Experimental Setup
Attenuation Measurements Setup | ||
[[File:|400px|thumb|none|He3 tube mounting for attenuation measurements ]] | [[File: |300px|thumb|none| ]] | |
[[File:|300px|thumb|none| ]] | [[File: |300px|thumb|none| ]] |
The He3 detector is shrouded in BorAl, in a triangular configuration with +1250V to the pre-amp (this voltage is based on previous work by Jeff Burggraf and David Friant).
The thermalizing box is designed as regular poly surrounding the Cf-252 source on all sides except the side in the direction of the He3 tube.
The thermalizer is considered a regular poly sheet that can be slid in and out of the experimental setup.
The detector is ~30cm from the Cf-252 source. This allows for enough distance to insert 8" (20.32cm) of green borated poly, a 1" (2.54cm) thermalizing sheet, and still have the BorAl in the correct position.
No lead was included in the setup as it was not necessary.
The discriminator threshold was not changed during this experiment as it is in the center of the plateau work done by David Friant. See plot below.
Measurements
Background Measurements
I will be performing background runs overnight due to the low amount of background counts. This is necessary to get an acceptable level of uncertainty in the measurements as the count rate looks to be on the order of .5Hz.
The first run background run is taking place the evening of 6/25/21 and will end the following day.
Background run lasted for 60000 seconds and took place overnight. Run ended at approximately 11:45am 6/26/21
1250V Source Out Data | |||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cf-252 Button Source | Counts | Run Time (Sec) | Count Rate (Hz) | Uncertainty |
1 | Detector surrounded by BorAl at +1250V | 255 | No | 722 | 60000 | 0.0120 | 0.000448 |
Source In
No borated poly, just BorAl shielding around detector
1250V Source In Data | |||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cf-252 Source | Counts | Run Time (Sec) | Count Rate (Hz) | Uncertainty |
1 | Detector on BorAl at +1250V | 255 | Yes | 236 | 300 | ||
2 | Detector on BorAl at +1250V | 255 | Yes | 263 | 300 | ||
3 | Detector on BorAl at +1250V | 255 | Yes | 251 | 300 | ||
4 | Detector on BorAl at +1250V | 255 | Yes | 203 | 300 | ||
5 | Detector on BorAl at +1250V | 255 | Yes | 218 | 300 |
Green Poly
1" Green poly (one sheet)
1250V source in | |||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cf-252 Source | Counts | Run Time (Sec) | Count Rate (Hz) | Uncertainty |
1 | Detector on poly at +1250V | 255 | Yes | 165 | 300 | 0.55 | 0.0428 |
2 | Detector on poly at +1250V | 255 | Yes | 219 | 300 | 0.73 | 0.0493 |
3 | Detector on poly at +1250V | 255 | Yes | 197 | 300 | 0.657 | 0.0468 |
4 | Detector on poly at +1250V | 255 | Yes | 220 | 300 | 0.733 | 0.0494 |
5 | Detector on poly at +1250V | 255 | Yes | 199 | 300 | 0.663 | 0.0470 |
2" Green poly (one brick) no thermalizer
1250V source in | |||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cf-252 Source | Counts | Run Time (Sec) | Count Rate (Hz) | Uncertainty |
1 | Detector on poly at +1250V | 255 | Yes | 154 | 300 | 0.513 | 0.0414 |
2 | Detector on poly at +1250V | 255 | Yes | 163 | 300 | 0.543 | 0.0426 |
3 | Detector on poly at +1250V | 255 | Yes | 150 | 300 | 0.5 | 0.0408 |
4 | Detector on poly at +1250V | 255 | Yes | 125 | 300 | 0.417 | 0.0373 |
5 | Detector on poly at +1250V | 255 | Yes | 154 | 300 | 0.513 | 0.0414 |
3" Green poly (one brick & sheet) no thermalizer
1250V source in | |||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cf-252 Source | Counts | Run Time (Sec) | Count Rate (Hz) | Uncertainty |
1 | Detector on poly at +1250V | 255 | Yes | 300 | |||
2 | Detector on poly at +1250V | 255 | Yes | 300 | |||
3 | Detector on poly at +1250V | 255 | Yes | 300 | |||
4 | Detector on poly at +1250V | 255 | Yes | 300 | |||
5 | Detector on poly at +1250V | 255 | Yes | 300 |
4" Green poly (two bricks) no thermalizer
1250V source in | |||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cf-252 Source | Counts | Run Time (Sec) | Count Rate (Hz) | Uncertainty |
1 | Detector on poly at +1250V | 255 | Yes | 153 | 600 | 0.255 | 0.0412 |
