Difference between revisions of "A W thesis"

From New IAC Wiki
Jump to navigation Jump to search
Line 521: Line 521:
 
{| border="1"  |cellpadding="20" cellspacing="0  
 
{| border="1"  |cellpadding="20" cellspacing="0  
 
|-
 
|-
| Energy || Trig || <math>N_0</math> (before efficiency) || <math>N_0</math> (after efficiency) || <math>T_final</math> ||<math> N(t_final)</math>|| <math>\frac{N(t_final)}{N_0}</math>
+
| Energy || Trig || <math>N_0</math> (before efficiency) || <math>N_0</math> (after efficiency) || <math>T_{final}</math> ||<math> N(t_{final})</math>|| <math>\frac{N(t_{final})}{N_0}</math>
 
|-
 
|-
 
|  rowspan="2" | 898 || sing ||1.281e-8 +/- 2.316e-9 || 1.003e-8 +/- 1.813e-9 || 383.63 <math> {+199.70} \atop { -63.50}</math>||8.286e-10 +/- 4.566e-10|| 0.0826
 
|  rowspan="2" | 898 || sing ||1.281e-8 +/- 2.316e-9 || 1.003e-8 +/- 1.813e-9 || 383.63 <math> {+199.70} \atop { -63.50}</math>||8.286e-10 +/- 4.566e-10|| 0.0826

Revision as of 15:23, 13 October 2014

Y-88 source #130113-004

File:Y-88 activation log.pdf

Foil weight: 0.0496 g

Foil # of atoms: 3.398e20 atoms


Here we have: P = E [math]\times[/math] I

Here, [P] Watts, [E] MeV, and [I] [math]\mu[/math]A

Accelerator Settings
beam energy 22 MeV
Rep Rate: 170 Hz
I_peak: 80 mA/pulse
pulse width: 4 [math]\mu[/math]s

Time on for one second is 1 us. and Time off is 0.999999 s The average current 0.999999 * 0 + 0.000004*170*80mA = 54.4 [math]\mu[/math]A.

P = (22 MeV) [math]\times[/math] (54.4 [math]\mu[/math]A) = 1.2 k Watt.

Introduction

Methods of determining atomic concentration in Material

Neutron Activation Analysis (NAA)

Neutron activation analysis (NAA) is a high-sensitivity and non-destructive multi-elemental analytical technique used for analysis of major, minor, and trace elements in samples. NAA is significantly different from other spectroscopic analytical techniques in that it is based on nuclear transitions rather than electronic transitions. The basic essentials to carry out an analysis of samples by NAA are a source of neutrons to bombard the sample, instruments suitable for detecting gamma rays, and knowledge of the reactions that occur when neutrons interact with target nuclei.

The sequence of events taking place during the most common type of nuclear reaction for NAA, namely neutron capture, is illustrated in the figure. (make figure) Upon irradiation, a neuron interacts with the target nucleus via a non-elastic collision. A compound nucleus forms in an excited and usually unstable state. This unfavourable state will almost instantaneously de-excite into a more stable configuration by emitting one or more characteristic prompt gamma rays. In most cases, this new and stable configuration yields a radioactive nucleus. The newly formed radioactive nucleus now decays by the emission of one of more characteristic delayed gamma rays. This decay process is at a much slower rate according to unique half-life of the radioactive nucleus. Measurement, in principle, falls into two categories: (1) prompt gamma-ray neutron activation analysis (PGNAA), where measurements are taken during the irradiation of a sample, or (2) the more common delayed gamma-ray neutron activation analysis (DGNAA), where measurements follow radioactive decay.

A range of different neutron sources can be used for NAA. These include reactors, accelerators, fusors, and radioisotopic neutron emitters. Since they have high fluxes of neutrons from uranium fission, nuclear reactors offer the highest available sensitivities for most elements.

There are several types of detectors and configurations employed in NAA. Most are designed to detect the emitted gamma radiation. The instrumentation most commonly used consists of scintillation type or semiconductor type detector(s), associated electronics, and a computer-based, multi-channel analyser. Scintillation type detectors use a radiation-sensitive crystal, usually thallium-doped sodium iodide (NaI(Tl)). Hyperpure or intrinsic germanium (HpGe) detectors that operate at liquid nitrogen temperatures (77 degrees K) are the semiconductor type most commonly operated for NAA.

There are a few drawbacks to the use of NAA. Even though the technique is non-destructive, the irradiated sample will remain radioactive for great lengths of time depending on the half-lives, requiring handling and disposal protocols. Also, not all labs have convenient access to a suitable reactor for a neutron source to irradiate a sample. As with any other analytical method, NAA is also not universal. For instance, the determination of low-Z elements, such as C, N, O, F, or several other elements such as Mg, Si, Ca, Ti, Ni, Sr, Y, Zr, Nb, Sn, and Tl, is not sufficiently sensitive or impossible.

