Difference between revisions of "04-22-2015"
Jump to navigation
Jump to search
(4 intermediate revisions by the same user not shown) | |||
Line 1: | Line 1: | ||
− | ==Pu-Be source test== | + | ===Pu-Be source test=== |
PMT detector #4 HV at -1000 Volts | PMT detector #4 HV at -1000 Volts | ||
Line 5: | Line 5: | ||
Pu-Be 18 gram source# M1197 | Pu-Be 18 gram source# M1197 | ||
− | Run 4546 had a discriminator threshold set to 80 mV and the terminators were out. Cosmic rays test. | + | * Run 4546 had a discriminator threshold set to 80 mV and the terminators were out. Cosmic rays test. |
{| cellpadding="2" style="border: 1px solid darkgray;" | {| cellpadding="2" style="border: 1px solid darkgray;" | ||
Line 15: | Line 15: | ||
|} | |} | ||
− | Run 4547 had a discriminator threshold set to 80 mV and the terminators were out. DAQ rate is 100 Hz. Pu-Be source in. | + | * Run 4547 had a discriminator threshold set to 80 mV and the terminators were out. DAQ rate is 100 Hz. Pu-Be source in. |
{| cellpadding="2" style="border: 1px solid darkgray;" | {| cellpadding="2" style="border: 1px solid darkgray;" | ||
Line 25: | Line 25: | ||
|} | |} | ||
− | Run 4548 had a discriminator threshold set to 80 mV and the terminators were in. DAQ rate is 100 Hz. Pu-Be source in. | + | * Run 4548 had a discriminator threshold set to 80 mV and the terminators were in. DAQ rate is 100 Hz. Pu-Be source in. |
{| cellpadding="2" style="border: 1px solid darkgray;" | {| cellpadding="2" style="border: 1px solid darkgray;" | ||
Line 35: | Line 35: | ||
|} | |} | ||
− | Run 4549 had a discriminator threshold set to 40 mV and the terminators were in. DAQ rate is 2800 Hz. Pu- | + | * Run 4549 had a discriminator threshold set to 40 mV and the terminators were in. DAQ rate is 2800 Hz. Pu-Be source in. |
{| cellpadding="2" style="border: 1px solid darkgray;" | {| cellpadding="2" style="border: 1px solid darkgray;" | ||
Line 45: | Line 45: | ||
|} | |} | ||
− | Run 4550 had a discriminator threshold set to 40 mV and the terminators were out. DAQ rate is 2800 Hz. Pu- | + | * Run 4550 had a discriminator threshold set to 40 mV and the terminators were out. DAQ rate is 2800 Hz. Pu-Be source in. |
{| cellpadding="2" style="border: 1px solid darkgray;" | {| cellpadding="2" style="border: 1px solid darkgray;" | ||
Line 55: | Line 55: | ||
|} | |} | ||
− | Run 4551 had a discriminator threshold set to 40 mV and the terminators were out. DAQ rate is 500 Hz. Pu- | + | * Run 4551 had a discriminator threshold set to 40 mV and the terminators were out. DAQ rate is 500 Hz. Pu-Be source in, moved away from the detector by 26". |
{| cellpadding="2" style="border: 1px solid darkgray;" | {| cellpadding="2" style="border: 1px solid darkgray;" | ||
Line 65: | Line 65: | ||
|} | |} | ||
− | Run 4552 had a discriminator threshold set to 40 mV and the terminators were in. DAQ rate is 500 Hz. Pu- | + | * Run 4552 had a discriminator threshold set to 40 mV and the terminators were in. DAQ rate is 500 Hz. Pu-Be source in, moved away from the detector by 26". |
{| cellpadding="2" style="border: 1px solid darkgray;" | {| cellpadding="2" style="border: 1px solid darkgray;" | ||
Line 75: | Line 75: | ||
|} | |} | ||
− | Run 4553 had a discriminator threshold set to 20 mV and the terminators were in. DAQ rate is 1850 Hz. Pu-Be source in, moved away from the detector by 26". | + | * Run 4553 had a discriminator threshold set to 20 mV and the terminators were in. DAQ rate is 1850 Hz. Pu-Be source in, moved away from the detector by 26". |
{| cellpadding="2" style="border: 1px solid darkgray;" | {| cellpadding="2" style="border: 1px solid darkgray;" | ||
