Difference between revisions of "HRRL 7-31-09"
(38 intermediate revisions by the same user not shown) | |||
Line 84: | Line 84: | ||
==TDC Calibration== | ==TDC Calibration== | ||
+ | Second pulse width is 80 ns. | ||
+ | ;Conclusion | ||
+ | |||
+ | In order to get actual time between the start and the stop pulse, one needs to subtract the width of the second signal. In this case it was 80 ns. | ||
{| border="1" cellpadding="20" cellspacing="0" | {| border="1" cellpadding="20" cellspacing="0" | ||
|- | |- | ||
− | | Run Number || Scope Image || TDC Histogram || Calculated Time (ns) | + | | Run Number || Scope Image || TDC Histogram || Calculated Time (ns) || Time measured on the scope ns (between the fall of the first pulse and rise of the second pulse) || Time measured on the scope ns (between the fall of the first pulse and fall of the second pulse) || Pulse width (ns) |
+ | |- | ||
+ | | 1238 || [[File:imageTDCCalibration_1.png|200px]] || [[File:imageTDCCalibration_1.gif|200px]] || 169.5 || 160 || 80 || 80 | ||
|- | |- | ||
− | | | + | | 1239 || [[File:imageTDCCalibration_2.png|200px]] || [[File:imageTDCCalibration_2.gif|200px]] || 135.7 || 130 || 50 || 80 |
|- | |- | ||
− | | | + | | 1240 || [[File:imageTDCCalibration_3.png|200px]] || [[File:imageTDCCalibration_3.gif|200px]] || 111.5 || 110 || 30 || 80 |
|- | |- | ||
− | | | + | | 1328 || [[File:imageTDCCalibration_4.png|200px]] || [[File:imageTDCCalibration_4.gif|200px]] || 122 || 120 || -50 || 170 |
|} | |} | ||
==Actual Measurements== | ==Actual Measurements== | ||
+ | |||
+ | |||
+ | We had runs for different beam energy (2 && 4 MeV) and collimator size(~0.2 && ~0.8 cm).<br> | ||
+ | Table below shows TDC measurements for all cases for Metalica and Plastika. | ||
{| border="1" cellpadding="20" cellspacing="0" | {| border="1" cellpadding="20" cellspacing="0" | ||
Line 108: | Line 118: | ||
|TDC Histogram | |TDC Histogram | ||
|- | |- | ||
− | | 1296 || 2 || 0.8 || 1400:-700:980 || 50 || [[File:TDCSpectrum_1296.gif| | + | | 1296 || 2 || 0.8 || 1400:-700:980 || 50 || [[File:TDCSpectrum_1296.gif|200px]][[File:TDCSpectrum_1296_1.gif|200px]] |
|- | |- | ||
− | | 1297 || 4 || 0.8 || 1400:-700:980 || 50 || [[File: | + | | 1297, 1298 && 1299 || 4 || 0.8 || 1400:-700:980 || 50 || [[File:TDCSpectrum_1297_8_9.gif|200px]][[File:TDCSpectrum_1297_8_9_1.gif|200px]] |
|- | |- | ||
− | | | + | | 1304 && 1305 || 4 || 0.2 || 1400:-700:980 || 50 || [[File:TDCSpectrum_1304-1305.gif|200px]][[File:TDCSpectrum_1304-1305_1.gif|200px]] |
|- | |- | ||
− | | | + | | 1306 || 2 || 0.8 || 1400:-700:980 || 50 || [[File:TDCSpectrum_1306.gif|200px]][[File:TDCSpectrum_1306_1.gif|200px]] |
− | |||
− | |||
− | |||
− | |||
|} | |} | ||
+ | |||
+ | |||
+ | Plastika - 120 ns, Metalica - 170 ns. | ||
+ | |||
+ | [[File:HRRL_7-31-09_MetalicaPlastika_Start.png|250px]][[File:HRRL_7-31-09_MetalicaPlastika_Start_1.png|250px]] | ||
+ | |||
+ | |||
+ | ;Example | ||
+ | |||
+ | [[File:TDCSpectrum_1304_5_gaussian_fitting.gif|200px]] | ||
+ | |||
+ | For Metalica on the TDC histogram mean value is 307, but the time between the pulse of the PMT+RFGate and Metalica is 307-170=137 ns. | ||
+ | For Plastika it is 226.4-120=106.4ns. So time difference between this two pulses is around ~30 ns(the same on the scope). | ||
+ | |||
+ | |||
+ | The drift velocity for electrons in drift chambers is ~ 50μm / ns. [[Media:mestayer.pdf]] | ||
+ | [edit] Calculated result using TDC data | ||
+ | |||
+ | The size of the drift chamber cell is 0.86 cm. | ||
+ | |||
