Pu-Be 18 gram source# M1197
 
Pu-Be 18 gram source# M1197
  
Run 4546 had a discriminator threshold set to 80 mV and the terminators were out. Cosmic rays test.
+
* Run 4546 had a discriminator threshold set to 80 mV and the terminators were out. Cosmic rays test.
  
 
{| cellpadding="2" style="border: 1px solid darkgray;"
 
{| cellpadding="2" style="border: 1px solid darkgray;"
Line 15: Line 15:
 
|}
 
|}
  
Run 4547 had a discriminator threshold set to 80 mV and the terminators were out.  DAQ rate is 100 Hz. Pu-Be source in.
+
* Run 4547 had a discriminator threshold set to 80 mV and the terminators were out.  DAQ rate is 100 Hz. Pu-Be source in.
  
 
{| cellpadding="2" style="border: 1px solid darkgray;"
 
{| cellpadding="2" style="border: 1px solid darkgray;"
Line 25: Line 25:
 
|}
 
|}
  
Run 4548 had a discriminator threshold set to 80 mV and the terminators were in.  DAQ rate is 100 Hz. Pu-Be source in.
+
* Run 4548 had a discriminator threshold set to 80 mV and the terminators were in.  DAQ rate is 100 Hz. Pu-Be source in.
  
 
{| cellpadding="2" style="border: 1px solid darkgray;"
 
{| cellpadding="2" style="border: 1px solid darkgray;"
Line 35: Line 35:
 
|}
 
|}
  
Run 4549 had a discriminator threshold set to 40 mV and the terminators were in.  DAQ rate is 2800 Hz. Pu-B source in.
+
* Run 4549 had a discriminator threshold set to 40 mV and the terminators were in.  DAQ rate is 2800 Hz. Pu-Be source in.
  
 
{| cellpadding="2" style="border: 1px solid darkgray;"
 
{| cellpadding="2" style="border: 1px solid darkgray;"
Line 45: Line 45:
 
|}
 
|}
  
Run 4550 had a discriminator threshold set to 40 mV and the terminators were out.  DAQ rate is 2800 Hz. Pu-B source in.
+
* Run 4550 had a discriminator threshold set to 40 mV and the terminators were out.  DAQ rate is 2800 Hz. Pu-Be source in.
  
 
{| cellpadding="2" style="border: 1px solid darkgray;"
 
{| cellpadding="2" style="border: 1px solid darkgray;"
Line 55: Line 55:
 
|}
 
|}
  
Run 4551 had a discriminator threshold set to 40 mV and the terminators were out.  DAQ rate is 500 Hz. Pu-B source in, moved away from the detector by 26".
+
* Run 4551 had a discriminator threshold set to 40 mV and the terminators were out.  DAQ rate is 500 Hz. Pu-Be source in, moved away from the detector by 26".
  
 
{| cellpadding="2" style="border: 1px solid darkgray;"
 
{| cellpadding="2" style="border: 1px solid darkgray;"
Line 65: Line 65:
 
|}
 
|}
  
Run 4552 had a discriminator threshold set to 40 mV and the terminators were in.  DAQ rate is 500 Hz. Pu-B source in, moved away from the detector by 26".
+
* Run 4552 had a discriminator threshold set to 40 mV and the terminators were in.  DAQ rate is 500 Hz. Pu-Be source in, moved away from the detector by 26".
  
 
{| cellpadding="2" style="border: 1px solid darkgray;"
 
{| cellpadding="2" style="border: 1px solid darkgray;"
Line 75: Line 75:
 
|}
 
|}
  
Run 4553 had a discriminator threshold set to 20 mV and the terminators were in.  DAQ rate is 1850 Hz. Pu-Be source in, moved away from the detector by 26".
+
* Run 4553 had a discriminator threshold set to 20 mV and the terminators were in.  DAQ rate is 1850 Hz. Pu-Be source in, moved away from the detector by 26".
  
