min | closed | off | 8dB |
7836 | 5.5h | open | on | 11 dB |
10/10/14
10/02/14
QDC's and Peak sensing's spectra distinguish between shutter open and shutter close as the source is on, I noticed it from yesterday's and today's measurements.
Also the spectra shows a difference in the number of count and the number of channels as the source on or off as the shutter is open. more measurements is needed to calculate the STDEV.
The oscilloscope picture is shows the gate and signal details,
date | time run ended | Source | run number | notes |
10/02/14 | 10:45 | On | 7792 | each run time is 20 min. |
10/02/14 | 11:07 | off | 7793 | 22', taking the source off directly then measuring the charge does not show any difference in channel number but shows difference in counts. |
10/02/14 | 16:30 | off | 7795 | 20', after 5 1/2 h the detector started to show a little difference again between shutter open source is on and when the source is off. |
10/03/14 | 8:00 | off | 7795 | 20' |
10/03/14 | 11:21 | off | 7795 | 20' equilibrium without source. |
10/06/14 | 09:05 | off | 7802 | 20' equilibrium without source. |
10/07/14 | 7:00 | off | 7806 | lower charge is detected |
10/07/14 | 11:19 | off | 7807 | higher charge is detected (TSO's stuff were checking on the source) |
GEM_HV= 2870 Volts, Drift HV = 3470 Volts
09/30/14
Plot of shutter open/closed (NO SOURCE) counter rate -vs- date
- Alpha's Primaries
Plot of Number of electrons collected -vs- Energy of (alpha, beta, and gamma) from U-233
- Electrons
[K Paludan et al 1997 J. Phys. B: At. Mol. Opt. Phys. 30 L581 doi:10.1088/0953-4075/30/17/005 ]
- Gamma
09/29/14
- Shutter open/close data plot for 2.87 3.48 kV GEM /Cathode
[8] The reference gives percentages of the emitted alpha particles as U-233 -> Th229 File:Alpha percentages.txt
- Roy's detector infomation and measurements
U-233 metal deposited source is measured by Protean Instrument corporation gaseous detector, has a model number of WPC9450 (serial number: 0915723)and uses (P10) gas mixture, as shown below:
Shutter position | Alpha particles /min. | Beta particles /min. |
Open | 6879 | 900 |
Close | 1 | 38 |
- 09/20/14
Rate of ( alpha, photon, beta) -vs- energy for U-233
Primary electron ionization -vs- (alpha,photon, beta) energy
- Alpha Secondaries
- Photo-Absorption Secondaries
The reason that the graph started from 30 keV is the lowest gamma energy emitted by U-233 or Cf-252 is higher than that energy.
- Electron Ionization
Using NIST data base [9]
we got the following data file for CO2 File:CO2 e ionization xsection.txt, For Ar, the ref. [10] that measured the ionization xsection.
09/20/14
The figure below shows the change in the signal as the GEM capacitor charges, at a specific fixed voltage it reaches saturation (equilibrium),
If the capacitor does not reach equilibrium, the signal of the detector is expected to change with time.
What may forbid the GEM capacitor to reach the equilibrium?
- The circuit board voltage fluctuations.
- The following reference describes almost the same conditions as those of our detector when their detector is in operation to detect simultaneously photons with alpha particles.
Nuclear Instruments and Methods in Physics Research A 471 (2001) 151–155
File:Gas electron multiplier for portal imaging wallmark.pdf
The author commented in the conclusion "The studies show that GEMs can operate at extremely high rates (>10^6 Hz/mm^2) with no sign of degradation and stability loss due to radiation damage. However, it was discovered that the maximum achievable gain for all planar gaseous detectors drops with the beam intensity "
"In real clinical operation the detector can operate safely with a gain of 10^2 in the GEM closest to the collector".
Our detector has a rate of 100-170 Hz (without and with the source), considering the detector age, will this rate cause an instability?
- High detector rate.
The gain will decrease when the count rate increases, if GEMs' voltage is at the point where gain is stable with the high rate, the detector output will be reproducible. As mentioned above, a gain of 10^2 made it enough for imaging with a well-quenched gas, I doubt we need to increase the gain more than that.
9/18/14
Determine best Cathode HV that produces the largest separation of the source ON/OFF signal.
The detector results are not reproducible, as the voltage is at 3.4 kV, the QDC spectrum is different as the source on; so QDC can not distinguish if the source is on or when it is off the detector.
Something is wrong!!!!
We never have a reproducibility problem before using the QDC until I start using V1495, Can we borrow the older module just to test the reproducibility is still a problem.
It is very unlikely that it is the v1495. The V1495 only tells the DAQ to read out a module. You can test the DAQ by injecting a pulse with a known charge and look for it in the QDC spectrum.
It is VERY important to have scope pictures showing the difference between source ON/OFF. Then you use scalers to check that your trigger pulse is able to see a difference between source ON/OFF Then you do the DAQ measurements.
If you don't follow the above proceedure then you will be building a pyramid on quicksand.
The same case when the cathode voltage increased to 3.6 kV.
Then perform a set of measurements (shutter open/closed and source ON/OFF) to establish reproduceability. Make sure you record the scaler count rates.
9/17/14
Measure the charge for several values of the Cathode HV keeping the GEM preamplifier voltage and gas flow rate constant.
- CATH 3.2 KV
- CATH 3.1 KV
All the runs have the same duration 20 min.
- CATH 3.4 KV
Try to take scope picture to show difference between source on and off signal that are being measured by the QDC.
9/16/14
- Condition of above results
Shutter is Open
Red is Cf-252 source ON Run 7724
Green is Cf-252 source OFF run 7726
HV_GEM= -2930 Volts HV_Cathode=-3400 Volts
Gas Flow rate = 0.1 ft^3/hr
- Goal
- Can the Cf-252 source ON signal be changed?
Change the Cathode voltage to try and turn off the signal when the source is on.
- Result
Yes, when the cathode voltage is decreased to -3100V, QDC histogram does not show any difference in the collected charge as the shutter is open with the source on it, and when the shutter is open without the source.
The figure below shows the change in the QDC spectrum when the cathode voltage is -3.2 kV. <references/>