Inductively coupled plasma mass spectrometry(ICP-MS)

Inductively coupled plasma atomic emission spectroscopy(ICP-OES)

Atomic Absorption Spectrometry (multi-element AAS)

Particle-induced X-ray Emission (PIXE)

Another method, uses X-ray from a synchrotron light source and look at the de-excitation of atomic electrons to measure the atomic number. Reports are that they can measure pico-gram quantities.

table of detection limits -vs- Method

Coincidence Counting Setup

A_W_CAA_apparatus

Y-88 CAA

Run6980_Y88

Run7023_Y88

Run7108_Y88

Run7161_Y88

Run7204_Y88

Run7236_Y88



Run7023 Y88 scaled.png Run7023 OverlayCoin Cuts.png

Run7023 OverlayCoin Cuts log.png

Using our Y-88 source, our set-up allows us to perform runs with the detectors in coincidence (AND mode) or singles (OR mode). The figure on the left shows the HpGe detector's coincidence events (red) that occurred within a 200 ns timing window. This graph has been overlaid with the same detector’s single events (blue). We can see that by having the detectors in coincidence, the noise is reduced and several peaks, which do not have multiple photons in coincidence, are removed.

The coincidence run does not show the two energy peaks associated with Y-88 decay alone. Even though we require coincidence for the system to trigger, there is still noise. These peaks could be Compton events contaminating the coincidence and accidentals. Photons can loss a portion of their energy traveling to the detector. They do not necessarily have to deposit all of their energy into the detector either. The photon can Compton scatter out. It is not possible to tell the difference between a photon that deposited all of its energy from a scattered photon, since the speed of light is so fast and the resolution of the detector so poor. The photons can, however, be distinguished if an energy cut is applied. In order to remove the Compton events and the accidentals, we can require the photon energy to be completely deposited into the detector. This can be achieved by placing an energy cut on the NaI detector around the high energy Y-88 peak. This will remove other unwanted photon energies. The graph on the right is an overlay of the coincidence signal before (blue) and after (red) an energy cut is applied to the companion detector (NaI detector). The noise seen at low energies is greatly reduced after including the cut and improved the signal.



Background subtraction

898 keV analysis

The raw Histograms and fits may be found here A_W_thesis_old#898_keV_signal_table

The Measured half life is

Y-88 Half-life

Days [math]T_{1/2}[/math] Trig Integral Statistical Error Systematic Error Systematic Error without Coincidence Energy Cut
108 1.01 sing 1.364 [math]\pm 0.062 [/math] [math] {+0.073} \atop { -0.078}[/math] N/A
coin 0.0052 [math]\pm 2.864e-4 [/math] [math] {+6.39e-4} \atop { -1.8e-3}[/math] [math] {+1.86e-4} \atop { -1.84e-4}[/math]
159 1.49 sing 0.8697 [math]\pm 0.055 [/math] [math] {+0.057} \atop { -0.077}[/math] N/A
coin 0.0042 [math]\pm 2.263e-4 [/math] [math] {+4.43e-4} \atop { -6.45e-4}[/math] [math] {+8.3e-5} \atop { -7.82e-5}[/math]
229 2.15 sing 0.4733 [math]\pm 0.058 [/math] [math] {+0.213} \atop { -0.191}[/math] N/A
coin 0.0034 [math]\pm 2.587e-4 [/math] [math] {+5.62e-4} \atop { -1.24e-3}[/math] [math] {+1.41e-4} \atop { -2.17e-4}[/math]
245 2.30 sing 0.5807 [math]\pm 0.049 [/math] [math] {+0.07} \atop { -0.133}[/math] N/A
coin 0.0015 [math]\pm 5.454e-5 [/math] [math] {+3.68e-4} \atop { -1.53e-3}[/math] [math] {+1.27e-5} \atop { -1.47-5}[/math]



Y-88 Halflife Update.png



The signal to noise ratio is

Days [math]T_{1/2}[/math] Trig SNR Statistical Error Systematic Error
108 1.01 sing 4.368 [math]\pm 0.252 (5.77%) [/math] [math] {+0.232} \atop { -0.249}[/math]
coin 46.181 [math]\pm 4.204 (9.10%)[/math] [math] {+1.653} \atop { -1.635}[/math]
159 1.49 sing 8.228 [math]\pm 0.999 (12.14%)[/math] [math] {+0.536} \atop { -0.724}[/math]
coin 37.30 [math]\pm 1.132 (3.03%)[/math] [math] {+0.739} \atop { -0.694}[/math]
229 2.15 sing 4.478 [math]\pm 2.047 (45.71%) [/math] [math] {+2.008} \atop { -1.811}[/math]
coin 30.195 [math]\pm 2.480 (8.21%)[/math] [math] {+1.253} \atop { -1.926}[/math]
245 2.30 sing 5.494 [math]\pm 1.415 (25.75%)[/math] [math] {+0.665} \atop { -1.263}[/math]
coin 37.994 [math]\pm 4.287 (11.28%)[/math] [math] {+0.322} \atop { -0.372}[/math]

The first data point for the singles 898 keV was not included in the signal to noise fit. At the time of measurement, the noise was larger. This could be from contaminates with short half-lives in the sample that decayed away. Only elements with longer half-lives, such as Y-88, remained. The background of the remaining three data points were averaged.