Line 85: | Line 85: | ||
|} | |} | ||
− | Run 4554 had a discriminator threshold set to 20 mV and the terminators were in. DAQ rate is 625 Hz. Pu-Be source in, moved away from the detector by 52". | + | * Run 4554 had a discriminator threshold set to 20 mV and the terminators were in. DAQ rate is 625 Hz. Pu-Be source in, moved away from the detector by 52". |
{| cellpadding="2" style="border: 1px solid darkgray;" | {| cellpadding="2" style="border: 1px solid darkgray;" | ||
Line 95: | Line 95: | ||
|} | |} | ||
− | Run 4555 had a discriminator threshold set to 20 mV and the terminators were out. DAQ rate is 650 Hz. Pu-Be source in, moved away from the detector by 52". | + | * Run 4555 had a discriminator threshold set to 20 mV and the terminators were out. DAQ rate is 650 Hz. Pu-Be source in, moved away from the detector by 52". |
− | [[File:r4555_fake_stop.png | | + | {| cellpadding="2" style="border: 1px solid darkgray;" |
+ | |- border="0" | ||
+ | | [[File:r4555_fake_stop.png|340px|Region 1]] | ||
+ | | [[File:r4555_det4_up.png|340px|Region 2]] | ||
+ | |- align="center" | ||
+ | | Fake stop|| Second hit | ||
+ | |} | ||
− | + | * Run 4556 had a discriminator threshold set to 10 mV and the terminators were out. DAQ rate is 1400 Hz. Pu-Be source in, moved away from the detector by 52". | |
+ | {| cellpadding="2" style="border: 1px solid darkgray;" | ||
+ | |- border="0" | ||
+ | | [[File:r4556_fake_stop.png|340px|Region 1]] | ||
+ | | [[File:r4556_det4_up.png|340px|Region 2]] | ||
+ | |- align="center" | ||
+ | | Fake stop|| Second hit | ||
+ | |} | ||
− | Run | + | * Run 4557 had a discriminator threshold set to 10 mV and the terminators were in. DAQ rate is 1400 Hz. Pu-B source in, moved away from the detector by 52". |
− | [[File: | + | {| cellpadding="2" style="border: 1px solid darkgray;" |
+ | |- border="0" | ||
+ | | [[File:r4557_fake_stop.png|340px|Region 1]] | ||
+ | | [[File:r4557_det4_up.png|340px|Region 2]] | ||
+ | |- align="center" | ||
+ | | Fake stop|| Second hit | ||
+ | |} | ||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
PMT detector #4 HV at -1400 Volts | PMT detector #4 HV at -1400 Volts | ||
− | Run 4558 had a discriminator threshold set to 10 mV and the terminators were in. DAQ rate is 4670 Hz. Pu-Be source in, moved away from the detector by 52". | + | * Run 4558 had a discriminator threshold set to 10 mV and the terminators were in. DAQ rate is 4670 Hz. Pu-Be source in, moved away from the detector by 52". |
− | [[File:r4558_fake_stop.png | | + | {| cellpadding="2" style="border: 1px solid darkgray;" |
+ | |- border="0" | ||
+ | | [[File:r4558_fake_stop.png|340px|Region 1]] | ||
+ | | [[File:r4558_det4_up.png|340px|Region 2]] | ||
+ | |- align="center" | ||
+ | | Fake stop|| Second hit | ||
+ | |} | ||
− | + | * Run 4559 had a discriminator threshold set to 10 mV and the terminators were in. DAQ rate is 2700 Hz. Pu-Be source in, moved away from the detector by 91". | |
− | + | {| cellpadding="2" style="border: 1px solid darkgray;" | |
+ | |- border="0" | ||
+ | | [[File:r4559_fake_stop.png|340px|Region 1]] | ||
+ | | [[File:r4559_det4_up.png|340px|Region 2]] | ||
+ | |- align="center" | ||
+ | | Fake stop|| Second hit | ||
+ | |} | ||
− | + | * Run 4560 had a discriminator threshold set to 10 mV and the terminators were out. DAQ rate is 2700 Hz. Pu-Be source in, moved away from the detector by 91". | |
− | + | {| cellpadding="2" style="border: 1px solid darkgray;" | |
− | + | |- border="0" | |
− | + | | [[File:r4560_fake_stop.png|340px|Region 1]] | |
