Difference between revisions of "Cf252 runs with HPGe"
(80 intermediate revisions by 2 users not shown) | |||
Line 2: | Line 2: | ||
= Cf-252 source description= | = Cf-252 source description= | ||
− | + | Serial Number: FTC-CFZ-431 | |
+ | Z=98 | ||
+ | Mass:A document indicates that on 03/19/1993 the Cf-252 mass was 4.2<math>\mu g</math>. ( <math>1\mu g</math> is expected to emit <math>2.3 \times 10^6 n/s</math>) | ||
+ | |||
Half Life= 2.652 years | Half Life= 2.652 years | ||
− | Neutron fluence: | + | Neutron fluence:A document indicates that on 03/19/1993 the Cf-252 the neutron rate was measured(?) to be <math>10^{7} n/s</math>. Based on the mass one would expect <math>4 \times (2.3 \times 10^6 n/s) = 9.2 \times 10^6 n/s</math>. |
+ | |||
+ | Neutron energy spectrum (from A.B.Smith et.al, Physics. Rev. Vol 108 #2 pg 411-414, 1957 [[File:PhysRev108p411957.pdf]]): | ||
+ | |||
+ | [[File:Cf-252_NeutronE_PhysRevV108Pg412_1957.png | 100 px]] | ||
Decay: Cf-252 undergoes alpha decay to Curium (cm-248) 96.9% of the time. The rest of the time it will undergo spontaneous Fission emitting 3.7 neutrons on average per fission. | Decay: Cf-252 undergoes alpha decay to Curium (cm-248) 96.9% of the time. The rest of the time it will undergo spontaneous Fission emitting 3.7 neutrons on average per fission. | ||
+ | |||
+ | |||
+ | |||
+ | The Fission Fragment Yield as a function Fission fragment Atomic Number (<math>Z</math>) and Atomic Mass Number (<math>A</math>) from Wahl, A., At. Data and Nuc. Data Tab., vol. 39, (1988) . | ||
+ | |||
+ | [[File:Cf252-FissionProductsRate_NucDatTable1988.png | 200 px]] | ||
+ | |||
+ | most likely observable isotopes: | ||
+ | |||
+ | 40<Z<45 and 53<Z<58 | ||
+ | |||
+ | ==Previously measured Gamma spec== | ||
+ | |||
+ | [[File:Cf-252GammaSpec_100keV-1MeVA.gif | 100 px]][[File:Cf-252GammaSpec_1-2MeV.gif | 100 px]][[File:Cf-252GammaSpec_2-3MeV.gif | 100 px]][[File:Cf-252GammaSpec_4-5MeV.gif | 100 px]] | ||
+ | |||
+ | Fragments to try and observe: | ||
+ | |||
+ | <math>{A \atop Z} X_N</math> = | ||
+ | Ac-228 | ||
+ | |||
+ | <math>{214 \atop 83} Bi_{131}</math> | ||
+ | Bi-214 | ||
+ | |||
+ | Tl-208 | ||
== HpGe distribution== | == HpGe distribution== | ||
+ | |||
+ | |||
+ | Test Run 5603 may be seeing Bi-214 line 1145 keV | ||
+ | |||
+ | [[File:r5602-5603Overlay.png | 100 px]] | ||
==HpGe Neutron Damage== | ==HpGe Neutron Damage== | ||
Line 21: | Line 57: | ||
+ | ==HpGe Singles measurement== | ||
+ | |||
+ | Measure the energy distribution observed by the HpGe detector | ||
+ | |||
+ | Determine optimal neutron shielding between HpGe detector and Cf-252 source | ||
+ | |||
+ | ==Small Scintillator position== | ||
+ | |||
+ | Determine how close the small scintillator can be positioned without having detected neutrons overwhelm detector photons. Want to be sure the small scintillator is detecting photons most of the time | ||
+ | |||
+ | == Determine timing resolution of HpGe Detector== | ||
+ | Below is a time difference plot of coincidence 511 keV photons from Na-22. The Na-22 source is 3" from both the timing scintillator and the HPGe detector. The following photon energy cut is applied: 508 < erg < 514. All coincidence events disappear if the source is removed, so there are virtually no accidentals. (run 5736). | ||
+ | |||
+ | [[File:HPGe timing.png|700px]] | ||
+ | |||