For the 898 keV coincidence signal to noise fit, the last data point was not included. The length of the run and activities in the room, such as the nitrogen tank refilling, during counting caused the peak to be spread over a larger energy range. This also meant that the background noise would be larger. The background of the first three data points were averaged.


SNR BW Update.png


The following plot is the signal to noise ratio with only the data points used to obtain a fit.


SNR BW2 Update.png


[math]T_{1/2}[/math]= 130.39 +/- 26.26 (s) and 108.18 +/- 4.89 (c)

[math]N_0[/math] = 1.327e-8 +/- 2.335e-9 (s) and 1.434e-11 +/- 7.986e-13 (c)

[math]T_{sing} = 383.63 {{+199.70} \atop { -63.50} }days[/math] = the time when SNR=0 using singles

[math]T_{coinc} = 483.60 {{+129.60} \atop { -70.95} }days [/math] = the time when SNR=0 using coinc

[math]N_{sing}[/math] = 1.096e-9 +/- 6.025e-10 = lowest measurable activity using singles

[math]N_{coinc}[/math] = 6.182e-13 +/- 1.692e-13= lowest measurable activity using coinc

[math]\frac{N_{sing}}{N_0}[/math]=0.0856 +/- 0.0494

[math]\frac{N_{coinc}}{N_0}[/math]=0.0096 +/- 0.0027

1836.1 Analysis

The raw Histograms and fits may be found here A_W_thesis_old#1836_keV_signal_table

The Measured half life is

Y-88 Half-life

Days [math]T_{1/2}[/math] Trig Integral Statistical Error Systematic Error Systematic Error without Coincidence Energy Cut
108 1.01 sing 0.8414 [math]\pm 0.048 [/math] [math]\pm 0.021[/math] N/A
coin 0.0062 [math]\pm 3.13e-4 [/math] [math] {+3.91e-4} \atop { -2.64e-3}[/math] [math] {+1.14e-4} \atop { -1.11e-4}[/math]
159 1.49 sing 0.5015 [math]\pm 0.042 [/math] [math] {+0.028} \atop { -0.026}[/math] N/A
coin 0.0046 [math]\pm 2.37e-4 [/math] [math] {+4.98e-4} \atop { -2.87e-3}[/math] [math] {+6.66e-5} \atop { -7.43e-5}[/math]
229 2.15 sing 0.2822 [math]\pm 0.059[/math] [math] {+0.053} \atop { -0.056}[/math] N/A
coin 0.0062 [math]\pm 4.60e-4[/math] [math] {+1.29e-3} \atop { -6.07e-3}[/math] [math] {+2.39e-4} \atop { -3.63e-4}[/math]
245 2.30 sing 0.2449 [math]\pm 0.032 [/math] [math] {+0.028} \atop { -0.033}[/math] N/A
coin 0.0016 [math]\pm 5.63e-5 [/math] [math] {4.72e-4} \atop { -1.23e-3}[/math] [math] {+1.20e-4} \atop { -1.21e-4}[/math]



Y-88 Halflife Update.png



The signal to noise ratio is

Days [math]T_{1/2}[/math] Trig SNR Statistical Error Systematic Error
108 1.01 sing 42.07 [math]\pm 10.78 (25.62%)[/math] [math] {+1.035} \atop { -1.031}[/math]
coin 40.666 [math]\pm 1.236 (3.04%) [/math] [math] {+0.748} \atop { -0.728}[/math]
159 1.49 sing 21.52 [math]\pm 1.722 (8.23%) [/math] [math] {+1.218} \atop { -1.136}[/math]
coin 30.172 [math]\pm 2.480 (8.22%) [/math] [math] {+0.437} \atop { -0.489}[/math]
229 2.15 sing 4.64 [math]\pm 1.161 (25.03%) [/math] [math] {+0.870} \atop { -0.925}[/math]
coin 40.666 [math]\pm 3.095 (7.61%) [/math] [math] {+1.566} \atop { -2.379}[/math]
245 2.30 sing 5.43 [math]\pm 0.980 (18.05%) [/math] [math] {+0.620} \atop { -0.722}[/math]
coin 0.7569 [math]\pm 0.057 (7.57%) [/math] [math] {+0.0568} \atop { -0.0572}[/math]

For the 1836 keV coincidence signal to noise fit, the last data point was not included. The length of the run and activities in the room, such as the nitrogen tank refilling, during counting caused the peak to be spread over a larger energy range. This also meant that the background noise would be large. The background of the first three data points were averaged.


SNR BW Update.png


The following plot is the signal to noise ratio with only the data points used to obtain a fit.


