Difference between revisions of "Y88-08-22-13"
Jump to navigation
Jump to search
(→Runs) |
|||
| Line 39: | Line 39: | ||
| Run Number || Date || Running Time || Desc | | Run Number || Date || Running Time || Desc | ||
|- | |- | ||
| − | |5825 || Y-88 130100-010, only counting with detector 180-3 (no TDC) 19 min 7 sec long, 500 Hz | + | |5825 || || || Y-88 130100-010, only counting with detector 180-3 (no TDC) 19 min 7 sec long, 500 Hz |
|- | |- | ||
| − | |5901 || Y-88 130100-010, only counting with detector 180-3 20 min 6 sec long, 184 Hz | + | |5901 || || || Y-88 130100-010, only counting with detector 180-3 20 min 6 sec long, 184 Hz |
|- | |- | ||
| − | |5905 || Y-88 130100-010 changed post amp 8.92 to 4.62 , only counting with detector 180-3 15 min 28 sec long, 70 Hz | + | |5905 || || || Y-88 130100-010 changed post amp 8.92 to 4.62 , only counting with detector 180-3 15 min 28 sec long, 70 Hz |
|- | |- | ||
| − | |6022 || Y-88 1 uCi 1219-33-5, Using only Det B , 4 min 35sec long, 219 Hz | + | |6022 || 9/16 || || Y-88 1 uCi 1219-33-5, Using only Det B , 4 min 35sec long, 219 Hz |
|- | |- | ||
| − | |6023 || Y-88 1 uCi 1219-33-5, Using only Det B , 3 min 3 sec long, Hz ,Moved source up 2" closer, no difference, I think I am just measureing background with this source | + | |6023 || 9/16 || || Y-88 1 uCi 1219-33-5, Using only Det B , 3 min 3 sec long, Hz ,Moved source up 2" closer, no difference, I think I am just measureing background with this source |
|- | |- | ||
| − | |6025 || Y-88 1 uCi 130100-010, Using only Det B , 3 min 4 sec long, 223 Hz , not much above background here | + | |6025 || 9/15|| || Y-88 1 uCi 130100-010, Using only Det B , 3 min 4 sec long, 223 Hz , not much above background here |
|- | |- | ||
| − | |6029 || Y-88 1 uCi 130100-010, Using only Det B , '''increase CFD threshold to 49 mv (~430 keV) ''', 17 min 28 sec long, Hz , The shielded HpGe detector saw both Y-88 lines and had a FWHM width of 3 keV on the 898 line I get a FWHM = 3.44 keV | + | |6029 || 9/16 || || Y-88 1 uCi 130100-010, Using only Det B , '''increase CFD threshold to 49 mv (~430 keV) ''', 17 min 28 sec long, Hz , The shielded HpGe detector saw both Y-88 lines and had a FWHM width of 3 keV on the 898 line I get a FWHM = 3.44 keV |
|- | |- | ||
Revision as of 20:24, 1 November 2013
run 6063 lines : 898 keV, 1836 seen run 6064 background lines are close to the above lines
| Energy (keV) | Relative Intensity |
| 850.6 | 0.065 |
| 898.042 | 93.7 |
| 1382.2 | 0.021 |
| 1836.063 | 99.2 |
| 2734.0 | 0.71 |
| 3219.7 | 0.0070 |
E(γ) Coincidence
850.6 898.042, 1836.063, 2734.0 850.6 898.042, 1836.063, 2734.0 898.042 850.6, 850.6, 1836.063 898.042 1836.063 1382.2 1836.063 1382.2 1836.063 1836.063 850.6, 850.6, 898.042, 898.042, 1382.2, 1382.2 2734.0 850.6, 850.6
Runs
| Run Number | Date | Running Time | Desc |
| 5825 | Y-88 130100-010, only counting with detector 180-3 (no TDC) 19 min 7 sec long, 500 Hz | ||
| 5901 | Y-88 130100-010, only counting with detector 180-3 20 min 6 sec long, 184 Hz | ||
| 5905 | Y-88 130100-010 changed post amp 8.92 to 4.62 , only counting with detector 180-3 15 min 28 sec long, 70 Hz | ||
| 6022 | 9/16 | Y-88 1 uCi 1219-33-5, Using only Det B , 4 min 35sec long, 219 Hz | |
| 6023 | 9/16 | Y-88 1 uCi 1219-33-5, Using only Det B , 3 min 3 sec long, Hz ,Moved source up 2" closer, no difference, I think I am just measureing background with this source | |
| 6025 | 9/15 | Y-88 1 uCi 130100-010, Using only Det B , 3 min 4 sec long, 223 Hz , not much above background here | |
| 6029 | 9/16 | Y-88 1 uCi 130100-010, Using only Det B , increase CFD threshold to 49 mv (~430 keV) , 17 min 28 sec long, Hz , The shielded HpGe detector saw both Y-88 lines and had a FWHM width of 3 keV on the 898 line I get a FWHM = 3.44 keV |
run 5828
Measured singles rate of Y-88 foils that was irradiated on 8/22/13
run 6063
calibrated draw command
ntuple->Draw("(-49.0214+ADC9*1.10056-ADC9*ADC9*2.70094e-05)>>(4096,0,4096)")