 
{| cellpadding="2" style="border: 1px solid darkgray;"
 
{| cellpadding="2" style="border: 1px solid darkgray;"
Line 85: Line 85:
 
|}
 
|}
  
Run 4554 had a discriminator threshold set to 20 mV and the terminators were in.  DAQ rate is 625 Hz. Pu-Be source in, moved away from the detector by 52".
+
* Run 4554 had a discriminator threshold set to 20 mV and the terminators were in.  DAQ rate is 625 Hz. Pu-Be source in, moved away from the detector by 52".
  
 
{| cellpadding="2" style="border: 1px solid darkgray;"
 
{| cellpadding="2" style="border: 1px solid darkgray;"
Line 95: Line 95:
 
|}
 
|}
  
Run 4555 had a discriminator threshold set to 20 mV and the terminators were out.  DAQ rate is 650 Hz. Pu-Be source in, moved away from the detector by 52".
+
* Run 4555 had a discriminator threshold set to 20 mV and the terminators were out.  DAQ rate is 650 Hz. Pu-Be source in, moved away from the detector by 52".
  
 
{| cellpadding="2" style="border: 1px solid darkgray;"
 
{| cellpadding="2" style="border: 1px solid darkgray;"
Line 105: Line 105:
 
|}
 
|}
  
Run 4556 had a discriminator threshold set to 10 mV and the terminators were out.  DAQ rate is 1400 Hz. Pu-Be source in, moved away from the detector by 52".
+
* Run 4556 had a discriminator threshold set to 10 mV and the terminators were out.  DAQ rate is 1400 Hz. Pu-Be source in, moved away from the detector by 52".
  
 
{| cellpadding="2" style="border: 1px solid darkgray;"
 
{| cellpadding="2" style="border: 1px solid darkgray;"
Line 115: Line 115:
 
|}
 
|}
  
Run 4557 had a discriminator threshold set to 10 mV and the terminators were in.  DAQ rate is 1400 Hz. Pu-B source in, moved away from the detector by 52".
+
* Run 4557 had a discriminator threshold set to 10 mV and the terminators were in.  DAQ rate is 1400 Hz. Pu-B source in, moved away from the detector by 52".
  
 
{| cellpadding="2" style="border: 1px solid darkgray;"
 
{| cellpadding="2" style="border: 1px solid darkgray;"
Line 128: Line 128:
 
  PMT detector #4 HV at -1400 Volts
 
  PMT detector #4 HV at -1400 Volts
  
Run 4558 had a discriminator threshold set to 10 mV and the terminators were in.  DAQ rate is 4670 Hz. Pu-Be source in, moved away from the detector by 52".
+
* Run 4558 had a discriminator threshold set to 10 mV and the terminators were in.  DAQ rate is 4670 Hz. Pu-Be source in, moved away from the detector by 52".
  
 
{| cellpadding="2" style="border: 1px solid darkgray;"
 
{| cellpadding="2" style="border: 1px solid darkgray;"
Line 138: Line 138:
 
|}
 
|}
  
Run 4559 had a discriminator threshold set to 10 mV and the terminators were in.  DAQ rate is 2700 Hz. Pu-Be source in, moved away from the detector by 91".
+
* Run 4559 had a discriminator threshold set to 10 mV and the terminators were in.  DAQ rate is 2700 Hz. Pu-Be source in, moved away from the detector by 91".
  
 
{| cellpadding="2" style="border: 1px solid darkgray;"
 
{| cellpadding="2" style="border: 1px solid darkgray;"
Line 148: Line 148:
 
|}
 
|}
  
Run 4560 had a discriminator threshold set to 10 mV and the terminators were out.  DAQ rate is 2700 Hz. Pu-Be source in, moved away from the detector by 91".
+
* Run 4560 had a discriminator threshold set to 10 mV and the terminators were out.  DAQ rate is 2700 Hz. Pu-Be source in, moved away from the detector by 91".
  