SNR BW2 Update.png


[math]T_{1/2}[/math]= 106.16 +/- 8.63 (s) and 100.10 +/- 4.0 (c)

[math]N_0[/math] = 1.910e-9 +/- 2.185e-10 (s) and 1.012e-11 +/- 3.710e-13 (c)

[math]T_{sing} = 264.94 {{+4.77} \atop { -3.81} }days[/math] = the time when SNR=0 using singles

[math]T_{coinc} = 305.63 {{+66.64} \atop { -38.80} }days [/math] = the time when SNR=0 using coinc

[math]N_{sing}[/math] = 3.413e-10 +/- 3.780e-11= lowest measurable activity using singles

[math]N_{coinc}[/math] = 1.388e-12 +/- 3.068e-13= lowest measurable activity using coinc

[math]\frac{N_{sing}}{N_0}[/math]=0.0354 +/- 0.0056

[math]\frac{N_{coinc}}{N_0}[/math]=0.0170 +/- 0.0038

Error analysis

1.) rate energy cuts

2.) Energy calibration error

3.) Linear background fit error

4.) Gaussian integration of Signal - background peak

5.) ignorable error from run length time

6.) Coincidence energy cut.

7.) dividing by average background for SNR

8.) correction ratio for 3rd data point

9.) error in half life to find N0 for the detection limit extrapolation

10.) error extrapolation efficiency to higher energy lines for N0 calculation

Energy Window systematic error

The signal is measured by subtracting a count rate that is summed over an energy interval. The choice of this integration interval, energy window, is a source of systematic error. In order to calculate the error associated with the energy window, its range was adjusted based on the statistical variance of the background subtracted count rate fit to a gaussian. The integral was measured for window ranges chosen according to an integer interval of the standard deviation [math] \mu \pm n \sigma [/math], where n = 3,4,5. The difference between the largest and mid-value was taken as the positive standard deviation, and the lowest was used for the standard deviation negative fluctuation. The resulting errors were divided by the signal to obtain a percent error as shown in column 1 of TableXXX.

Energy calibration systematic error

Prior to each measurement of the Y-88 sample's activity, the detectors where calibrated using several standard sources with known activity and dominant energy lines. This was accomplished by counting certified samples of Cs-137, Co-60, and Na-22 as well as Ba-133 and Mn-54. The observed mean ADC channel number was correlated with the expected energy line from each of the above sources. The correlation was measured using a standard linear fit algorithm. The linear fit algorithm determines the uncertainty of each fit parameter, the slope and the y-intercept. The effects of this uncertainty on the signal was analysed and is referred to as the energy calibration error.

YIntCal ADC7 5.png SlopeCal ADC7 5.png

The uncertainty in the energy calibration just before a singles measurement was used to investigate the impact of the calibration on the measured signal for a given energy window of integration. While the singles measurements took less than 10 minutes, the coincidence measurement required more than 24 hours. The long term variance of the energy calibration shown in Figure ZZ.XXX was used to estimate the impact of the energy calibration drift during the coincidence measurements. The fit's y-intercept has a temporal variance corresponding to a relative uncertainty of about 1.865%; the slope is about 0.608%. The effects of this error on the coincidence signal were analysed. In both cases, the range of the energy widow is held constant while the slope and the y-intercept where adjusted to account for all possible combinations with and without their respective uncertainty. The integral value was recorded after every fit parameter change and the max, min, and mean values were obtained. The difference between the largest and mean value was taken as the positive standard deviation, and the lowest was used for the standard deviation's negative fluctuation. The resulting errors were divided by the signal to obtain a percent error as shown in column 2 of TableXXX.

Linear Fit Systematic error

Run7107 898 raw.png

The signal is determine by subtracting the underlying background. A estimate of this background is made by fitting the observed count rate with the function below :

[math] f(x)=A \textstyle \int_{\mu-2\sigma}^{\mu+2\sigma} e^{-\frac{1}{2}\left(\frac{x-\mu}{\sigma}\right)^2} \operatorname{d}\!x +B x +C[/math].

The above function contains a Gaussian representing the measured signal and a linear function for the background. The above figure XX.YY shows the fit applied. The linear fit parameters are used to create a histogram that is subtracted from the raw signal histogram, resulting in a histogram of just a signal. The energy window range used for the fit was adjusted and several fits performed to determine the uncertainty of the fit's slope and y-intercept. This systematic error, based on the fits dependence on the energy window is shown in the table under the column labeled "Linear Fit of Background". The fit parameters where adjusted to determine the maximum and minimum area of the background. For minimum background area, it was required that the area be positive. The difference between the maximum and unadjusted area value was taken as the positive standard deviation's fluctuation, and the minimum was used for the standard deviation's negative fluctuation. The resulting errors were divided by the signal to obtain a percent error as shown in column 3 of TableXXX.

Systematic error of the Coincidence Energy Cut

Figure AAA.XXX below show the coincidence measurements made before(blue) and after(red) energy cuts are applied on the NaI signal to reduce the accidental coincidence events and photons that fail to deposit all their energy in the NaI detector due to compton scattering.

