Difference between revisions of "Faraday Cup Analysis"
Line 8: | Line 8: | ||
For distribution over all ADC channel: | For distribution over all ADC channel: | ||
− | <math> ADC^{avg}=\frac{\sum_{i=1}^{16}{Q_{i}*i}}{\sum_{i=1}^{16}{Q_{ | + | <math> ADC^{avg}=\frac{\sum_{i=1}^{16}{Q_{i}*i}}{\sum_{i=1}^{16}{Q_{i}}}</math><br> |
??? <math> ADC^{err}=\frac{\sum_{i=1}^{16}{q_{channel}*i}}{\sum_{i=1}^{16}{ADCErr_{channel}}}</math><br> | ??? <math> ADC^{err}=\frac{\sum_{i=1}^{16}{q_{channel}*i}}{\sum_{i=1}^{16}{ADCErr_{channel}}}</math><br> | ||
Revision as of 19:54, 13 April 2010
FC analysis using ADC channel current distribution
For each ADC channel:
(total charge per one channel)
(error in charge distribution per channel)
So for each channel we have (Q +- q)
For distribution over all ADC channel:
???
FC analysis using pulse by pulse ADC channel mean value distribution
For each beam pulse:
For distribution over all beam pulses:
Here is:
1. ADC# = bridge#
2. Pulse# = ReadOut# = Entry# = Event#
Some examples of ADC mean value distribution. Here are:
1. x axis: ADC mean value for one pulse
2. y axis: number of pulse w/ that ADC mean value
Faraday Cup 3D plot
Below is the plot of the charge in Faraday cup (pC) as a function of magnet current (vertical axis, A) (basically magnetic field) and ADC (horizontal axis).
Faraday Cup ADC channel distribution
Faraday Cup "Rain"
Error Analysis
The ADC measures the charge deposited on each of the 16 Aluminum blocks. The ADC is a 12 bit ADC with a max input of 400 pc. THe means that the charge per channel is:
Run1477 -5A
Raw ADC result for channel 8 :
ADC 8 using channel -> Coul conversion:
The above histogram shows an RMS of 66.74 pC.
Comment
- The ADC basically counts the number of electrons collected by the aluminum FC bricks and transfered through the cables to the ADC. This is a poisson process with a large number of trials leading to a large mean value
The histogram shows
- Mean =
- RMS =
- What does this mean?
- The theoretical distribution would be Guassian with . The above suggests that the beam charge delivered to the FC is not following this statistical parent distribution. Most likely the beam current is changing while you are measuring the charge with the FC.
- Is the charge lost or is it just moving to different FC channels
Within 100 RF pulses the total charge on the FC drops from 600 to 100 pC. It seems the beam current is very unstable.
There also appears to be a gaussian distribution centered around 600 pC and a wider one centered around 100 pC.
Could it be that I can put a cut on total beam current and see if the charge moves between FC elements?
Let see what the total charge looks like for 3 cases.