 
{| cellpadding="2" style="border: 1px solid darkgray;"
 
{| cellpadding="2" style="border: 1px solid darkgray;"
Line 158: Line 158:
 
|}
 
|}
  
Run 4561 had a discriminator threshold set to 10 mV and the terminators were out.  DAQ rate is 200 Hz. Cosmic rays test.
+
* Run 4561 had a discriminator threshold set to 10 mV and the terminators were out.  DAQ rate is 200 Hz. Cosmic rays test.
  
 
{| cellpadding="2" style="border: 1px solid darkgray;"
 
{| cellpadding="2" style="border: 1px solid darkgray;"

Latest revision as of 23:53, 25 April 2015

Pu-Be source test

PMT detector #4 HV at -1000 Volts

Pu-Be 18 gram source# M1197

  • Run 4546 had a discriminator threshold set to 80 mV and the terminators were out. Cosmic rays test.
Region 1 Region 2
Fake stop Second hit
  • Run 4547 had a discriminator threshold set to 80 mV and the terminators were out. DAQ rate is 100 Hz. Pu-Be source in.
Region 1 Region 2
Fake stop Second hit
  • Run 4548 had a discriminator threshold set to 80 mV and the terminators were in. DAQ rate is 100 Hz. Pu-Be source in.
Region 1 Region 2
Fake stop Second hit
  • Run 4549 had a discriminator threshold set to 40 mV and the terminators were in. DAQ rate is 2800 Hz. Pu-Be source in.
Region 1 Region 2
Fake stop Second hit
  • Run 4550 had a discriminator threshold set to 40 mV and the terminators were out. DAQ rate is 2800 Hz. Pu-Be source in.
Region 1 Region 2
Fake stop Second hit
  • Run 4551 had a discriminator threshold set to 40 mV and the terminators were out. DAQ rate is 500 Hz. Pu-Be source in, moved away from the detector by 26".
Region 1 Region 2
Fake stop Second hit
  • Run 4552 had a discriminator threshold set to 40 mV and the terminators were in. DAQ rate is 500 Hz. Pu-Be source in, moved away from the detector by 26".
Region 1 Region 2
Fake stop Second hit
  • Run 4553 had a discriminator threshold set to 20 mV and the terminators were in. DAQ rate is 1850 Hz. Pu-Be source in, moved away from the detector by 26".
Region 1 Region 2
Fake stop Second hit
  • Run 4554 had a discriminator threshold set to 20 mV and the terminators were in. DAQ rate is 625 Hz. Pu-Be source in, moved away from the detector by 52".
Region 1 Region 2
Fake stop Second hit
  • Run 4555 had a discriminator threshold set to 20 mV and the terminators were out. DAQ rate is 650 Hz. Pu-Be source in, moved away from the detector by 52".
Region 1 Region 2
Fake stop Second hit
  • Run 4556 had a discriminator threshold set to 10 mV and the terminators were out. DAQ rate is 1400 Hz. Pu-Be source in, moved away from the detector by 52".
Region 1 Region 2
Fake stop Second hit
  • Run 4557 had a discriminator threshold set to 10 mV and the terminators were in. DAQ rate is 1400 Hz. Pu-B source in, moved away from the detector by 52".
Region 1 Region 2
Fake stop Second hit


PMT detector #4 HV at -1400 Volts
  • Run 4558 had a discriminator threshold set to 10 mV and the terminators were in. DAQ rate is 4670 Hz. Pu-Be source in, moved away from the detector by 52".
Region 1 Region 2
Fake stop Second hit
  • Run 4559 had a discriminator threshold set to 10 mV and the terminators were in. DAQ rate is 2700 Hz. Pu-Be source in, moved away from the detector by 91".
Region 1 Region 2
Fake stop Second hit
  • Run 4560 had a discriminator threshold set to 10 mV and the terminators were out. DAQ rate is 2700 Hz. Pu-Be source in, moved away from the detector by 91".
Region 1 Region 2
Fake stop Second hit
  • Run 4561 had a discriminator threshold set to 10 mV and the terminators were out. DAQ rate is 200 Hz. Cosmic rays test.
Region 1 Region 2
Fake stop Second hit
1.8 g Pu-Be source should be used next time