Difference between revisions of "Y88-08-22-13"

From New IAC Wiki
Jump to navigation Jump to search
Line 41: Line 41:
 
|5825 ||  Aug 23 (1377299361) || 1143  || || || Y-88 130100-010, only counting with detector 180-3 (no TDC) 19 min 7 sec long,  500 Hz
 
|5825 ||  Aug 23 (1377299361) || 1143  || || || Y-88 130100-010, only counting with detector 180-3 (no TDC) 19 min 7 sec long,  500 Hz
 
|-
 
|-
|5901 || Sept 3 (1378246063)  || 1201|| Y-88 130100-010, only counting with detector 180-3 20 min 6 sec long,  184 Hz
+
|5901 || Sept 3 (1378246063)  || 1201||  || ||Y-88 130100-010, only counting with detector 180-3 20 min 6 sec long,  184 Hz
 
|-
 
|-
|5905 || Sept 3 (1378249601) ||923 || Y-88 130100-010 changed post amp 8.92 to 4.62 , only counting with detector 180-3 15 min 28 sec long,  70 Hz
+
|5905 || Sept 3 (1378249601) ||923 ||  || ||Y-88 130100-010 changed post amp 8.92 to 4.62 , only counting with detector 180-3 15 min 28 sec long,  70 Hz
 
|-
 
|-
|6022 ||  9/16 (1379343487)|| 298|| Y-88  1 uCi 1219-33-5, Using only Det B ,  4 min  35sec long,  219  Hz  
+
|6022 ||  9/16 (1379343487)|| 298||  || ||Y-88  1 uCi 1219-33-5, Using only Det B ,  4 min  35sec long,  219  Hz  
 
|-
 
|-
|6023 || 9/16 (1379343834)|| 178|| Y-88  1 uCi 1219-33-5, Using only Det B ,  3  min 3 sec long,    Hz ,Moved source up 2" closer, no difference, I think I am just measureing background with this source
+
|6023 || 9/16 (1379343834)|| 178||   || ||Y-88  1 uCi 1219-33-5, Using only Det B ,  3  min 3 sec long,    Hz ,Moved source up 2" closer, no difference, I think I am just measureing background with this source
 
|-
 
|-
|6025 ||  9/16(1379344382)|| 214|| Y-88  1 uCi 130100-010, Using only Det B ,  3  min 4 sec long,  223  Hz , not much above background here
+
|6025 ||  9/16(1379344382)|| 214||  || ||Y-88  1 uCi 130100-010, Using only Det B ,  3  min 4 sec long,  223  Hz , not much above background here
 
|-
 
|-
|6029 || 9/16(1379346383) || 980|| Y-88  1 uCi 130100-010, Using only Det B , '''increase CFD threshold to 49 mv (~430 keV) ''', 17 min 28 sec long,    Hz , The shielded HpGe detector saw both Y-88 lines and had a FWHM width of 3 keV on the 898 line I get a FWHM = 3.44 keV
+
|6029 || 9/16(1379346383) || 980||  || ||Y-88  1 uCi 130100-010, Using only Det B , '''increase CFD threshold to 49 mv (~430 keV) ''', 17 min 28 sec long,    Hz , The shielded HpGe detector saw both Y-88 lines and had a FWHM width of 3 keV on the 898 line I get a FWHM = 3.44 keV
 
|-
 
|-
|6562 || 10/31(1383237042) || 760|| Y-88  1 uCi 130100-010, Using only Det B ,  
+
|6562 || 10/31(1383237042) || 760||  || ||Y-88  1 uCi 130100-010, Using only Det B ,  
 
|-
 
|-
  

Revision as of 22:50, 4 November 2013

run 6063 lines : 898 keV, 1836 seen run 6064 background lines are close to the above lines


Energy (keV) Relative Intensity
850.6 0.065
898.042 93.7
1382.2 0.021
1836.063 99.2
2734.0 0.71
3219.7 0.0070


E(γ) Coincidence

  850.6	898.042, 1836.063, 2734.0
  850.6	898.042, 1836.063, 2734.0
  898.042	850.6, 850.6, 1836.063
  898.042	1836.063
 1382.2	1836.063
 1382.2	1836.063
 1836.063	850.6, 850.6, 898.042, 898.042, 1382.2, 1382.2
 2734.0	850.6, 850.6

Runs

Run Number Date (Seconds since 1970) Running Time (sec) 898 Rate 1898 rate Calibration Desc
5825 Aug 23 (1377299361) 1143 Y-88 130100-010, only counting with detector 180-3 (no TDC) 19 min 7 sec long, 500 Hz
5901 Sept 3 (1378246063) 1201 Y-88 130100-010, only counting with detector 180-3 20 min 6 sec long, 184 Hz
5905 Sept 3 (1378249601) 923 Y-88 130100-010 changed post amp 8.92 to 4.62 , only counting with detector 180-3 15 min 28 sec long, 70 Hz
6022 9/16 (1379343487) 298 Y-88 1 uCi 1219-33-5, Using only Det B , 4 min 35sec long, 219 Hz
6023 9/16 (1379343834) 178 Y-88 1 uCi 1219-33-5, Using only Det B , 3 min 3 sec long, Hz ,Moved source up 2" closer, no difference, I think I am just measureing background with this source
6025 9/16(1379344382) 214 Y-88 1 uCi 130100-010, Using only Det B , 3 min 4 sec long, 223 Hz , not much above background here
6029 9/16(1379346383) 980 Y-88 1 uCi 130100-010, Using only Det B , increase CFD threshold to 49 mv (~430 keV) , 17 min 28 sec long, Hz , The shielded HpGe detector saw both Y-88 lines and had a FWHM width of 3 keV on the 898 line I get a FWHM = 3.44 keV
6562 10/31(1383237042) 760 Y-88 1 uCi 130100-010, Using only Det B ,

run 5825

On 8/23 at measured singles rate of Y-88 foils that was irradiated on 8/22/13

R5825 ADC5 raw.png

run 6063

calibrated draw command

ntuple->Draw("(-49.0214+ADC9*1.10056-ADC9*ADC9*2.70094e-05)>>(4096,0,4096)")


PAA_8-22-13