Difference between revisions of "S6 Testing and Performance Summary"

From New IAC Wiki
Jump to navigation Jump to search
Line 1: Line 1:
 
=Testing results=
 
=Testing results=
  
== Wire Signal Testing ==
+
== Signal Testing for the Wires ==
  
 
Signal testing performed to all S6's wires for cosmic rays, the testing was simultaneously with the wire burning process. In general the signal is detected when the voltage on the sense wire at least was 1200V, guard's 535V, and field's -750V.  
 
Signal testing performed to all S6's wires for cosmic rays, the testing was simultaneously with the wire burning process. In general the signal is detected when the voltage on the sense wire at least was 1200V, guard's 535V, and field's -750V.  
Line 26: Line 26:
 
|SL1 ||STB7-G1202272 ||  6,83  || 23,33,34,69,78,92 ||1,2,82  || 49-64 N/A
 
|SL1 ||STB7-G1202272 ||  6,83  || 23,33,34,69,78,92 ||1,2,82  || 49-64 N/A
 
|-
 
|-
 
+
|SL2 ||STB1-K1202516 || 4  || N/A || 1,2,90 ||
 +
|-
 +
|SL2 ||STB7-K1201717||  N/A  || 10,19,21,43 ||  N/A  ||
  
 
|}
 
|}

Revision as of 02:01, 5 August 2013

Testing results

Signal Testing for the Wires

Signal testing performed to all S6's wires for cosmic rays, the testing was simultaneously with the wire burning process. In general the signal is detected when the voltage on the sense wire at least was 1200V, guard's 535V, and field's -750V.

The results are shown in the following table:

Super Layer STP Board Wires without any Signal Hot wires High Noise Wires Notes
SL1 STB1-G1202270 4 24,38,66,87,93 1,17 Sense wire voltage 1380V
SL1 STB2-G1202266 1,49,82 38,40,51,65 N/A Sense wire voltage 1360V or 1500V
SL1 STB3-G1202273 N/A 1,2,39,78 N/A
SL1 STB4-G1202275 N/A 61 61
SL1 STB5-G1202271 N/A 8,60,88,89 8,17,18,33,34,49,50,65,66,81,82,
SL1 STB6-G1202274 N/A N/A 1,2
SL1 STB7-G1202272 6,83 23,33,34,69,78,92 1,2,82 49-64 N/A
SL2 STB1-K1202516 4 N/A 1,2,90
SL2 STB7-K1201717 N/A 10,19,21,43 N/A