Difference between revisions of "CLAS12 R1 S1 QA"

From New IAC Wiki
Jump to navigation Jump to search
Line 34: Line 34:
 
|}
 
|}
  
[[File:DChitDisplay_3831,png | 200 px]]
+
[[File:DChitDisplay_3831.png | 200 px]]
  
 
[[R3811_TDCspectra]]
 
[[R3811_TDCspectra]]

Revision as of 16:41, 16 February 2013

HV burn in

Begin burn in after at least 3 volume changes.


File:High Voltage Region 1 Sector 1 data.xls

Measured Sense wire output

CLAS12_R1_S1_QA_wireOut


Sense Wire Scope pictures on DAQ machine

Cosmic Runs

CLAS12R1_CosmicTestStand

3811

Cosmic rays were used to measure drift chamber time spectra with respect to scintillators placed above and below the drift chambers (more details are at the wiki location CLAS12R1_CosmicTestStand)


CLAS6 DCrawTDCspectrum.gif R3811 TDC51 Time.png

R3831 3833TDC51 TimeL1W9.png

Fig 23 from "The CLAS drift chamber system", M. Mestayer, et. al., NIM A, Volume 449, Issues 1–2, 11 July 2000, Pages 81–111 cosmic run from 1/10 - 1/14. Signal cable 4 (ch. 48-63). 16 channels with leading edge discriminators (-80 mV, 100 ns pulse width). Cosmics runs with lower discriminator thresholds

DChitDisplay 3831.png

R3811_TDCspectra

3833

R3833_TDCspectra

Cosmic measurements

Run # Date Description
3806 1/10/13 3 hour cosmic run, 16 channels with leading edge discriminators (-80 mV, 100 ns pulse width)
3809 1/10/13 cosmic run on 1/10. Signal cable 4 (ch. 48-63). 16 channels with leading edge discriminators (-80 mV, 100 ns pulse width).
3810 1/10/13 cosmic run from 1/10 - 1/14. Signal cable 4 (ch. 48-63). 16 channels with leading edge discriminators (-80 mV, 100 ns pulse width). Scintillator paddle F1/F2 on table above Sector 1, paddle I1/I2 below.
3811 1/10/13
3831 2/1/13 Changed disciminator from 80 mV to 60 mV, We improved the window, installed HV & STB boards, purged and started taking cosmic data again, DC is sandwhiched between 2 scints with PMTs aligned (PMT on only one end of Scint is turned on)
3833 2/4/13 Changed disciminator from 60 mV to 30 mV
3834 2/11/13 Changed disciminator from 30 mV to 60 mV

Readout Map

Wire PostAmpOut (CH 1-96) TDC Channel
L W
5 1 1 1
3 1 2 2
1 1 3 3
6 1 4 4
4 1 5 5
2 1 6 6
5 2 7 7
3 2 8 8
1 2 9 9
6 2 10 10
4 2 11 11
2 2 12 12
5 3 13 13
3 3 14 14
1 3 15 15
6 3 16 16
4 3 17 17
2 3 18 18
5 4 19 19
3 4 20 20
1 4 21 21
6 4 22 22
4 4 23 23
2 4 24 24
5 5 25 25
3 5 26 26
1 5 27 27
6 5 28 28
4 5 29 29
2 5 30 30
5 6 31 31
3 6 32 32
1 6 33 33
6 6 34 34
4 6 35 35
2 6 36 36
5 7 37 37
3 7 38 38
1 7 39 39
6 7 40 40
4 7 41 41
2 7 42 42
5 8 43 43
3 8 44 44
1 8 45 45
6 8 46 46
4 8 47 47
2 8 48 48
5 9 49 49
3 9 50 50
1 9 51 51
6 9 52 52
4 9 53 53
2 9 54 54
5 10 55 55
3 10 56 56
1 10 57 57
6 10 58 58
4 10 59 59
2 10 60 60
5 11 61 61
3 11 62 62
1 11 63 63
6 11 64 64
4 11 65 65
2 11 66 66
5 12 67 67
3 12 68 68
1 12 69 69
6 12 70 70
4 12 71 71
2 12 72 72
5 13 73 73
3 13 74 74
1 13 75 75
6 13 76 76
4 13 77 77
2 13 78 78
5 14 79 79
3 14 80 80
1 14 81 81
6 14 82 82
4 14 83 83
2 14 84 84
5 15 85 85
3 15 86 86
1 15 87 87
6 15 88 88
4 15 89 89
2 15 90 90
5 16 91 91
3 16 92 92
1 16 93 93
6 16 94 94
4 16 95 95
2 16 96 96

Measurements with source

4/20/12 1hour counts:

SL2 HVB1 Column 18 Row 5

-With source 14459 pulses/1502 sec


4/23/12 SL1 HVB7 column 203 row 4

-With source 50681 pulses/3760 sec.


4/25/12 1 hour counts:

SL2 HVB4 Column 118 row 5

- Without source: 33145 pulses/3610 Sec.

- With source(CS-137):38262 pulses/3600 Sec.

- Without source: 37548 pulses/3715 Sec.

Preamp voltage:5V

Background measurements

HV off 4/5/12 Background counts:

27 pulses/61806 Sec.

SL2 Column 132 row 2

Preamp voltage 6.3V

Discriminator threshold 46mV


Preamp on= 6.3

4/9/12: Back ground count measurements

4531 counts/69564 seconds

2044 counts/24620 seconds


4/10/12 Background counts:

3 pulses/ 74500 sec.

SL1 HVB7 Column 196 Row 2

Preamp voltage 6.3 V

Discriminator threshold 46mV


4/16/12 Background counts:

280 pulses/245850 sec

SL1 HVB7 column 226 row 2

preamp voltage 5V

Discriminator threshold 46mV


4/17/12 Background counts:

6 pulses/62407 sec

SL2 HVB7 column 226 row 8

preamp voltage 5V

discriminator threshold 46mV


4/20/12 Background counts:

63 pulses/ 71724 Sec.

SL1 HVB7 column 226 row 8

Preamp voltage 5V

Discriminator threshold 46mV


4/23/12 Background counts:

41078 pulse/ 245104 sec.

SL1 HVB7 Column 203 Row 4

Preamp voltage 5V

Discriminator threshold 46mV

Cosmic measurements

4/24/12 Cosmic count:

SL1 HVB1 Column 3 Row 10

14534 pulses/11547 sec.

Preamp voltage 5V


4/26/12 Overnight Cosmic pulse count:

SL2 HVB4 Column 118 Row 5

53122 pulses/49793 Sec.

Preamp voltage 5V



CLAS12_R1