Channel 17 is a dose measuring detector on in the HRRL experimental cell. | Channel 17 is a dose measuring detector on in the HRRL experimental cell. | ||
− | == Calculating Number of | + | == Calculating Number of Incident Electrons Per Second == |
We have electron beam of: | We have electron beam of: | ||
Line 269: | Line 269: | ||
<math> N = f*I*\Delta t/e= 110*0.026*1*10^{-6}/(1.6*10^{-19}) = 1.7875*10^{13} </math> | <math> N = f*I*\Delta t/e= 110*0.026*1*10^{-6}/(1.6*10^{-19}) = 1.7875*10^{13} </math> | ||
− | |||
− | |||
=Calculation= | =Calculation= |
Revision as of 03:28, 31 May 2010
Building a new positron beamline in the HRRL cell might require moving the linac itself. R adiation levels were measured on Feb 4, 2009 by M. Balzer and G. Stancari to determine the impact of repositioning the HRRL RF cavity in the accelerator room. \
The HRRL was used to accelerator electrons to a 15 MeV beam energy, 20 mA peak current, 1 kHz repetition rate, and 30 ns pulse width. Five OSL dosimeters (For Details on OSL dosimeter, look [[1]]) were placed at each of 15 locations within the cell identified in the sketch below.
Media:dosimeter-locations-20090204.pdf
Readings in mrad are reported in the following table. They were taken before exposure (first column) and after a couple of minutes of machine tuning (second column). Readings in the third column were taken 94 minutes after the second reading. During these 94 minutes, the machine was running with the settings mentioned above.
Media:exposure-measurements-20090204.pdf
Dose rates in mrad/hr at each of the 15 locations can be estimated by subtracting column 2 from column 3, averaging over the 5 dosimeters, and multiplying the result by (60 min/hr) / (94 min). RMS spreads refer to variations within each group of 5.
Location | Dose rate | RMS spread |
---|---|---|
(mrad/hr) | (mrad/hr) | |
1 | 9 | 4 |
2 | 396 | 57 |
3 | 7940 | 204 |
4 | 2831 | 117 |
5 | 4408 | 373 |
6 | 29339 | 3332 |
7 | 72517 | 687 |
8 | 36507 | 4746 |
9 | 5 | 5 |
10 | 37 | 4 |
11 | 164 | 40 |
12 | 2734 | 313 |
13 | 5828 | 120 |
14 | 7793 | 2579 |
15 | 62431 | 27155 |
OSL dosimeters
Dosimeters are placed approximately 1-m off the floor. Large face of the dosimeters were facing HRRL.
Diameter of a dosimeter is measured as:
d = 7.4 mm, 7.6 mm, 7.5 mm, 7.3 mm, 7.3 mm, 7.4 mm, 7.6 mm.
so,
d = 7.44
0.13 mm.Experimental Cell Dose Measurement with HRRL
This experiment is done on 5/17/2010, at 10:00 am.
Dr. Khalid Chouffani wrote the note. Here is photocopy of his note:
page #1
page #2
Detector Used on Channel 16
Link to detector information from producer: [2]
For more detailed information go to: ludlums_Gamma_Area_Monitor_moedl_45-9
16 MeV Beam Energy
Experimental cell measurement with HRRL.
Conditions are:
Rep rate: 40 Hz
Pulse width: 1
E=(10-17) MeV
Peak current of beam (mA) | Energy of beam (MeV) | Measurement on Channel 17, Experimental Cell (mrad) | Measurement on Channel 16, Accelerator cell (mrad) |
100 | 16 | 4.5 | 0.245.5 | 2
69.2 | 16 | 4 | 0.235, 42.5 |
60.8 | 16 | 3.5 | 0.129.2, 28.9 |
58.0 | 16 | 2.8, 2.9 | 26, 26.7 |
47.6 | 16 | 2.3 | 19.6, 21.3 |
33.2 | 16 | 1.8, 1.7 | 19.5, 19.0, 13.7 |
Beam bear through 1/4 inches diameter collimator with 4 inches lead.
16 MeV Beam Energy
Conditions are:
Rep rate: 40 Hz
Peak current of beam (mA) | Energy of beam (MeV) | Measurement on Channel 17, Experimental Cell (mrad) | Measurement on Channel 16, Accelerator cell (mrad) |
136 | 16 | 0.3 | 99.2, 98.7 |
10 MeV Beam Energy
Conditions are:
Rep rate: 110 Hz
E=(8-16) MeV
Peak current of beam (mA) | Energy of beam (MeV) | Measurement on Channel 17, Experimental Cell (mrad) | Measurement on Channel 16, Accelerator cell (mrad) |
48.4 | 10 | 4.6, 4.7 | 117.6, 117.5 |
36.8 | 10 | 3.8, 3.9 | 104.4, 103.5, 102.9 |
30.8 | 10 | 3.1 | 0.172.8, 72.9, 72.5 |
26 | 10 | 2.5 | 56.7, 56.4, 54.6 |
Fit for channel 17
Order of the fit | Fit equations | Chi square | p-value |
1st order | 24.91 | 0.000004 |
Fit for channel 16
Order of the fit | Fit equations | Chi square | p-value |
1st order | 437.5 | less than 0.0001 |
Channel 17 Dose Estimation in New Position
Channel 17 is a dose measuring detector on in the HRRL experimental cell.
Calculating Number of Incident Electrons Per Second
We have electron beam of:
Frequency: f (Hz)
Peak current: I (mAmp = 0.001 Amp)
Pulse width:
t ( seconds )So, how many electrons we have in each second?
By Q=It, we have
Where N is the total electron numbers hits target per second, e is electron charge and f, I and
are given above. So
Number of Particles Per Second in test by M. Balzer on Feb 4, 2009
In radiation levels measured on Feb 4, 2009 by M. Balzer and G. Stancari
Five OSL dosimeters were placed at each of 15 locations in the cell. (What is OSL? Need to ask M. Balzer)
The HRRL was set at 15 MeV beam energy, 20 mA peak current, 1 kHz repetition rate, and 30 ns pulse width.
Number of Particles Per Second in "Experimental Cell Dose Measurement with HRRL" on 5/17/201
This is the experiment we have done on 5/17/201 to measure dose on the detector placed in the corner of the doorway.
Conditions are: Pulse width: 1
, Rap rate: 110 Hz.We measured dose for several peak current. One of the peak current is 26 mA. For this peak current:
Calculation
Assume that the position 13 at radiation test on Feb 4, 2009 by M. Balzer and G. Stancari will shift to detector position at corner on experiment that is done on 5/17/2010.
Radiation test on Feb 4, 2009 by M. Balzer and G. Stancari used OSL dosimeter.
The area of a OSL dosimeter is:
Experimental Cell Dose Measurement with HRRL on 5/17/2010 used Ludlums Gamma Area Monitor moedl 45-9 for channel 16. Ludlums Gamma Area Monitor moedl 45-9 has dimension: 7.6 x 25.7 cm (3 x 10.1 in.) (Dia x L).
The area of a Ludlums Gamma Area Monitor moedl 45-9 is:
The ratio of detectors detecting area is:
Assuming 100% efficiency for both detectors, their detecting are ratio is their dose ratio:
Back to Positrons