{|border="1"  | cellpadding="20" cellspacing="0  class="wikitable"
 
{|border="1"  | cellpadding="20" cellspacing="0  class="wikitable"
 
|-
 
|-
|Run # || date || Sense HV ||Guard HV || Field HV || Eff W1 || Eff W2 || Eff W3 || Eff W4 || Eff W5 || Eff W6  
+
|Run # || date || Sense HV ||Guard HV || Field HV || Eff W1 || Eff W2 || Eff W3 || Eff W4 || Eff W5 || Eff W6
 
|-
 
|-
 
|'''4146'''|| 5/29/13 || 1500|| -750|| 550 || 0.995195 || 0.994042 || 0.993335 || 0.992137 || 0.993069 || 0.996362
 
|'''4146'''|| 5/29/13 || 1500|| -750|| 550 || 0.995195 || 0.994042 || 0.993335 || 0.992137 || 0.993069 || 0.996362
Line 11: Line 11:
 
|'''4157'''|| 6/21/13 || 1300 || -650 || 473 || 0.345951 || 0.174296 || 0.163732 || 0.15669 || 0.171655 || 0.34243
 
|'''4157'''|| 6/21/13 || 1300 || -650 || 473 || 0.345951 || 0.174296 || 0.163732 || 0.15669 || 0.171655 || 0.34243
 
|-
 
|-
|'''4158'''|| 6/24/13 || 1400 || 509 || -700 || 0.380383 || 0.165072 || 0.174641 || 0.160287 || 0.165072 || 0.361244
+
|'''4158'''|| 6/24/13 || 1400 || 509 || -700 || 0.380383 || 0.165072 || 0.174641 || 0.160287 || 0.165072 || 0.361244  
|-
 
|4159|| 6/25/13 || 58 mv Threshold, S:F:G=1400:-700:509 ,  run started at 11:06:05 on 6/25/13, ended at  11:12:55 on 6/26/13. ROC
 
|-
 
|4160|| 6/26/13 || 58 mv Threshold, S:F:G=1400:-700:509 ,  run started at 13:26:51 on 6/26/13, ER failed
 
|-
 
|4161|| 6/26/13 || 58 mv Threshold, S:F:G=1400:-700:509 ,  run started at 21:13:27 on 6/26/13, ended at 09:16:32 on 6/27/13, TDC40->TDC43 have events that are way out of time, restart run
 
|-
 
|4162|| 6/27/13 || 58 mv Threshold, S:F:G=1400:-700:509 ,  run started at 09:17:02 on 6/27/13, ended at  on TDC40->TDC43 still have events that are way out of time, APlatform:ERROR Lost connection to the container = daq2.physics.isu.edu_admin. 15:39:54 06/24
 
|-
 
|4163|| 6/27/13 || 58 mv Threshold, S:F:G=1400:-700:509 ,  run started at 09:17:02 on 6/27/13, ended at  on
 
|-
 
|4165|| 7/3/13 || 58 mv Threshold, S:F:G=1400:-700:509 ,  ADB cards were swapped to see if the out of time events in TDC ch 40->43 move to the channels in the other ADC, they did not, switch the cables back and check the ADC cables.
 
|-
 
|4166|| 7/5/13 || 58 mv Threshold, S:F:G=1400:-700:509 ,  swapped the ADB output cables such that ADB outputs in TDC chan 32->47 are now going into TDC channels 81->96.  The problems with ut of time hits moved with the ADC output cables.  The question now; Is it the cable, the connectors, or the TDC?
 
 
|-
 
|-
 
|}
 
|}
 +
 +
[[File:row1.png]]
 +
 +
[[File:row2.png]]
 +
 +
[[File:row3.png]]
 +
 +
[[File:row4.png]]
 +
 +
[[File:row5.png]]
 +
 +
[[File:row6.png]]
  
 
[[CLAS12_R1_S1_QA#Efficiency_measurement]]
 
[[CLAS12_R1_S1_QA#Efficiency_measurement]]

Latest revision as of 23:07, 13 August 2013

58 mv Threshold

Run # date Sense HV Guard HV Field HV Eff W1 Eff W2 Eff W3 Eff W4 Eff W5 Eff W6
4146 5/29/13 1500 -750 550 0.995195 0.994042 0.993335 0.992137 0.993069 0.996362
4155 6/17/13 1250 -625 458 0.909917 0.826014 0.818991 0.817118 0.934067 0.891657
4157 6/21/13 1300 -650 473 0.345951 0.174296 0.163732 0.15669 0.171655 0.34243
4158 6/24/13 1400 509 -700 0.380383 0.165072 0.174641 0.160287 0.165072 0.361244

Row1.png

Row2.png

Row3.png

Row4.png

Row5.png

Row6.png

CLAS12_R1_S1_QA#Efficiency_measurement