Yes, when the cathode voltage decreased  to -3200V, QDC histogram does not show a difference in in the collected charge as the shutter is open and the source is on, and when shutter is open without the source the source on it.  
+
Yes, when the cathode voltage is decreased  to -3200V, QDC histogram does not show any difference in the collected charge as the shutter is open with the source is on it, and when shutter is open without the source.  
  
 
[[File: QDC_source_on_off_7728_7729.png | 300 px]]
 
[[File: QDC_source_on_off_7728_7729.png | 300 px]]
 +
 +
  
 
[[Neutron_TGEM_Detector_Abdel]]
 
[[Neutron_TGEM_Detector_Abdel]]

Revision as of 23:28, 16 September 2014

9/16/14

QDC source on off 7724 7726.png

Condition of above results

Shutter is Open

Red is Cf-252 source ON Run 7724

Green is Cf-252 source OFF run 7726

HV_GEM= -2930 Volts HV_Cathode=-3400 Volts

Gas Flow rate = 0.1 ft^3/hr

Goal
Can the Cf-252 source ON signal be changed?

Change the Cathode voltage to try and turn off the signal when the source is on.

Result

Yes, when the cathode voltage is decreased to -3200V, QDC histogram does not show any difference in the collected charge as the shutter is open with the source is on it, and when shutter is open without the source.

QDC source on off 7728 7729.png


Neutron_TGEM_Detector_Abdel