Run7108 CoinOverlay.png Run7108 1836 CoinOverlay.png

To analyze the error association with coincidence counting only, the signal on the NaI detector was fit to a gaussian distribution. The [math]\sigma[/math] of the gaussian fit was used to adjust the energy cut limits. The energy window was set at [math]\text{lower/upper limit} \pm \sigma [/math]. The intergal values were recorded for each instance and the difference between the maximum and unadjusted integral values were taken as the positive standard deviation's fluctuation while the minimums were used for the standard deviation's negative fluctuation. The resulting errors were divided by the coincidence signal to obtain a percent error as shown in column 5 of TableXXX. The limited energy resolution of the NaI detector makes this uncertainty the dominant systematic error. If two HpGe detectors are used, then this uncertainty would decrease substantially will be far less than the remaining dominant systematic error; the energy Gaussian integral uncertainty. We report this systematic uncertainty but it is not propagated for the final result in order consider the case of a coincidence apparatus using two HpGe detectors.


Half-life

898 keV Singles

Days [math]T_{1/2}[/math] Trig Energy Window Energy Calibration Linear Fit of Background Gaussian Integration of Signal Total Systematic Error
108 1.01 sing [math] {+0.367%} \atop { -0.660%}[/math] [math] {+1.026%} \atop { -1.83%}[/math] [math]\pm 0.183%[/math] [math]\pm 5.132%[/math] [math] {+5.32%} \atop { -5.71%}[/math]
159 1.49 sing [math] {+0.713%} \atop { -3.68%}[/math] [math] {+0.816%} \atop { -2.438%}[/math] [math] {+0.747%} \atop { -0.724%}[/math] [math]\pm 6.29%[/math] [math] {+6.52%} \atop { -8.80%}[/math]
229 2.15 sing [math] {+32.83%} \atop { -23.83%}[/math] [math] {+11.87%} \atop { -16.35%}[/math] [math] {+2.7%} \atop { -2.58%}[/math] [math]\pm 3.68%[/math] [math] {+44.94%} \atop { -40.44%}[/math]
245 2.30 sing [math] {+1.19%} \atop { -8.28%}[/math] [math] {+4.24%} \atop { -12.0%}[/math] [math]\pm 1.72%[/math] [math]\pm 10.68%[/math] [math] {+12.1%} \atop { -22.99%}[/math]


898 keV Coincidence

Days [math]T_{1/2}[/math] Trig Energy Window Energy Calibration Linear Fit of Background Gaussian Integration of Signal Coincidence Energy Total Systematic Error Systematic Error without Coincidence Energy Cut
108 1.01 coin [math] {+0.56%} \atop { -0.404%}[/math] [math] \pm 0.31%[/math] [math] {+0.142%} \atop { -0.136%}[/math] [math]\pm 3.46%[/math] [math] {+10.92%} \atop { -33.65%}[/math] [math] {+12.29%} \atop { -34.62%}[/math] [math] {+3.58%} \atop { -3.54%}[/math]
159 1.49 coin [math] {+0.405%} \atop { -0.048%}[/math] [math] \pm 0.38%[/math] [math] {+0.048%} \atop { -0.047%}[/math] [math]\pm 1.81%[/math] [math] {+9.6%} \atop { -14.8%}[/math] [math] {+10.55%} \atop { -15.36%}[/math] [math] {+1.98%} \atop { -1.86%}[/math]
229 2.25 coin [math] {+0.588%} \atop { -3.12%}[/math] [math] {+1.03%} \atop { -2.0%}[/math] [math] {+0.237%} \atop { -0.231%}[/math] [math]\pm 3.82%[/math] [math] {+14.5%} \atop { -31.2%}[/math] [math] {+16.53%} \atop { -36.47%}[/math] [math] {+4.15%} \atop { -6.38%}[/math]
245 2.30 coin [math] {+0.2%} \atop { -0.067%}[/math] [math] {+0.13%} \atop { -0.53%}[/math] [math] {+0.296%} \atop { -0.291%}[/math] [math]\pm 0.719%[/math] [math] {+24.2%} \atop { -101.3%}[/math] [math] {+24.53%} \atop { -102.0%}[/math] [math] {+0.847%} \atop { -0.98%}[/math]


1836.1 keV Singles

Days [math]T_{1/2}[/math] Trig Energy Window Energy Calibration Linear Fit of Background Gaussian Integration of Signal Total Systematic Error
108 1.01 sing [math] {+0.119%} \atop { -0.095%}[/math] [math]\pm 0.33%[/math] [math] {+0.606%} \atop { -0.582%}[/math] [math]\pm 2.34%[/math] [math] {+2.46%} \atop { -2.45%}[/math]
159 1.49 sing [math] {+0.199%} \atop { -0.798%}[/math] [math] {+2.81%} \atop { -1.42%}[/math] [math]\pm 0.279%[/math] [math]\pm 4.79%[/math] [math] {+5.66%} \atop { -5.28%}[/math]
229 2.15 sing [math] {+10.21%} \atop { -7.48%}[/math] [math] {+4.32%} \atop { -8.65%}[/math] [math] {+3.26%} \atop { -3.01%}[/math] [math] {+11.37%} \atop { -11.34%}[/math] [math] {+18.74%} \atop { -19.94%}[/math]
245 2.30 sing [math] {+0.245%} \atop { -5.31%}[/math] [math]\pm 1.674%[/math] [math] {+2.246%} \atop { -2.205%}[/math] [math] {+11.025%} \atop { -11.107%}[/math] [math] {+11.414%} \atop { -13.303%}[/math]