+ | <math>Time = \frac{1000 {\rm ns}}{2^{12}{\rm chan}} \times {\rm chan} ns</math> | ||
+ | |||
+ | <math>{t_{Drift}}^{maximum} = \frac{CellSize}{v_{Drift}} = \frac{0.86 \times 10^{-2} m}{50 \mu m/ns} = 172 ns</math> | ||
+ | |||
+ | So the maximum drift time in DCs(Metalica && Plastika) can be 172 ns. | ||
+ | |||
+ | Our data(above example) shows that for Metalica it is ~137+/-100ns and for Plastika it is ~106+/? | ||
=VFAT runs= | =VFAT runs= | ||
Line 154: | Line 187: | ||
[[File:Pb_Attenuation_Photon.pdf]] | [[File:Pb_Attenuation_Photon.pdf]] | ||
+ | |||
+ | =Experimental Setup= | ||
+ | |||
+ | [[File:0-8cm_collimator.jpg|250px]][[File:0-2cm_collimator.jpg|250px]] | ||
+ | |||
+ | [[File:Detectors_7-31-09_HRRL_ExperimentSetup.gif|850px]] | ||
+ | |||
+ | |||
+ | [[File:Detectors_7-31-09_HRRL_ExperimentSetup_1.jpg|550px]] |
Latest revision as of 20:22, 12 August 2009
Default HRRL running Conditions
Pulse Width | Rep Rate | Energy | |
8 mA | 100 ns | 1 Hz | 3 MeV |
cable names
CH ID | Description |
2223A1 | Front Scintillator |
2223A2 | Rear Scintilaltor |
2223A3 | ADC Gate From counting house to Experimental Cell |
2223A4 | TDC start pulse from counting house to experimental cell (PMT front + beam Gate) |
2223A5 | Coincidenc between front + rear Scint + Gun pulser to scaler channel 8 |
2223A6 | Metalica sense wire 4 |
2223A7 | Delayed rear PMT pulse to ADC |
2223A8 | Plastica sense wire 4 |
2223A9 | Plastica to scaler channel 4 |
2223A10 | Front scintillator to ADC |
2223A11 | Metalica scaler channel 1 |
2223A12 | DAQ trigger |
DAQ Ntuple names
Ntuple Variable | Description |
ADC5 | Front Scintilator |
ADC1 | Metalica sense wire 4 |
ADC16 | Rear Scintilator |
ADC13 | Plastica sense wire 4 |
TDC1 | Metalica sense wire 4 |
TDC3 | Plastika sense wire 4 |
ADC + Scaler + TDC + V1495
Run Number | Distance to Metallica's Feedthrough tip (cm) | Metalica scaler counts scaler ch 4 | Plastika Scaler Counts scaler ch 1 | Rear PMT + Front PMT + Gun Pulse Counts Scaler Ch 8 | HV S:F:G | Discriminator (mV) | Metalica counts/PMT+Gun pulse |
1263 | 19 | 0 | 1400:-700:980 | 50 |
TDC Measurements
TDC Calibration
Second pulse width is 80 ns.
- Conclusion
In order to get actual time between the start and the stop pulse, one needs to subtract the width of the second signal. In this case it was 80 ns.
Actual Measurements
We had runs for different beam energy (2 && 4 MeV) and collimator size(~0.2 && ~0.8 cm).
Table below shows TDC measurements for all cases for Metalica and Plastika.
Plastika - 120 ns, Metalica - 170 ns.
- Example
For Metalica on the TDC histogram mean value is 307, but the time between the pulse of the PMT+RFGate and Metalica is 307-170=137 ns. For Plastika it is 226.4-120=106.4ns. So time difference between this two pulses is around ~30 ns(the same on the scope).
The drift velocity for electrons in drift chambers is ~ 50μm / ns. Media:mestayer.pdf
[edit] Calculated result using TDC data
The size of the drift chamber cell is 0.86 cm.
So the maximum drift time in DCs(Metalica && Plastika) can be 172 ns.
Our data(above example) shows that for Metalica it is ~137+/-100ns and for Plastika it is ~106+/?
VFAT runs
4 MeV beam energy, big collimator.
GEM HV settings:
=3425/3125.Run Number | Card 1 Threshold | Card 2 Threshold |
1300 | 151 | 116 |
1301 | 153 | 118 |
1302 | 155 | 120 |
1303 | 157 | 122 |
GEM HV settings:
=3425/3125.4 MeV beam energy, small collimator.
Run Number | Card 1 Threshold | Card 2 Threshold |
1305 | 151 | 124 |
File:Pb Attenuation Photon.pdf