| Date |
Time (24h) |
HV Board ID |
Guard (V) |
Field (V) |
Sense (V) |
Sense ([math]\mu[/math]A) |
Field ([math]\mu[/math]A) |
comments
|
| 03/11/13 |
16:50 |
S6 SL2 HVB5 |
50 |
-50 |
50 |
0.000 |
0.065 |
Probed traces and pins connected to boots for fields/guard, all are showing voltage.
|
| 03/11/13 |
16:56 |
S6 SL2 HVB5 |
536 |
-750 |
1500 |
11.5 |
11.7 |
Peaked around 13 [math]\mu[/math]A initially, but is rapidly decreasing after sense wires stabilized at operating voltage. Trip is set to 50 [math]\mu[/math]A, no wires tripped during voltage rise at 5 V/s.
|
| 03/11/13 |
17:11 |
S6 SL2 HVB5 |
536 |
-750 |
1500 |
10.7 |
10.8 |
|
| 03/11/13 |
17:32 |
S6 SL2 HVB5 |
536 |
-750 |
1500 |
10.2 |
10.3 |
|