+ | == search for fission fragment gammas == | ||
+ | I completed a run with the CF252 source where all events were 3-fold coinc between the HPGe detector, a small timing scintillator nearby the source, and a hit in at least one neutron detector (run 5743). Neutrons were then discriminated in software by ToF, which is made possible by means of the timing scintillator. | ||
+ | |||
+ | I compared the photon energies of events with a neutron hit, to energies from a non-coincidence background subtracted CF252 spectrum (runs 5775 & 5773). I made a histogram for both cases and normalized the integrals to unity in oder to allow me to search for gammas that occur overwhelmingly when a neutron is coincident. I failed to see any pronounced energies occurring only when in coincidence with a neutron. | ||
+ | |||
+ | |||
+ | [[File:CoincVSNONCoinceCF252.png| 850px]] | ||
+ | |||
+ | ==Estimate run time for efficiency measurement== | ||
=Runlist= | =Runlist= | ||
Line 37: | Line 97: | ||
|- | |- | ||
| 5614 || 654 || 15 || 0 || 13885 || Co-60 source. Coincidence between HPGe and timing PMT || Calibration | | 5614 || 654 || 15 || 0 || 13885 || Co-60 source. Coincidence between HPGe and timing PMT || Calibration | ||
+ | |- | ||
+ | | colspan="7" | '''Total n flux: 13885 n/cm^2''' | ||
+ | |} | ||
+ | |||
+ | |||
+ | ==01/01/17== | ||
+ | CF-252 source is kept at a fixed distance while the thickness of poly between src and HPGe is increased by increments of 2". | ||
+ | {| border="1" style="border-collapse: collapse; padding: 5px; text-align: center " | ||
+ | !Run # !! RunTime [sec] !! distance[cm] !! n flux <math>[\frac{n}{cm^2}]</math><br><math>=\frac{n_{rate}t}{4\pi d^2}</math> <br>([[Cf-252 rates table|<small>Cf252 rates</small>]]) || Integral <br>n flux <br> <math>[\frac{n}{cm^2}] </math> ||configuration || num events || notes | ||
+ | |- | ||
+ | | 5633 || 372 || rowspan="6" |49 || 235 || 235 || No poly. || 303,929 || | ||
+ | |- | ||
+ | | 5634 || 371 || 235 || 470 || 2" poly. || 250,495 || | ||
+ | |- | ||
+ | | 5635 || 371 || 235 || 705 || 4" poly. || 202,942 || | ||
+ | |- | ||
+ | | 5637 || 397 ||256 || 956 || 6" poly. ||181,667 || | ||
+ | |- | ||
+ | | 5638 || 371 || 235 || 1190 || 8" poly. || 147,801 || | ||
+ | |- | ||
+ | | 5639 || 371 || 235 || 1425 || 10" poly. || 134,095 || | ||
+ | |- | ||
+ | | 5641 || 110 || - || - || - || -- || || CO-60 + Na-22 cal. run | ||
+ | |- | ||
+ | | 5642 || 291 || - || - || - || -- || ||background run | ||
+ | |- | ||
+ | |} | ||
+ | |||
+ | ==01/12/17 - 01/18/17== | ||
+ | |||
+ | Discriminator threshold for timing PMT, neutron PMT's, and HPGE are 50, 35, & 70 mV, respectively. When calculating neutron flux I reduced the result by factor of 4 because of the 4" of poly shielding the HPGe detector. | ||
+ | |||
+ | {| border="1" style="border-collapse: collapse; padding: 5px; text-align: center " | ||
+ | !Run # !! RunTime [sec] !! distance[cm] !! n flux <math>[\frac{n}{cm^2}]</math><br><math>=\frac{n_{rate}t}{4\pi d^2}</math> <br>([[Cf-252 rates table|<small>Cf252 rates</small>]]) || Integral <br>n flux <br> <math>[\frac{n}{cm^2}] </math> ||configuration || notes | ||
+ | |- | ||
+ | | 5743 || 57515 || rowspan="3" | 10 || 218000|| 218000 || CF252 src. 4" Poly. 3-fold coincidence between timing scint., HPGe, and at least one neutron PMT. || - | ||
+ | |- | ||