1836.1 keV Coincidence

Days [math]T_{1/2}[/math] Trig Energy Window Energy Calibration Linear Fit of Background Gaussian Integration of Signal Coincidence Energy Cut Total Systematic Error Systematic Error without Coincidence Energy Cut
108 1.01 coin [math] {+0.613%} \atop { -0.55%}[/math] [math] {+0.52%} \atop { -0.5%}[/math] [math]\pm 0.04%[/math] [math]\pm 1.45%[/math] [math] {+5.02%} \atop { -41.45%}[/math] [math] {+6.31%} \atop { -42.58%}[/math] [math] {+1.84%} \atop { -1.79%}[/math]
159 1.49 coin [math] {+0.435%} \atop { -0.196%}[/math] [math] {+0.261%} \atop { -0.804%}[/math] [math] {+0.074%} \atop { -0.073% }[/math] [math]\pm 1.267%[/math] [math] {+10.07%} \atop { -61.3%}[/math] [math] {+10.83%} \atop { -62.39%}[/math] [math] {+1.45%} \atop { -1.62%}[/math]
229 2.15 coin [math] {+0.629%} \atop { -2.55%}[/math] [math] {+1.11%} \atop { -2.19%}[/math] [math] {+0.244%} \atop { -0.239%}[/math] [math]\pm 3.44%[/math] [math] {+18.71%} \atop { -93.06%}[/math] [math] {+20.81%} \atop { -97.90%}[/math] [math] {+3.85%} \atop { -5.85%}[/math]
245 2.30 coin [math] {+0.125%} \atop { -0.438%}[/math] [math] \pm 0.188%[/math] [math]\pm 0.386% [/math] [math]\pm 7.5% [/math] [math] {+28.2%} \atop { -75.6%}[/math] [math] {29.5%} \atop { -76.88%}[/math] [math] {+7.5%} \atop { -7.56%}[/math]

Systematic error in the SNR

The singles measurement of the 898 keV peak had an initial high background that decays quickly and was more or less constant for singles measurements take 159 days after irradiation. We assumed that this reflected the presence of activated material with half lives longer that 245 days. As a result the background for the last three measurements was averaged and used in the denominator for the SNR. The first singles measurement has a large background due to activated nuclei with short half lives. This background decreased substantially after about 51 days when the next measurement was taken (the half-life of Y-88 is 106.6 days). The singles background near the 1836 keV line was measured to be around 0.5 Hz and fluctuated less over the experiment compared to the background around the 898 peak. The coincidence measurements had such little background, due to the nature of the apparatus, that the background was averaged over the first three measurements for both the 898 and the 1836 peaks separately. The background for the final coincidence measurements was ignored because of the increased noise produced when the detectors nitrogen dewar was filled during the 5 day background measurement.

Background Update.png

The systematic uncertainty of the background was not analysed and propagated into the final SNR uncertainty. The signal's total systematic error, described above, was used for the final uncertainty.


Signal to Noise Ratio

Energy (keV) Trig Variance of Averaged Background
898 sing [math]\pm 39.16%[/math]
coin [math]\pm 50.48%[/math]
1836.1 sing N/A
coin [math]\pm 153.50%[/math]

The background's systematic error is being ignored.

Background contribution to the SNR error

The analysis of the background uncertainty is described in the above section, "Linear Fit Systematic error." The background's relative statistical uncertainty shown in column 5 of TableXXX is calculated by varying the integral of the background by the uncertainties from the fit and dividing by the background used for the SNR ratio calculation. Column 6 presents the uncertainty of the averaged background as mentioned in the previous section. The variance the coincidence measurement is considerable larger than the individual coincidence measurements. The background in each singles measurement is smaller than the average variance indicating that the individual background measurements could have been made longer to improve their statistical uncertainty.

898 keV Background

Days [math]T_{1/2}[/math] Trig Background Linear Fit of Background Error Statistical Error of Averaged Background
108 1.01 sing 0.3123 [math]\pm 0.801%[/math] N/A
coin 9.518e-5 [math] {+7.73%} \atop { -7.45%}[/math] [math]\pm 50.01%[/math]
159 1.49 sing 0.0777 [math] {+8.37%} \atop { -8.11%}[/math] [math]\pm 5.9%[/math]
coin 8.732e-5 [math] {+2.30%} \atop { -2.28%}[/math] [math]\pm 54.51%[/math]
229 2.15 sing 0.1532 [math] {+8.36%} \atop { -7.96%}[/math] [math]\pm 3.0%[/math]
coin 1.554e-4 [math] {+5.17%} \atop { -5.06%}[/math] [math]\pm 30.63%[/math]
245 2.30 sing 0.0862 [math]\pm 11.60%[/math] [math]\pm 5.3%[/math]
coin 3.948e-5 [math] {+11.24%} \atop { -11.06%}[/math] [math]\pm 120.57%[/math]