− | + | | [[File:r4560_det4_up.png|340px|Region 2]] | |
− | [[File:r4560_fake_stop.png | | + | |- align="center" |
− | + | | Fake stop|| Second hit | |
− | [[File:r4560_det4_up.png | | + | |} |
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
+ | * Run 4561 had a discriminator threshold set to 10 mV and the terminators were out. DAQ rate is 200 Hz. Cosmic rays test. | ||
+ | {| cellpadding="2" style="border: 1px solid darkgray;" | ||
+ | |- border="0" | ||
+ | | [[File:r4561_fake_stop.png|340px|Region 1]] | ||
+ | | [[File:r4561_det4_up.png|340px|Region 2]] | ||
+ | |- align="center" | ||
+ | | Fake stop|| Second hit | ||
+ | |} | ||
;1.8 g Pu-Be source should be used next time | ;1.8 g Pu-Be source should be used next time |
Latest revision as of 23:53, 25 April 2015
Pu-Be source test
PMT detector #4 HV at -1000 Volts
Pu-Be 18 gram source# M1197
- Run 4546 had a discriminator threshold set to 80 mV and the terminators were out. Cosmic rays test.
Fake stop | Second hit |
- Run 4547 had a discriminator threshold set to 80 mV and the terminators were out. DAQ rate is 100 Hz. Pu-Be source in.
Fake stop | Second hit |
- Run 4548 had a discriminator threshold set to 80 mV and the terminators were in. DAQ rate is 100 Hz. Pu-Be source in.
Fake stop | Second hit |
- Run 4549 had a discriminator threshold set to 40 mV and the terminators were in. DAQ rate is 2800 Hz. Pu-Be source in.
Fake stop | Second hit |
- Run 4550 had a discriminator threshold set to 40 mV and the terminators were out. DAQ rate is 2800 Hz. Pu-Be source in.
Fake stop | Second hit |
- Run 4551 had a discriminator threshold set to 40 mV and the terminators were out. DAQ rate is 500 Hz. Pu-Be source in, moved away from the detector by 26".
Fake stop | Second hit |
- Run 4552 had a discriminator threshold set to 40 mV and the terminators were in. DAQ rate is 500 Hz. Pu-Be source in, moved away from the detector by 26".
Fake stop | Second hit |
- Run 4553 had a discriminator threshold set to 20 mV and the terminators were in. DAQ rate is 1850 Hz. Pu-Be source in, moved away from the detector by 26".
Fake stop | Second hit |
- Run 4554 had a discriminator threshold set to 20 mV and the terminators were in. DAQ rate is 625 Hz. Pu-Be source in, moved away from the detector by 52".
Fake stop | Second hit |
- Run 4555 had a discriminator threshold set to 20 mV and the terminators were out. DAQ rate is 650 Hz. Pu-Be source in, moved away from the detector by 52".
Fake stop | Second hit |
- Run 4556 had a discriminator threshold set to 10 mV and the terminators were out. DAQ rate is 1400 Hz. Pu-Be source in, moved away from the detector by 52".
Fake stop | Second hit |
- Run 4557 had a discriminator threshold set to 10 mV and the terminators were in. DAQ rate is 1400 Hz. Pu-B source in, moved away from the detector by 52".
Fake stop | Second hit |
PMT detector #4 HV at -1400 Volts
- Run 4558 had a discriminator threshold set to 10 mV and the terminators were in. DAQ rate is 4670 Hz. Pu-Be source in, moved away from the detector by 52".
Fake stop | Second hit |
- Run 4559 had a discriminator threshold set to 10 mV and the terminators were in. DAQ rate is 2700 Hz. Pu-Be source in, moved away from the detector by 91".
Fake stop | Second hit |
- Run 4560 had a discriminator threshold set to 10 mV and the terminators were out. DAQ rate is 2700 Hz. Pu-Be source in, moved away from the detector by 91".
Fake stop | Second hit |
- Run 4561 had a discriminator threshold set to 10 mV and the terminators were out. DAQ rate is 200 Hz. Cosmic rays test.
Fake stop | Second hit |
- 1.8 g Pu-Be source should be used next time