A few Characteristic scope pictures of event triggered on the chamber output.  The trigger threshold was set to ~-14 mV for all but the last picture which was set at -17 mV.  Signals beyond -14 mV are well outside the intrinsic detector noise and considered ionization events.
+
A few Characteristic scope pictures of event triggered on the chamber output.  The trigger threshold was set to ~-14 mV for all but the last picture which was set at -17 mV.  Signals beyond -14 mV are beond the intrinsic detector noise and considered ionization events.
  
 
[[Image:MiniChamber_Singles_A_10-9-07.jpg | 400 px]]
 
[[Image:MiniChamber_Singles_A_10-9-07.jpg | 400 px]]

Revision as of 01:51, 10 October 2007

Signals were observed in the LDS on the Mini DC prototype sense wires capacitively coupled to a digital oscilloscope. The drift chamber was filled with a Ar/CO2 (75/25) gas mixture.

The apparatus

MiniChamber Setup 10-9-07.jpg

The following Voltages were established on the Drift chamber wires

HV settings
Wire Volts
Sense 1800
Field -750
Guard 750

The Scintillator mounted on top of the Chamber was set to -900 Volts

Below is a scope picture representing the observation of a coincidence event between a particle which passed through both the top scintillator and the Mini Drift Chamber. The scope was triggered on the scintillator output. The top trace is the output of the drift chamber from one of the sense wires capacitively coupled to the scope and terminated in 1 Meg Ohm. The bottom trace is the fast output from the Scintilaltor.

MiniChamber CosmicCoincidence 10-9-07.jpg

Noise Level Measured by scope in infinite persistance mode indicate that the Mini Prototype with plastic feedthroughs (Plastica) has noise levels spanning +/- 10 mV. A PMT connected to a scintillator shows noise levels around+/- 2 mV.


PlasticaNoise 10-9-07.jpg


A few Characteristic scope pictures of event triggered on the chamber output. The trigger threshold was set to ~-14 mV for all but the last picture which was set at -17 mV. Signals beyond -14 mV are beond the intrinsic detector noise and considered ionization events.

MiniChamber Singles A 10-9-07.jpg

MiniChamber Singles B 10-9-07.jpg

MiniChamber Singles C 10-9-07.jpg

MiniChamber Singles D 10-9-07.jpg

MiniChamber Singles E 10-9-07.jpg

MiniChamber Singles F 10-9-07.jpg

MiniChamber Singles G 10-9-07.jpg


Back to MiniPrototype Page