2 | Detector on poly at +1250V | 255 | Yes | 134 | 600 | 0.223 | 0.0386 |
3 | Detector on poly at +1250V | 255 | Yes | 124 | 600 | 0.207 | 0.0371 |
4 | Detector on poly at +1250V | 255 | Yes | 145 | 600 | 0.242 | 0.0401 |
5 | Detector on poly at +1250V | 255 | Yes | 133 | 600 | 0.222 | 0.0384 |
6" Green poly (three bricks) no thermalizer
1250V source in | |||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cf-252 Source | Counts | Run Time (Sec) | Count Rate (Hz) | Uncertainty |
1 | Detector on poly at +1250V | 255 | Yes | 115 | 900 | 0.128 | 0.0119 |
2 | Detector on poly at +1250V | 255 | Yes | 137 | 900 | 0.152 | 0.0130 |
3 | Detector on poly at +1250V | 255 | Yes | 149 | 900 | 0.166 | 0.0136 |
4 | Detector on poly at +1250V | 255 | Yes | 151 | 900 | 0.168 | 0.0137 |
5 | Detector on poly at +1250V | 255 | Yes | 116 | 900 | 0.129 | 0.0120 |
8" Green poly (four bricks) no thermalizer
1250V source in | |||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cf-252 Source | Counts | Run Time (Sec) | Count Rate (Hz) | Uncertainty |
1 | Detector on poly at +1250V | 255 | Yes | 143 | 1200 | 0.119 | 0.00997 |
2 | Detector on poly at +1250V | 255 | Yes | 133 | 1200 | 0.111 | 0.00961 |
3 | Detector on poly at +1250V | 255 | Yes | 152 | 1200 | 0.127 | 0.0102 |
4 | Detector on poly at +1250V | 255 | Yes | 154 | 1200 | 0.128 | 0.0103 |
5 | Detector on poly at +1250V | 255 | Yes | 149 | 1200 | 0.124 | 0.0102 |
Regular Poly
1" Regular Poly
1250V source in | |||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cf-252 Source | Counts | Run Time (Sec) | Count Rate (Hz) | Uncertainty |
1 | Detector on poly at +1250V | 255 | Yes | 300 | |||
2 | Detector on poly at +1250V | 255 | Yes | 300 | |||
3 | Detector on poly at +1250V | 255 | Yes | 300 | |||
4 | Detector on poly at +1250V | 255 | Yes | 300 | |||
5 | Detector on poly at +1250V | 255 | Yes | 300 |
2" Regular Poly (one brick)
1250V source in | |||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cf-252 Source | Counts | Run Time (Sec) | Count Rate (Hz) | Uncertainty |
1 | Detector on poly at +1250V | 255 | Yes | 300 | |||
2 | Detector on poly at +1250V | 255 | Yes | 300 | |||
3 | Detector on poly at +1250V | 255 | Yes | 300 | |||
4 | Detector on poly at +1250V | 255 | Yes | 300 | |||
5 | Detector on poly at +1250V | 255 | Yes | 300 |
3" Regular Poly (one brick & sheet)
1250V source in | |||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cf-252 Source | Counts | Run Time (Sec) | Count Rate (Hz) | Uncertainty |
1 | Detector on poly at +1250V | 255 | Yes | 300 | |||
2 | Detector on poly at +1250V | 255 | Yes | 300 | |||
3 | Detector on poly at +1250V | 255 | Yes | 300 | |||
4 | Detector on poly at +1250V | 255 | Yes | 300 | |||
5 | Detector on poly at +1250V | 255 | Yes | 300 |
4" Regular Poly (two bricks)
1250V source in | |||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cf-252 Source | Counts | Run Time (Sec) | Count Rate (Hz) | Uncertainty |
1 | Detector on poly at +1250V | 255 | Yes | 300 | |||
2 | Detector on poly at +1250V | 255 | Yes | 300 | |||
3 | Detector on poly at +1250V | 255 | Yes | 300 | |||
4 | Detector on poly at +1250V | 255 | Yes | 300 | |||
5 | Detector on poly at +1250V | 255 | Yes | 300 |
6" Regular Poly (three bricks)
1250V source in | |||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cf-252 Source | Counts | Run Time (Sec) | Count Rate (Hz) | Uncertainty |
1 | Detector on poly at +1250V | 255 | Yes | 300 | |||
2 | Detector on poly at +1250V | 255 | Yes | 300 | |||
3 | Detector on poly at +1250V | 255 | Yes | 300 | |||
4 | Detector on poly at +1250V | 255 | Yes | 300 | |||
5 | Detector on poly at +1250V | 255 | Yes | 300 |
8" Regular Poly (four bricks)
1250V source in | |||||||
Run Number | Configuration | Discriminator Threshold (mV) | Cf-252 Source | Counts | Run Time (Sec) | Count Rate (Hz) | Uncertainty |
1 | Detector on poly at +1250V | 255 | Yes | 300 | |||
2 | Detector on poly at +1250V | 255 | Yes | 300 | |||
3 | Detector on poly at +1250V | 255 | Yes | 300 | |||
4 | Detector on poly at +1250V | 255 | Yes | 300 | |||
5 | Detector on poly at +1250V | 255 | Yes | 300 |