Difference between revisions of "Paste painting procedure"
		
		
		
		
		
		Jump to navigation
		Jump to search
		
				
		
		
	
| (66 intermediate revisions by one other user not shown) | |||
| Line 1: | Line 1: | ||
| − | =  | + | =Purpose=  | 
| − | |||
| − | + | This document description the application proceedures to safely apply the ED-7100 carbon paste.  | |
| − | + | [[File:ED7100_Series_paste.pdf]]  | |
| + | |||
| + | =Requirements for safe handling=  | ||
| + | Before opening the paste please read  the ED-7100 Safety Data Sheet located at   | ||
| − | + | :http://wiki.iac.isu.edu/index.php/File:ED7100_OSHA_MSDS.pdf  | |
| − | |||
| + | Be sure that the following are available:  | ||
| − | + | ;Ventilation:  | |
| + | An externally venting fume hood is recommended when working with this paste or a work area with very good ventilation.  | ||
| − | ;  | + | ;Poor Ventilation area:  | 
| + | Wear a well fitted NIOSH approved or industrial type canister mask.  | ||
| − | + | :http://www1.mscdirect.com/CGI/NNSRIT?PMPXNO=1651332&PMT4NO=0  | |
| − | |||
| − | |||
| − | + | :http://www1.mscdirect.com/Disposable-Respirators-And-Masks/Respiratory-Protection/Safety/s0000002069.HTML  | |
| − | + | ;The following are required for applying the paste:  | |
| − | + | # Gloves,goggles.  | |
| − | + | #suitable protection clothes.  | |
| − | + | # sack of sand or absorber for cleaning the area.  | |
| − | ; a-   | + | =Paste Application Procedure=  | 
| + | ; a- Review the "Requirements for Safe Handling" section preceding this one before using the paste.[http://wiki.iac.isu.edu/index.php/Paste_painting_procedure#Requirements_for_safe_handling]  | ||
; b- Equipment And Tools:  | ; b- Equipment And Tools:  | ||
| − | + | #ED-7100 paste  | |
| − | + | # isobutanol for cleaning,  | |
| − | + | #CLR A/F lens blue frame revolution dust goggles  | |
| − | + | #27"WD 36"L vinyl clear ansell chem-resist apron  | |
| − | + | #35"WD 45"L neoprene Black ansell chem-resist apron  | |
| − | + | #8511 N90 3M particulate respirator  | |
| − | + | #Nitrile disposal gloves  | |
| − | + | ||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| − | |||
| + | ; d- Application steps:  | ||
| + | #put on mask and apron (use goggles if spraying paste using printer)  | ||
| + | #establish a work area under the ventilatin hood in the Laboratory for Detector Science#Add the paste to copper plate.(using a brush or a semi-mechanical printer)  | ||
| + | #mix paste (if reducing is needed) under the hood for sufficient ventilation  | ||
| + | #use applicator to smear paste on pc-board  | ||
| + | #Clean the paste residues using isobutanol cleaner.  | ||
| + | # Cure for 1 - 2h 30min. at 150 - 165 degrees Celsius.(Contact the company for IR-curing)  | ||
| − | + | Back to the main page [http://wiki.iac.isu.edu/index.php/Neutron_TGEM_Detector_Abdel#10.2F27.2F09_.28Paste_painting_procedure.29]  | |
Latest revision as of 15:55, 5 March 2012
Purpose
This document description the application proceedures to safely apply the ED-7100 carbon paste. File:ED7100 Series paste.pdf
Requirements for safe handling
Before opening the paste please read the ED-7100 Safety Data Sheet located at
Be sure that the following are available:
- Ventilation
 
An externally venting fume hood is recommended when working with this paste or a work area with very good ventilation.
- Poor Ventilation area
 
Wear a well fitted NIOSH approved or industrial type canister mask.
- The following are required for applying the paste
 
- Gloves,goggles.
 - suitable protection clothes.
 - sack of sand or absorber for cleaning the area.
 
Paste Application Procedure
- a- Review the "Requirements for Safe Handling" section preceding this one before using the paste.[1]
 
- b- Equipment And Tools
 
- ED-7100 paste
 - isobutanol for cleaning,
 - CLR A/F lens blue frame revolution dust goggles
 - 27"WD 36"L vinyl clear ansell chem-resist apron
 - 35"WD 45"L neoprene Black ansell chem-resist apron
 - 8511 N90 3M particulate respirator
 - Nitrile disposal gloves
 
- d- Application steps
 
- put on mask and apron (use goggles if spraying paste using printer)
 - establish a work area under the ventilatin hood in the Laboratory for Detector Science#Add the paste to copper plate.(using a brush or a semi-mechanical printer)
 - mix paste (if reducing is needed) under the hood for sufficient ventilation
 - use applicator to smear paste on pc-board
 - Clean the paste residues using isobutanol cleaner.
 - Cure for 1 - 2h 30min. at 150 - 165 degrees Celsius.(Contact the company for IR-curing)
 
Back to the main page [2]