Imaging System for Emittance Measurement of IAC Medical Linac
Abstract
Introduction
We used 3 lenses to collect OTR light from Al target. They have focal length of 100 mm, 500 mm and 50 mm, they were called collector lens, fine tune lens and main focus lens respectively. Collector lens has the middle focal length, collects most of the light from the source. Fine tune lens has the biggest focal length, act as fine tune of magnification. Main focus lens: has the smallest focal length, has biggest focusing strength, so act as main focus of light and focuses light to CCD.
Experiment Setup
Camera cage system was located below the OTR target. There are 3 lenses used to focus lights from target to the CCD camera. Target can be pushed into or taken out of the beamline by the actuator at the top.