Difference between revisions of "05/12/2011 random coincidence cosmic test"
Line 1: | Line 1: | ||
[https://wiki.iac.isu.edu/index.php/N%27s_detector_calibration Go Back] | [https://wiki.iac.isu.edu/index.php/N%27s_detector_calibration Go Back] | ||
+ | The voltages and thresholds are: | ||
+ | {| border="1" cellspacing="0" style="text-align: center; width: 800px; height: 100px;" | ||
+ | ! scope="col" width="50" | | ||
+ | ! scope="col" width="50" | PMT voltage, V | ||
+ | ! scope="col" width="50" | threshold, mV | ||
+ | ! scope="col" width="50" | discr. width, ns | ||
+ | |- | ||
+ | !PS1 | ||
+ | | -1000|| -80|| 20 | ||
+ | |- | ||
+ | !PS1 | ||
+ | | -1000|| -80|| 10 | ||
+ | |} | ||
The powers are: | The powers are: |
Revision as of 04:47, 14 May 2011
The voltages and thresholds are:
PMT voltage, V | threshold, mV | discr. width, ns | |
---|---|---|---|
PS1 | -1000 | -80 | 20 |
PS1 | -1000 | -80 | 10 |
The powers are:
PS1 and PS2: -1000 V (the old 2-channel TENNELEC TC952) n's: -1100 V (the new 2-channel C.A.E.N N471)
The thresholds are:
PS1 and PS2: -80 mV n's: -200 mV
The discriminator widths are:
PS1: -20 ns PS2: -10 ns n's: -20 ns
Let's do the random coincidence cosmic ray test 1.
PS1 and PS2 is about 30 cm away from each other. And let's look on the coincidence signal between PS1 and PS2.
- 70 min test was on May 12 2011
- 80 min test was on May 13 2011
total counts | time | counts/sec | |
---|---|---|---|
PS1 | 70 min | ||
PS2 | 70 min | ||
PS1+PS2 (30 cm away) | 70 min | ||
PS1 | 80 min | ||
PS2 | 80 min | ||
PS1+PS2 (30 cm away) | 80 min |
And below my scope pictures of this rear event:
Here the questions: All coincidences are true and caused by detectors or maybe some of them are false and caused by electronics (Tamuna said I need to subtract 3-4 coincidences from total 11 for 80 min test because she did something with electronics at that time)
Let's do the random coincidence cosmic ray test 2.
But now the distance between PS1 and PS2 is about 5 cm. And let's look on the coincidence signal between PS1 and PS2:
total counts | time | counts/sec | |
---|---|---|---|
PS1 | 60 min | ||
PS1 | 60 min | ||
PS1+PS2 (5 cm away) | 60 min |
And below my scope pictures of this rear event as before:
So what we are measured?
- ○ purely random events from different reaction?. But why does the coincidence rate depend from distance between PS1 and PS2? Statistical fluctuation? Electronic false events?
- ○ or probably some coincidence events caused by air shower originated from one particle nearby the two detectors. Now if detectors located closer to each other the rate should be higher because the detectors are closer to initial particle. As evidence note that rates for PS1 and PS1 for different distances are almost the same. But the coincidence rate PS1+PS2 increased almost in 3 times as distance reduced from 30 cm to 5 cm. (That need to be studied)
Also Sadiq did something at that time with gamma source. Maybe that effected somehow my measurements.