+ | | 5766 || 1546 || 6000 || 224000|| CF252 src. 4" Poly. '''2-fold''' coincidence between timing scint and HPGe || - | ||
+ | |- | ||
+ | | 5773 || 780 || 3000 || 227000 || CF252 src. 4" Poly. '''no coincidence'''|| - | ||
+ | |- | ||
+ | | 5775 || 780 || N/A ||rowspan="3" | 0|| rowspan="3"| 227000 || '''No src'''. 4" Poly. no coincidence|| Background run | ||
+ | |- | ||
+ | | 5769 ||21097 || rowspan="2" | 10 ||'''Na-22'''. 4" Poly. '''2-fold''' coincidence between timing scint and HPGe|| Na-22 calibration run | ||
+ | |- | ||
+ | | 5770 || 1211 ||'''C0-60'''. 4" Poly. 2-fold coincidence between timing scint and HPGe|| Co-22 calibration run | ||
+ | |||
+ | |||
+ | |||
|- | |- | ||
|} | |} | ||
Line 43: | Line 152: | ||
[[2nCor 44|go back]] | [[2nCor 44|go back]] | ||
− |
Latest revision as of 06:45, 19 January 2017
Detector efficiency measurements using a Cf-252 source.
Cf-252 source description
Serial Number: FTC-CFZ-431 Z=98 Mass:A document indicates that on 03/19/1993 the Cf-252 mass was 4.2
. ( is expected to emit )
Half Life= 2.652 years
Neutron fluence:A document indicates that on 03/19/1993 the Cf-252 the neutron rate was measured(?) to be
. Based on the mass one would expect .Neutron energy spectrum (from A.B.Smith et.al, Physics. Rev. Vol 108 #2 pg 411-414, 1957 File:PhysRev108p411957.pdf):
Decay: Cf-252 undergoes alpha decay to Curium (cm-248) 96.9% of the time. The rest of the time it will undergo spontaneous Fission emitting 3.7 neutrons on average per fission.
The Fission Fragment Yield as a function Fission fragment Atomic Number (
) and Atomic Mass Number ( ) from Wahl, A., At. Data and Nuc. Data Tab., vol. 39, (1988) .most likely observable isotopes:
40<Z<45 and 53<Z<58
Previously measured Gamma spec
Fragments to try and observe:
= Ac-228
Bi-214
Tl-208
HpGe distribution
Test Run 5603 may be seeing Bi-214 line 1145 keV
HpGe Neutron Damage
I made a page discussing neutron damage to the HPGe detector (Fast_neutron_damage_to_HPGe_Detector) as a function of the total amount of fast neutrons per irradiated on the detector face. is a factor of 10 below the point where damage was first observed in an experiment.
RunPlan
Object: Determine if an HpGe detector can be using to tag neutrons and identify the isotope species parent of the neutron
HpGe Singles measurement
Measure the energy distribution observed by the HpGe detector
Determine optimal neutron shielding between HpGe detector and Cf-252 source
Small Scintillator position
Determine how close the small scintillator can be positioned without having detected neutrons overwhelm detector photons. Want to be sure the small scintillator is detecting photons most of the time
Determine timing resolution of HpGe Detector
Below is a time difference plot of coincidence 511 keV photons from Na-22. The Na-22 source is 3" from both the timing scintillator and the HPGe detector. The following photon energy cut is applied: 508 < erg < 514. All coincidence events disappear if the source is removed, so there are virtually no accidentals. (run 5736).
search for fission fragment gammas
I completed a run with the CF252 source where all events were 3-fold coinc between the HPGe detector, a small timing scintillator nearby the source, and a hit in at least one neutron detector (run 5743). Neutrons were then discriminated in software by ToF, which is made possible by means of the timing scintillator.
