Roy's detector infomation and measurements
U-233 metal deposited source is measured by Protean Instrument corporation gaseous detector, has a model number of WPC9450 (serial number: 0915723)and uses (P10) gas mixture, as shown below:
Shutter position | Alpha particles /min. | Beta particles /min. |
Open | 6879 | 900 |
Close | 1 | 38 |
Rate of ( alpha, photon, beta) -vs- energy for U-233
Primary electron ionization -vs- (alpha,photon, beta) energy
- Alpha Secondaries
- Photo-Absorption Secondaries
The reason that the graph started from 30 keV is the lowest gamma energy emitted by U-233 or Cf-252 is higher than that energy.
- Electron Ionization
Using NIST data base [1]
we got the following data file for CO2 File:CO2 ionization xsection.dat, For Ar, the ref. [2] that measured the ionization xsection.
09/20/14
The figure below shows the change in the signal as the GEM capacitor charges, at a specific fixed voltage it reaches saturation (equilibrium),
If the capacitor does not reach equilibrium, the signal of the detector is expected to change with time.
What may forbid the GEM capacitor to reach the equilibrium?
- The circuit board voltage fluctuations.
- The following reference describes almost the same conditions as those of our detector when their detector is in operation to detect simultaneously photons with alpha particles.
Nuclear Instruments and Methods in Physics Research A 471 (2001) 151–155
File:Gas electron multiplier for portal imaging wallmark.pdf
The author commented in the conclusion "The studies show that GEMs can operate at extremely high rates (>10^6 Hz/mm^2) with no sign of degradation and stability loss due to radiation damage. However, it was discovered that the maximum achievable gain for all planar gaseous detectors drops with the beam intensity "
"In real clinical operation the detector can operate safely with a gain of 10^2 in the GEM closest to the collector".
Our detector has a rate of 100-170 Hz (without and with the source), considering the detector age, will this rate cause an instability?
- High detector rate.
The gain will decrease when the count rate increases, if GEMs' voltage is at the point where gain is stable with the high rate, the detector output will be reproducible. As mentioned above, a gain of 10^2 made it enough for imaging with a well-quenched gas, I doubt we need to increase the gain more than that.
9/18/14
Determine best Cathode HV that produces the largest separation of the source ON/OFF signal.
The detector results are not reproducible, as the voltage is at 3.4 kV, the QDC spectrum is different as the source on; so QDC can not distinguish if the source is on or when it is off the detector.
Something is wrong!!!!
We never have a reproducibility problem before using the QDC until I start using V1495, Can we borrow the older module just to test the reproducibility is still a problem.
It is very unlikely that it is the v1495. The V1495 only tells the DAQ to read out a module. You can test the DAQ by injecting a pulse with a known charge and look for it in the QDC spectrum.
It is VERY important to have scope pictures showing the difference between source ON/OFF. Then you use scalers to check that your trigger pulse is able to see a difference between source ON/OFF Then you do the DAQ measurements.
If you don't follow the above proceedure then you will be building a pyramid on quicksand.
The same case when the cathode voltage increased to 3.6 kV.
Then perform a set of measurements (shutter open/closed and source ON/OFF) to establish reproduceability. Make sure you record the scaler count rates.
9/17/14
Measure the charge for several values of the Cathode HV keeping the GEM preamplifier voltage and gas flow rate constant.
- CATH 3.2 KV
- CATH 3.1 KV
All the runs have the same duration 20 min.
- CATH 3.4 KV
Try to take scope picture to show difference between source on and off signal that are being measured by the QDC.
9/16/14
- Condition of above results
Shutter is Open
Red is Cf-252 source ON Run 7724
Green is Cf-252 source OFF run 7726
HV_GEM= -2930 Volts HV_Cathode=-3400 Volts
Gas Flow rate = 0.1 ft^3/hr
- Goal
- Can the Cf-252 source ON signal be changed?
Change the Cathode voltage to try and turn off the signal when the source is on.
- Result
Yes, when the cathode voltage is decreased to -3100V, QDC histogram does not show any difference in the collected charge as the shutter is open with the source on it, and when the shutter is open without the source.
The figure below shows the change in the QDC spectrum when the cathode voltage is -3.2 kV.