1836 keV Background

Days [math]T_{1/2}[/math] Trig Background Linear Fit of Background Error Statistical Error
108 1.01 coin 1.037e-4 [math] {+2.42%} \atop { -2.39% }[/math] [math]\pm 951.78%[/math]
159 1.49 coin 4.248e-5 [math] {+8.06%} \atop { -7.88% }[/math] [math]\pm 232.34%[/math]
229 2.15 coin 3.112e-4 [math] {+4.87%} \atop { -4.76%}[/math] [math]\pm 317.16%[/math]
245 2.30 coin 2.114e-3 [math]\pm 0.29% [/math] [math]\pm 46.69%[/math]


Singles Efficiency on 2/28/2014

A measurement of the HpGe detectors geometric efficiency for detecting the Y-88 energy lines at 898 keV and 1836.1 keV was performed in order to determine the absolute activity of the sample. The count rate for several sources of known activity was measured by the data acquisition system. Since each sample has decayed since the original calibration by the source vendor, column 4 of tableXX.YY. reports the expected rates from each calibrated sample that were calculated using the half life and the known activity of the standard sources. The characteristic energy peaks from each source was fit to a gaussian distribution. The variance from that fit was used to determine the range of integration for the count rate measurement. The integration energy range was selected to include counts over the energy interval from [math]\bar{E} - (2\sigma+\Delta \sigma)[/math] to [math]\bar{E} + (2\sigma+\Delta \sigma)[/math]. The measured rates for each calibrated source using the HpGe detector are listed in column 5. The efficiency given by the ratio of the rates in column 5 to the expected rates in column 4 are shown in column 6 of the same table. The percent efficiency of the standard sources was fit to a exponential curve shown in figureXX.YY. The singles efficiency for 898 keV and 1836.1 keV were extrapolated from the exponential equation and are recorded in tableXX.YY. This process was repeated for the NaI detector and the results can be seen in figureXX.YY and tablesXX.YY and XX.YY.


HpGe Detector

Run Source Energy Expected rate (Hz) HpGe Det B Rate (ADC 7) (Hz) HpGe Eff (%)
7027 Na-22 511 1589.218 +/- 49.425 18.07 +/- 0.111 1.137 +/- 0.0317
7025 Cs-137 661.657 2548.03 +/- 80.008 25.468+/- 0.172 0.9995 +/- 0.0321
7029 Mn-54 834.848 41.405 +/- 1.300 0.485 +/- 0.014 1.171 +/- 0.0427
7026 Co-60 1173.228 1876.619 +/- 58.926 11.03 +/- 0.106 0.588 +/- 0.0328
7027 Na-22 1274.537 888.264 +/- 27.625 4.857 +/- 0.034 0.5468 +/- 0.0319
7026 Co-60 1332.492 1878.167 +/- 58.974 9.584 +/- 0.023 0.510 +/- 0.0315


  1  Constant     6.36491e-01   4.70444e-02   1.00271e-05  -3.28621e-03
  2  Slope       -9.81232e-04   5.68236e-05   1.21109e-08  -1.59453e+00


SingEff NoMn.png


Extrapolated Efficiency

Source Energy HpGe Eff (%)
Y-88 898 0.7830 +/- 0.0471
Y-88 1836.1 0.3119 +/- 0.0310



Energy Trig [math]N_0[/math] (before efficiency) [math]N_0[/math] (after efficiency) [math]T_{final}[/math] [math] N(t_{final})[/math] [math]\frac{N(t_{final})}{N_0}[/math]
898 sing 1.281e-8 +/- 2.316e-9 1.003e-8 +/- 1.813e-9 383.63 [math] {+199.70} \atop { -63.50}[/math] 8.286e-10 +/- 4.566e-10 0.0826
coin 6.411e-11 +/- 4.449e-12 1.084e-11 +/- 7.519e-13 483.60 [math] {+129.60} \atop { -70.95}[/math] 4.673e-13 +/- 1.294e-13 0.0431
1836.1 sing 9.634e-9 +/- 1.129e-9 3.005e-9 +/- 3.521e-10 264.94 [math] {+4.77} \atop { -3.81}[/math] 5.369e-10 +/- 6.099e-11 0.1787
coin 8.147e-11 +/- 3.633e-12 1.591e-11 +/- 7.095e-13 305.63 [math] {+66.64} \atop { -38.80}[/math] 2.182e-12 +/- 4.855e-13 0.1371

NaI Detector

Run Source Energy Expected rate (Hz) NaI Det B Rate (ADC 7) (Hz) NaI Eff (%)
7027 Na-22 511 1589.218 +/- 49.425 14.393 +/- 0.715 0.9057 +/- 0.0586
7025 Cs-137 661.657 2548.03 +/- 80.008 22.8 +/- 0.9154 0.8948 +/- 0.0510
7026 Co-60 1173.228 1876.619 +/- 58.926 7.529 +/- 0.5503 0.4012 +/- 0.0796
7027 Na-22 1274.537 888.264 +/- 27.625 3.654 +/- 0.5266 0.4114 +/- 0.1474
7026 Co-60 1332.492 1878.167 +/- 58.974 6.784 +/- 0.5345 0.3612 +/- 0.0848


