Difference between revisions of "SIS3600 Trigger Supervisor"
(7 intermediate revisions by the same user not shown) | |||
Line 1: | Line 1: | ||
+ | =Manual= | ||
+ | |||
+ | [[Image:SIS3610_Manual.pdf]] | ||
+ | |||
=Setting address= | =Setting address= | ||
Line 8: | Line 12: | ||
| Pin || Setting | | Pin || Setting | ||
|- | |- | ||
− | |SW- | + | |SW-A32U || 3 |
|- | |- | ||
− | |SW- | + | |SW-A32L || 8 |
|- | |- | ||
− | |SW- | + | |SW-A24U || 3 |
|- | |- | ||
− | |SW- | + | |SW-A24L || 8 |
|- | |- | ||
|SW-A16 || 3 | |SW-A16 || 3 | ||
Line 126: | Line 130: | ||
= SIS3610 readout list (ROL) = | = SIS3610 readout list (ROL) = | ||
− | The readout list is a script that coda download to the ROC which | + | makelist gen_int.crl ppc -mlongcall |
+ | |||
+ | The readout list is a script that coda download to the ROC which | ||
= SIS3610 run configuration = | = SIS3610 run configuration = | ||
+ | = Input data word tests= | ||
+ | Run 876 | ||
+ | |||
+ | |||
+ | D0 | D1 | D2 | D3| D4SIS3610 Input word= fbfe = 64510 | ||
+ | SIS3610 Input word= 7bfe = 31742 | ||
+ | SIS3610 Input word= 3bfe = 15358 | ||
+ | SIS3610 Input word= 1bfe = 7166 | ||
+ | SIS3610 Input word= bfe = 3070 | ||
+ | SIS3610 Input word= 9fe = 2558 | ||
+ | SIS3610 Input word= 8fe = 2302 | ||
+ | SIS3610 Input word= 87e = 2174 | ||
+ | SIS3610 Input word= 8be = 2238 | ||
+ | SIS3610 Input word= 8ae = 2222 | ||
+ | SIS3610 Input word= 8a6 = 2214 | ||
+ | SIS3610 Input word= 8a2 = 2210 | ||
+ | SIS3610 Input word= 8a0 = 2208 | ||
− | [http:// | + | [http://wiki.iac.isu.edu/index.php/Data_Acquisition Back to DAQ] [[Data_Acquisition]] |
Latest revision as of 16:48, 3 April 2009
Manual
Setting address
Address | |
Pin | Setting |
SW-A32U | 3 |
SW-A32L | 8 |
SW-A24U | 3 |
SW-A24L | 8 |
SW-A16 | 3 |
There was also a of jumpers labeled "EN_A16, "EN_A24" and "EN_A32". The factory default setting was "EN_A32" so I moved the jumper to "EN_16".
I left the base address unchanged. Which means for A16 addressing mode the address will be 0x3800.
loading ROC library
A ROC library is created by compiling "c" code functions which make VxWorks system calls to modules in the VME.
compiling VxWorks Modules
In order to compile VxWorks modules you need the cross compiler ccppc which compiles binaries which will run on PPC chip sets used by the ROC. The cross compiled was distributed as a binary from the CODA group and is located in
/usr/local/gnu/x86-linux/bin/ccppc.
You also need the header files for the particular version of VxWorks you are using. I created a subdirectory called "ROClibs" in which I store all of these libraries and download them onto the ROC. The "ROCLibs" subdirectory is located under the CODA directory but isn't associated with a version of CODA since. I got the VxWorks header files from
/site/vxworks/5.5/ppc/target/h
and copied the whole "h/" subdirectory under "ROClibs"
Makefile.ppc
The Makefile below is used to compile the libraries
# # Description: # Makefile for PPC code # ifndef ARCH ARCH = VXWORKSPPC endif #example #ccppc -i adc1182.c -fno-builtin -fno-for-scope -fstrength-reduce -nostdinc -nostdinc -mcpu=604 -DCPU=PPC604 -DVXWORKS -D_GNU_TOOL -DVXWORKSPPC -I/usr/local/coda/2.5/common/include/ -I./h -o adc1182_ppc.o CC = ccppc LD = ldppc LIBDIR = $(CODA)/$(ARCH)/lib #DEFINES = -DVXWORKS -DVXWORKSPPC -DCPU=PPC604 -D_GNU_TOOL DEFINES = -mcpu=604 -DCPU=PPC604 -DVXWORKS -D_GNU_TOOL -DVXWORKSPPC INCS = -I$(CODA)/common/include/ -I./h CFLAGS = -fno-builtin -fno-for-scope -fstrength-reduce -mlongcall $(DEFINES) -I$(CODA)/common/include -I./h #OBJS = adc1182_ppc.o scale32Lib.o usrstrutils.o #OBJS = adc1182_ppc.o scale32Lib.o pas9740Lib.o OBJS = sis3610Lib.o all: $(OBJS) sis3610LIB.o: sis3610Lib.c sis3610.h $(CC) -i sis3610Lib.c $(CFLAGS) -o sis3610Lib_ppc.o clean: distclean distclean: rm -f $(OBJS) exports: install install: cp $(OBJS) $(CLON_VXWORKS)/code
sis3610Lib
The SIS3610 library source code is given below:
The header file is:
just type
make -f Makefile.ppc
to compile the library
Test Communication from ROC console
If the ROC didn't download the sis library from within the boot script then you will need to do it by hand from the ROC console.
->cd "/home/daq/CODA/ROClibs" ->ld < sis3610Lib.o
From the ROC console (minicom window) I tried talking to the SIS3610 using the library function "s3610Init"
-> s3610Init 0x3800 Initialized SIS3610 ID 0 at address 0x91003800 value = 0 = 0x0
The next test was to try and turn the user LED (labeled "U" on the front pannel) on and off.
From the ROC console I typed
-> s3610SetLED value = 0 = 0x0
and I saw the green LED light up
I then typed
-> s3610ClearLED value = 0 = 0x0
and I saw it turn off
SIS3610 readout list (ROL)
makelist gen_int.crl ppc -mlongcall
The readout list is a script that coda download to the ROC which
SIS3610 run configuration
Input data word tests
Run 876
D0 | D1 | D2 | D3| D4SIS3610 Input word= fbfe = 64510 SIS3610 Input word= 7bfe = 31742 SIS3610 Input word= 3bfe = 15358 SIS3610 Input word= 1bfe = 7166 SIS3610 Input word= bfe = 3070 SIS3610 Input word= 9fe = 2558 SIS3610 Input word= 8fe = 2302 SIS3610 Input word= 87e = 2174 SIS3610 Input word= 8be = 2238 SIS3610 Input word= 8ae = 2222 SIS3610 Input word= 8a6 = 2214 SIS3610 Input word= 8a2 = 2210 SIS3610 Input word= 8a0 = 2208