Difference between revisions of "Paste painting procedure"
Jump to navigation
Jump to search
(78 intermediate revisions by one other user not shown) | |||
Line 1: | Line 1: | ||
− | = | + | =Purpose= |
− | |||
− | + | This document description the application proceedures to safely apply the ED-7100 carbon paste. | |
− | + | [[File:ED7100_Series_paste.pdf]] | |
− | + | =Requirements for safe handling= | |
− | + | Before opening the paste please read the ED-7100 Safety Data Sheet located at | |
+ | :http://wiki.iac.isu.edu/index.php/File:ED7100_OSHA_MSDS.pdf | ||
− | + | Be sure that the following are available: | |
− | ; | + | ;Ventilation: |
+ | An externally venting fume hood is recommended when working with this paste or a work area with very good ventilation. | ||
− | + | ;Poor Ventilation area: | |
− | + | Wear a well fitted NIOSH approved or industrial type canister mask. | |
− | |||
− | + | :http://www1.mscdirect.com/CGI/NNSRIT?PMPXNO=1651332&PMT4NO=0 | |
+ | :http://www1.mscdirect.com/Disposable-Respirators-And-Masks/Respiratory-Protection/Safety/s0000002069.HTML | ||
− | |||
− | + | ;The following are required for applying the paste: | |
− | ; | ||
− | |||
− | ; a- | + | # Gloves,goggles. |
+ | #suitable protection clothes. | ||
+ | # sack of sand or absorber for cleaning the area. | ||
+ | |||
+ | =Paste Application Procedure= | ||
+ | ; a- Review the "Requirements for Safe Handling" section preceding this one before using the paste.[http://wiki.iac.isu.edu/index.php/Paste_painting_procedure#Requirements_for_safe_handling] | ||
; b- Equipment And Tools: | ; b- Equipment And Tools: | ||
− | + | #ED-7100 paste | |
− | + | # isobutanol for cleaning, | |
− | + | #CLR A/F lens blue frame revolution dust goggles | |
− | + | #27"WD 36"L vinyl clear ansell chem-resist apron | |
− | + | #35"WD 45"L neoprene Black ansell chem-resist apron | |
− | + | #8511 N90 3M particulate respirator | |
− | + | #Nitrile disposal gloves | |
− | + | ||
− | |||
− | |||
− | |||
− | |||
− | + | ; d- Application steps: | |
− | + | #put on mask and apron (use goggles if spraying paste using printer) | |
+ | #establish a work area under the ventilatin hood in the Laboratory for Detector Science#Add the paste to copper plate.(using a brush or a semi-mechanical printer) | ||
+ | #mix paste (if reducing is needed) under the hood for sufficient ventilation | ||
+ | #use applicator to smear paste on pc-board | ||
+ | #Clean the paste residues using isobutanol cleaner. | ||
+ | # Cure for 1 - 2h 30min. at 150 - 165 degrees Celsius.(Contact the company for IR-curing) | ||
− | |||
− | |||
− | [ | + | Back to the main page [http://wiki.iac.isu.edu/index.php/Neutron_TGEM_Detector_Abdel#10.2F27.2F09_.28Paste_painting_procedure.29] |
Latest revision as of 15:55, 5 March 2012
Purpose
This document description the application proceedures to safely apply the ED-7100 carbon paste. File:ED7100 Series paste.pdf
Requirements for safe handling
Before opening the paste please read the ED-7100 Safety Data Sheet located at
Be sure that the following are available:
- Ventilation
An externally venting fume hood is recommended when working with this paste or a work area with very good ventilation.
- Poor Ventilation area
Wear a well fitted NIOSH approved or industrial type canister mask.
- The following are required for applying the paste
- Gloves,goggles.
- suitable protection clothes.
- sack of sand or absorber for cleaning the area.
Paste Application Procedure
- a- Review the "Requirements for Safe Handling" section preceding this one before using the paste.[1]
- b- Equipment And Tools
- ED-7100 paste
- isobutanol for cleaning,
- CLR A/F lens blue frame revolution dust goggles
- 27"WD 36"L vinyl clear ansell chem-resist apron
- 35"WD 45"L neoprene Black ansell chem-resist apron
- 8511 N90 3M particulate respirator
- Nitrile disposal gloves
- d- Application steps
- put on mask and apron (use goggles if spraying paste using printer)
- establish a work area under the ventilatin hood in the Laboratory for Detector Science#Add the paste to copper plate.(using a brush or a semi-mechanical printer)
- mix paste (if reducing is needed) under the hood for sufficient ventilation
- use applicator to smear paste on pc-board
- Clean the paste residues using isobutanol cleaner.
- Cure for 1 - 2h 30min. at 150 - 165 degrees Celsius.(Contact the company for IR-curing)
Back to the main page [2]