SingEffNaI corrected.png


Source Energy NaI Eff (%)
Y-88 898 0.6262 +/- 0.0246
Y-88 1836.1 0.2158 +/- 0.0466

Coincidence Efficiency

Unlike the singles efficiency found above, the efficiency for a coincidence measurement relies on the combined efficiency of both detectors. A known Y-88 source was not available to measure the coincidence activity directly. Instead, the HpGe's geometric efficiency for detecting the Y-88 energy lines at 898 keV and 1836.1 keV was combined with the NaI detectors efficiency for the 1836.1 and 898 keV lines respectively. For example, to obtain the 898 keV coincidence efficiency, the HpGe singles efficiency at this energy line is multiplied by the NaI singles efficiency at the 1836.1 keV line. The uncertainties of the singles efficiency are added in quadrature and the resulting values recorded in column 4 of tableXX.YY. The reverse case was applied to find the coincidence efficiency at the 1836.1 keV line and the values are listed in column 4 of tableXX.YY.


Source HpGe Singles 898 keV Eff (%) NaI Singles 1836.1 keV Eff (%) HpGe Coincidence Eff (%)
Y-88 0.7830 +/- 0.0471 0.2158 +/- 0.0466 0.1690 +/- 0.0663


Source HpGe Singles 1836.1 keV Eff (%) NaI Singles 898 keV Eff (%) HpGe Coincidence Eff (%)
Y-88 0.3119 +/- 0.0310 0.6262 +/- 0.0246 0.1953 +/- 0.0396

The SNR measurement

Activation efficiency for Y-88

It is well understood that radioactive materials decay as an exponential function of time. We can therefore use the measured signals of the Y-88 sample fitted to an exponential curve to find the initial number of atoms in the foil. The exponential fit was extrapolated back to the initial time t=0 when the foil was activated. This value is in terms of rate (Hz) and we want to convert the units to Curie. The half-life of Y-88 is well know, 106.63 days, and was used to calculate the decay constant [math]\lambda[/math]. Using the the relationship between activity and the number density [math]A=\lambda N[/math] we can find [math]N_0[/math] for t=0.

The Y-88 foil's mass was weighted at the IAC and using Avogadro's number and the isotopic mass, it is calculated that the foil contains [math]3.398\times 10^{20}[/math] atoms of Y-88.

m=0.0496 g

[math]N_a=6.022 \times 10^{23} mol^{-1}[/math]

[math]m_i=87.9095011 \frac{g}{mol}[/math]

[math]Number of atoms = \frac{mN_a}{m_i}[/math]


Taking to ratio [math]\frac{Number of atoms}{N_0}[/math] will tell us how well the accelerator beam was at activating to sample.

Determining the lowest measurable activity

We want to know what is the lowest amount of measurable activity that can be detected with our setup. Using the signal to noise ratio fit, as seen in figureXX.YY, we can find the the time at which the SNR is zero. Beyond this point, we will not be able to distinguish the signal from the background noise. This final time can be plugged into the exponential decay function [math]N(t)=N_0e^{-\lambda t}[/math] to find the lowest number density, [math] N_{final} [/math], and therefore the lowest measurable activity.

Ba-133 CAA

BO-08-22-13

Useful commands

Converting CODA data file to ROOT

make sure the CODA and ROOT environmental variables are setup by source the following scripts

source ~/CODA/setup
source ~/ROOT/root/bin/thisroot.csh

Now change to the data subdirectory and execute the program to convert the data file to root

cd /data
~/CODA/CODAreader/ROOT_V5.30/V785V792/evio2nt -fr6994.dat > /dev/null

rename the file so it has the .root extension allowing ROOT to identify it in the browser

mv r6994 r6994.root


Calibration work

AW_ADC_7_Histogram

SlopeCal ADC7 5.png

ADC7 Average Slope = 0.601839 +/- 0.003662033

YIntCal ADC7 5.png

ADC7 Average Y-intercept = -51.1025 +/- 0.953309

AW_ADC_3_Histogram

SlopeCal ADC3 5.png

ADC3 Average Slope = 1.39597 +/- 0.029539420

YIntCal ADC3 5.png

ADC3 Average Y-Intercept = -133.007 +/- 2.66636

System's intrinsic err

Plot calibration parameters as a function of time

determine the variance of the parameters using several (>20) fits

Impact of higher order fits

Plot [math]E_{expected} - E_{fit} -vs- E_{expected}[/math]

Variance comes from several fits,.

Compare uncertainty when fit is E-vs-Channe to Channel-vs-E

Probably should use %error for the weighting

Concentration measurement

A comparison of the measure concentrations using singles and coincidence counting


PAA_Research

A_W_thesis_old