TGEMS

From New IAC Wiki
Revision as of 17:58, 23 January 2009 by Oborn (talk | contribs)
Jump to navigation Jump to search

TGEMS : defined as GEM detector which use thick PCboard (0.4 mm thick) size GEM preamplifiers with hole diameters between 0.5 and 1 mm.

PCboard spacing = 1.5 - 2 mm.

[1], Media:01352098.pdf

TGEM Detector HV setting

[2]Media:Development_of_Tracking_Detectors_with_industrially_produced_GEM_Foils.pdf

[math]\displaystyle{ V_{GEM} = 0.103 \times V_{Drift} }[/math]

[math]\displaystyle{ V_{GEM} }[/math] is between ~[300V, 600V] for [math]\displaystyle{ CF_4 }[/math] and for [math]\displaystyle{ Ar/CO_2 }[/math] it is [200V, 370V];

[math]\displaystyle{ V_{Drift} }[/math] voltage is ~[2900V, 5800V] ([1950V, 4000V]) for [math]\displaystyle{ CF_4 }[/math] ([math]\displaystyle{ Ar/CO_2 }[/math]).




Descriminated signals

Test Setup 1

Cathode HV (Volts) First THGEM foil(Volts) Second THGEM foil(Volts) Third THGEM foil(Volts) Third THGEM foil current [math]\displaystyle{ \mu }[/math]A
3000 1756 1657 1652 1273



I used 2 Ortec timing amps to integrate the signal and amplify more than what we did for the standard GEM.

The following setttings were

Parameter Dial setting
Timing Amp1 x20
Timing Amp1 Integration 50 ns
Timing Amp2 x4
Timing Amp2 Integration 20 ns
CFD Threshold 5.22
CFD Threshold ULD 5.28
CFD Threshold LLD 3.04






Neutron_TGEM_Detector_Abdel

Go back