Difference between revisions of "BriansRanLogbook"
Jump to navigation
Jump to search
Line 60: | Line 60: | ||
| 1.503 || 5.4 ± 2.22 || 105.7 ± 8.63 | | 1.503 || 5.4 ± 2.22 || 105.7 ± 8.63 | ||
|- | |- | ||
− | | 1.227 || 7.9 ± 3.14 || 109.8 | + | | 1.227 || 7.9 ± 3.14 || 109.8 ± 9.34 |
|- | |- | ||
| 1.009 || 89.9 ± 7.87 || 234.1 ± 20.06 | | 1.009 || 89.9 ± 7.87 || 234.1 ± 20.06 | ||
Line 71: | Line 71: | ||
2/10/23 | 2/10/23 | ||
Same procedure as 2/3/23, just changing HV applied to He3 detector to values of 1300V then 1100V. | Same procedure as 2/3/23, just changing HV applied to He3 detector to values of 1300V then 1100V. | ||
+ | |||
+ | |||
+ | {| class="wikitable" | ||
+ | |+ He3 Counts vs Voltage (1100 HV) | ||
+ | |- | ||
+ | ! Discriminator Voltage (V) !! Counts/10s (with source) !! Counts/10s (no source) | ||
+ | |- | ||
+ | | 1.971 || 4.6 ± 2.37 || 107.7 ± 13.17 | ||
+ | |- | ||
+ | | 1.775 || 7.7 ± 2.11 || 100 ± 10.70 | ||
+ | |- | ||
+ | | 1.486 || 5.5 ± 2.32 || 104.8 ± 12.79 | ||
+ | |- | ||
+ | | 1.248 || 4.7 ± 2.00 || 99.6 ± 7.09 | ||
+ | |- | ||
+ | | 1.013 || 75.2 ± 12.17 || 171 ± 12.82 | ||
+ | |- | ||
+ | | 0.762 || 3074 ± 143.0 || 2833 ± 894.7 | ||
+ | |- | ||
+ | | 0.509 || 19658 ± 424.6 || 40290 ± 2402 | ||
+ | |} | ||
+ | |||
+ | {| class="wikitable" | ||
+ | |+ He3 Counts vs Voltage (1300 HV) | ||
+ | |- | ||
+ | ! Discriminator Voltage (V) !! Counts/10s (with source) !! Counts/10s (no source) | ||
+ | |- | ||
+ | | 1.971 || 5.4 ± 1.78 || 111.3 ± 9.97 | ||
+ | |- | ||
+ | | 1.773 || 5.3 ± 1.89 || 97.3 ± 7.70 | ||
+ | |- | ||
+ | | 1.524 || 7.1 ± 4.51 || 99.7 ± 11.78 | ||
+ | |- | ||
+ | | 1.275 || 12.2 ± 3.55 || 117.3 ± 23.11 | ||
+ | |- | ||
+ | | 0.993 || 177.7 ± 8.18 || 351.9 ± 28.71 | ||
+ | |- | ||
+ | | 0.784 || 2691.2 ± 114.7 || 2763.7 ± 126.2 | ||
+ | |- | ||
+ | | 0.485 || 50346 ± 3831 || 53920 ± 4354 | ||
+ | |} |
Revision as of 19:54, 20 February 2023
This is where I write down what I'm doing
1/13/23 : Turned on HpGe HV to 1500 and learned about analog to digital signal processing
detected signal rate -vs- HpGe Voltage
Cs-137 source on face of HpGe detector
HV | Counts |
1492.67 | 1532.5 +/- 36.67 |
799.15 | 1336.5 +/- 2.121 |
396.83 | 1193.5 +/- 54.477 |
197.19 | 1035.5 +/- 40.305 |
4.88 | 562.5 +/- 2.121 |
NO source on face of HpGe detector
HV | Counts |
1492.67 | 167.5 +/- 7.778 |
799.15 | 149.4 +/- 34.648 |
396.83 | 127.5 +/- 7.778 |
197.19 | 121 +/- 4.243 |
4.88 | 332 +/- 35.355 |
HV on He3: 1208V
Amplifier Gain: 21.5
He3 Tube
1/27/23 : Turned on He3 detector, adjusted to find counts with Cf source in containment and out of containment
2/3/23 : Same as 1/27, adjusting discriminator voltage and recording counts with Cf-252 source. Encountered strange effects of higher counts with source away at low QD voltage. Sets of measurements notably inconsistent.
Discriminator Voltage (V) | Counts/10s (with source) | Counts/10s (no source) |
---|---|---|
2.003 | 4.4 ± 1.67 | 97.0 ± 8.49 |
1.788 | 4.8 ± 1.09 | 108.2 ± 10.89 |
1.503 | 5.4 ± 2.22 | 105.7 ± 8.63 |
1.227 | 7.9 ± 3.14 | 109.8 ± 9.34 |
1.009 | 89.9 ± 7.87 | 234.1 ± 20.06 |
0.799 | 2656 ± 251.1 | 2447 ± 183.0 |
0.485 | 55479 ± 1325 | 48360 ± 1938 |
2/10/23 Same procedure as 2/3/23, just changing HV applied to He3 detector to values of 1300V then 1100V.
Discriminator Voltage (V) | Counts/10s (with source) | Counts/10s (no source) |
---|---|---|
1.971 | 4.6 ± 2.37 | 107.7 ± 13.17 |
1.775 | 7.7 ± 2.11 | 100 ± 10.70 |
1.486 | 5.5 ± 2.32 | 104.8 ± 12.79 |
1.248 | 4.7 ± 2.00 | 99.6 ± 7.09 |
1.013 | 75.2 ± 12.17 | 171 ± 12.82 |
0.762 | 3074 ± 143.0 | 2833 ± 894.7 |
0.509 | 19658 ± 424.6 | 40290 ± 2402 |
Discriminator Voltage (V) | Counts/10s (with source) | Counts/10s (no source) |
---|---|---|
1.971 | 5.4 ± 1.78 | 111.3 ± 9.97 |
1.773 | 5.3 ± 1.89 | 97.3 ± 7.70 |
1.524 | 7.1 ± 4.51 | 99.7 ± 11.78 |
1.275 | 12.2 ± 3.55 | 117.3 ± 23.11 |
0.993 | 177.7 ± 8.18 | 351.9 ± 28.71 |
0.784 | 2691.2 ± 114.7 | 2763.7 ± 126.2 |
0.485 | 50346 ± 3831 | 53920 ± 4354 |