The 12 bit ADC conversion time is 5.7
s. A sliding scale technique is used to reduce differential nonlinearity by injecting an offset. This reduces the dynamic range to 0 3840. ADC values between 3841 and 4095 indicate an error.The TDC measures the time interval between the falling ECL inputs to the start/stop and the channel.
You can change the dynamic range of the TDC by setting bits using the library function c775setFSR.
Address | |
Pin | Setting |
SW 1 | 2 |
SW 2 | 2 |
SW 3 | 0 |
SW 4 | 0 |
ROC library
Compiling library
initialize TDC
-> c775Init(0x610000) Initialized TDC ID 0 at address 0x90610000