I compared the photon energies of events with a neutron hit, to energies from a non-coincidence background subtracted CF252 spectrum (runs 5775 & 5773). I made a histogram for both cases and normalized the integrals to unity in oder to allow me to search for gammas that occur overwhelmingly when a neutron is coincident. I failed to see any pronounced energies occurring only when in coincidence with a neutron.
Estimate run time for efficiency measurement
Runlist
12/29/16
No poly was used to shield the HPGe detector.
Run # | RunTime [sec] | distance[cm] | n flux (Cf252 rates) |
Integral n flux |
configuration | notes |
---|---|---|---|---|---|---|
5609 | 4060 | 30 | 6850 | 6850 | Cf252 in. Coincidence between HPGe and timing PMT | A ToF spectrum exists but is very hectic. Problem may be that the TDC stop is determined by a widened timing PMT pulse in coinc. with a HPGe pulse that has high timing jitter. |
none | 1250 | 26 | 2800 | 9655 | Not taking data during this time. | |
5611 | 2509 | 30 | 4230 | 13885 | Cf252 in. Coincidence between HPGe and timing PMT. TDC stop was changed to always be 1200 ns after PMT timing pulse, otherwise same setup as 5609. | ToF looks good. Neutron rate approx. 3 Hz for detector 90. |
5613 | 10025 | 15 | 0 | 13885 | Na-22 source. Coincidence between HPGe and timing PMT | Calibration |
5614 | 654 | 15 | 0 | 13885 | Co-60 source. Coincidence between HPGe and timing PMT | Calibration |
Total n flux: 13885 n/cm^2 |
01/01/17
CF-252 source is kept at a fixed distance while the thickness of poly between src and HPGe is increased by increments of 2".
Run # | RunTime [sec] | distance[cm] | n flux (Cf252 rates) |
Integral n flux |
configuration | num events | notes |
---|---|---|---|---|---|---|---|
5633 | 372 | 49 | 235 | 235 | No poly. | 303,929 | |
5634 | 371 | 235 | 470 | 2" poly. | 250,495 | ||
5635 | 371 | 235 | 705 | 4" poly. | 202,942 | ||
5637 | 397 | 256 | 956 | 6" poly. | 181,667 | ||
5638 | 371 | 235 | 1190 | 8" poly. | 147,801 | ||
5639 | 371 | 235 | 1425 | 10" poly. | 134,095 | ||
5641 | 110 | - | - | - | -- | CO-60 + Na-22 cal. run | |
5642 | 291 | - | - | - | -- | background run |
01/12/17 - 01/18/17
Discriminator threshold for timing PMT, neutron PMT's, and HPGE are 50, 35, & 70 mV, respectively. When calculating neutron flux I reduced the result by factor of 4 because of the 4" of poly shielding the HPGe detector.
Run # | RunTime [sec] | distance[cm] | n flux (Cf252 rates) |
Integral n flux |
configuration | notes |
---|---|---|---|---|---|---|
5743 | 57515 | 10 | 218000 | 218000 | CF252 src. 4" Poly. 3-fold coincidence between timing scint., HPGe, and at least one neutron PMT. | - |
5766 | 1546 | 6000 | 224000 | CF252 src. 4" Poly. 2-fold coincidence between timing scint and HPGe | - | |
5773 | 780 | 3000 | 227000 | CF252 src. 4" Poly. no coincidence | - | |
5775 | 780 | N/A | 0 | 227000 | No src. 4" Poly. no coincidence | Background run |
5769 | 21097 | 10 | Na-22. 4" Poly. 2-fold coincidence between timing scint and HPGe | Na-22 calibration run | ||
5770 | 1211 | C0-60. 4" Poly. 2-fold coincidence between timing scint and HPGe | Co-22 calibration run
|