10-30-08 GEM HRRL Photon Runs

From New IAC Wiki
Jump to navigation Jump to search

The Goal today is to observe the GEM detector output when hit with MeV photons

14 MeV photons

We are sending 14 MeV phtoton through a 4" thick brick wall collimator. A 4" Al brick in front of the HRRL exit port is used as a FC.


FC: 223B2 NaI: 223B5 Acc Gate: 223B8 GEMtrig: 223B3

Vdrift= 3700 Volt

VGem=3400 IGem=815 microA


The scope trace below shows output when electron beam current is 9 mA.

10-30-08 14MeV9mAHRRL GEM strip8.png

The scope trace below shows output when electron beam current is 18 mA.

10-30-08 18mAHRRL GEM strip8.png

The scope trace below shows output when electron beam current is 36 mA.

10-30-08 14MeV36mAHRRL GEM strip8.png

The scope trace below shows output when electron beam current is 55 mA.

10-30-08 14MeV55mAHRRL GEM strip8.png


Now Max Gain on GEM

Vdrift= 3800 Volt

VGem=3500 IGem=841 microA

The scope trace below shows output when electron beam current is 9 mA.

10-30-08 14MeV9mAHRRL GEM strip8 Vd3800.png

The scope trace below shows output when electron beam current is 4 mA.

10-30-08 14MeV4mAHRRL GEM strip8 Vd3800.png

10 MeV photons

The scope trace below shows output when electron beam current is 20 mA.

10-30-08 10MeV20mAHRRL GEM strip8 Vd3800.png

Peak amplitude is same but rate is much lower