Calibration
HpGe HV = 3700 kV
NaI Hv = 1300 V
Run # |
MPA file name |
Description
|
40 |
NaI040.mpa |
Na-22 source in front of HpGe and Co-60 on top of NaI detector
|
41 |
NaI019.mpa |
Co-60 source in front of HpGe and Na-22 on top of NaI detector
|
Last Dipole Sweep
Tune 10 Mev onto Tungsten and 4 MeV into Cell
We put in two thin pieces (thickness ~1/4" each) making a ~1/2" of lead angled at ~30 degrees wrt vertical between the NaI detector and the collimator to cut down on low-energy (< 200 keV) photons.
Chad Tune
Run # |
MPA file name |
W |
Ta |
Description
|
45 |
NaI045.mpa |
IN |
IN |
Dipole at 2 MeV (0.78 on Pot), 60 Hz rep rate
|
46 |
NaI046.mpa |
IN |
IN |
Dipole at 2 MeV (0.78 on Pot), 300 Hz rep rate, 116 [math]\pm[/math] 15 counts at 500 keV, Live=929 seconds
|
47 |
NaI047.mpa |
IN |
Out |
Dipole at 2 MeV (0.78 on Pot), 300 Hz rep rate, 53 [math]\pm[/math] 9 counts at 500 keV, Live=607 seconds, (Tant IN/Out ratio =116/53 [math]\times[/math] 607/929 = 1.4)
|
48 |
NaI048.mpa |
IN |
Out |
Dipole at 3 MeV (1.13 on Pot), 300 Hz rep rate, 90 [math]\pm[/math] 11 counts at 500 keV, Live=600 seconds, ignore NaI (cable not plugged into scope)
|
49 |
NaI049.mpa |
IN |
In |
Dipole at 3 MeV (1.13 on Pot), 300 Hz rep rate, 113[math]\pm[/math] 12 counts at 500 keV, Live=634 seconds,plugged NaI cable back into scope in middle of run. ((Tant IN/Out ratio =113/90 = 1.26 )
|
Accelerator Gun problem (ground wire shielding was worn away and creating path for short)
Kevin Tune
Restore tune with settings below.:
Device |
setting
|
RS1 |
-14.96
|
RS2 |
+0
|
FS1 |
0
|
FS2 |
8.58
|
0Q1 |
-14.75
|
0Q2 |
+2.26
|
45Q1 |
-0.62
|
45Q2 |
+1.88
|
L90Q1 |
-2.62
|
L90Q2 |
+2.83
|
Bend1 |
-3.42
|
Bend2 |
-5.3 (3 MeV)
|
Gun HV |
3 kV
|
Grid |
1.64
|
802 |
15.7 kV
|
SOL1 |
0
|
SOL2 |
3.58
|
Run # |
MPA file name |
W |
Ta |
Description
|
56 |
NaI056.mpa |
IN |
In |
Dipole at 3 MeV (1.11 on Pot), 300 Hz rep rate, (802 set to 15.7 kV. 48[math]\pm[/math] 8 counts at 500 keV, Live=600 seconds, grid = 1.64, Gun = 3 kV.
|
57 |
NaI057.mpa |
IN |
Out |
Dipole at 3 MeV (1.11 on Pot), 300 Hz rep rate, (802 set to 15.7 kV, 24 [math]\pm[/math]6 counts at 500 keV, Live=600 seconds, grid = 1.64, Gun = 3 kV. ( Tant In/out ratio = 2/1)
|
58 |
NaI058.mpa |
IN |
Out |
Dipole at 4 MeV (1.44 on Pot), 300 Hz rep rate, (802 set to 15.7 kV, 25[math]\pm[/math]6 counts at 500 keV, Live=600 seconds, grid = 1.64, Gun = 3 kV.
|
59 |
NaI059.mpa |
IN |
In |
Dipole at 4 MeV (1.44 on Pot), 300 Hz rep rate, (802 set to 15.7 kV, 25[math]\pm[/math]6 counts at 500 keV, Live=600 seconds, grid = 1.64, Gun = 3 kV.
|
60 |
NaI060.mpa |
IN |
In |
Dipole at 3 MeV (1.44 on Pot), 300 Hz rep rate, (802 set to 15.7 kV, 27[math]\pm[/math]6 counts at 500 keV, Live=600 seconds, grid = 1.64, Gun = 3 kV.
|
61 |
NaI061.mpa |
IN |
In |
Dipole at 3 MeV (1.44 on Pot), 300 Hz rep rate, (802 set to 15.7 kV, 1kG Permanent sweeper magnet upstream of last quad. 14 [math]\pm[/math]7 counts at 500 keV, Live=600 seconds, grid = 1.64, Gun = 3 kV.
|
63 |
NaI063.mpa |
IN |
In |
Quads after last dipole set to zero (all above runs have quad polarity for electrons), Dipole at 3 MeV (1.44 on Pot), 300 Hz rep rate, (802 set to 15.7 kV, 102[math]\pm[/math]11 counts at 500 keV, Live=600 seconds, grid = 1.64, Gun = 3 kV.
|
64 |
NaI064.mpa |
IN |
In |
swap polarity of Quads after last dipole set to: 45Q2= -1.37 L90Q1=+2.98 L90Q2=-0.98, Dipole at 3 MeV (1.44 on Pot), 300 Hz rep rate, (802 set to 15.7 kV, 64[math]\pm[/math]9 counts at 500 keV, Live=600 seconds, grid = 1.64, Gun = 3 kV.
|