Difference between revisions of "UnderGraduate Neutron Correlations Lab"

From New IAC Wiki
Jump to navigation Jump to search
Line 76: Line 76:
 
|}
 
|}
  
 +
=data 5/3/21=
 +
 +
Set up:
 +
* Lead: Detector surrounded on 4 sides by 1" lead.  Front end 4" of lead.  Back end exposed. 
 +
* Poly: 40cm of green poly between lead and source.  No poly elsewhere.
 +
 +
 +
Source in 1428
 +
Source out 1318
 +
Source out 1361
 +
Source out 1393
  
  
  
 
[[https://wiki.iac.isu.edu/index.php/Main_Page]]
 
[[https://wiki.iac.isu.edu/index.php/Main_Page]]

Revision as of 19:09, 3 May 2021

[[1]]


NaI detector setup


Creating ROOT data files


source ~/CODA/setup

/home/daq/CODA/CODAreader/ROOT_V5.34/PAA/evio2nt -fr6815.dat > /dev/null

Analyzing ROOT data file

root -l r8735.root

to draw a histogram for Peak ADC channel 4

PAA->Draw("PADC.PADC785N[4]>>(4096,0,4096)")

To scale the histogram using an energy calibration of 0.45 keV/channel

PAA->Draw("PADC.PADC785N[4]*0.45>>(4096,0,4096)")

To save thehistogram under the name Cs137

PAA->Draw("PADC.PADC785N[4]*0.45>>Cs137(4096,0,4096)")


To normalize the histogram by a run time of 2600 seconds

1/3600=0.0002778

Cs137->Scale(0.0002778)
Cs137->Draw();

run 6816 background (~300Hz) 1/47 = 0.021277

run 6817 signal (~600 Hz) 1/75 =0.013333


NaI Cs137 4-22-21.png

https://wiki.iac.isu.edu/index.php/File:NaI_Cs137_4-22-21.png

data 4/26/21

Run Number configuration
6819 Cs-137 source at location of Cf-252, no Cf-252 or Cd target in apparatus
6820 start time: 16:04:33 04/26/2021 Cs-137 out, Cd in , Cf-252 out. NaI at closest possible. stop time: 16:08:34 (241 seconds)
6821 start time: 16:11:49 04/26/2021 Cs-137 out, Cd in , Cf-252 in , Poly thermalizer 1/2" in. NaI at closest possible. stop time: 16:22:52 (663 sec)
6822 start time: 16:23:54 04/26/2021 Cs-137 out, Cd in , Cf-252 out , Poly thermalizer 1/2" in. NaI at closest possible. stop time:16:34:38 (673 sec)
6823 start time: 16:35:44 04/26/2021 Cs-137 out, Cd out , Cf-252 out , Poly thermalizer 1/2" in. NaI at closest possible. stop time:16:46
6824 start time: 16:35:44 04/26/2021 Cf-252 in , Poly thermalizer 1/2" in. ,Cd in sandwhich , NaI at closest possible. stop time:16:46
6825 start time: 16:53:55 04/26/2021 Cf-252 in 6 in away, 1" lead, , Poly thermalizer 1/2" in. ,Cd in sandwhich , NaI at closest possible. stop time:16:54:58
6826 start time: 16:57:39 04/26/2021 Cf-252 in 6 in away, 2" lead, , Poly thermalizer 1/2" in. ,Cd in sandwhich , NaI at closest possible. stop time:16:58:49
6827 start time: 16:59:13 04/26/2021 Cf-252 in 6 in away, 2" lead, , Poly thermalizer 1/2" in. ,Cd OUT sandwhich , NaI at closest possible. stop time:17:00:15

data 5/3/21

Set up:

  • Lead: Detector surrounded on 4 sides by 1" lead. Front end 4" of lead. Back end exposed.
  • Poly: 40cm of green poly between lead and source. No poly elsewhere.


Source in 1428 Source out 1318 Source out 1361 Source out 1393


[[2]]