|06/17/13 || 12:16 ||  536 || 1520 ||  -750 || 10.7 || 10.8 ||  
 
|06/17/13 || 12:16 ||  536 || 1520 ||  -750 || 10.7 || 10.8 ||  
 
|-
 
|-
|}
+
|06/17/13 || 12:23 ||  536 || 1530 ||  -750 || 14.3  || 14.4 ||
 +
|-
 +
|-
 +
|06/17/13 || 12:24 ||  536 || 1530 ||  -750 || 20.1  || 21.4 || field power supply tripped
 +
|-|}

Revision as of 18:30, 17 June 2013

Date Time (24h) Guard (V) Sense (V) Field (V) Sense ([math]\mu[/math]A) Field ([math]\mu[/math]A) comments
05/28/13 14:08 100 100 -100 0.02 0.00 Confirmed with initial setting of 30/40/50V that there are no crossed connections between wire types and that voltage supply is connected correctly.
05/28/13 16:34 536 1500 -750 45.5 45.5 Someone upped the voltage to full the full operating level without recording it. This is the current reading right now.
05/28/13 17:56 536 1500 -750 41.5 41.6
05/29/13 07:56 536 1500 -750 22.5 22.6
05/29/13 15:02 536 1500 -750 17.9 18.0
05/29/13 18:07 536 1500 -750 16.7 16.8
05/30/13 07:55 536 1500 -750 10.6 10.7
05/30/13 12:51 536 1500 -750 9.5 9.6
05/30/13 17:23 536 1500 -750 8.7 8.8
06/03/13 08:23 536 1500 -750 1.7 1.8
06/04/13 07:42 536 1500 -750 1.4 1.5
06/05/13 12:01 536 1500 -750 1.8 1.9 had to turn off HV to organize cables
06/05/13 17:56 536 1500 -750 1.8 1.9
06/06/13 07:56 536 1500 -750 1.8 1.9 seem to be stuck at this current draw?
06/10/13 17:29 536 1500 -750 1.5 1.6 I think we ran out of gas on the weekend, the windows were deflated a bit but now they deflected outward
06/11/13 07:50 536 1500 -750 3.2 3.3 current doubles? Humidity dropped from 40 to 25%.
06/12/13 13:39 536 1500 -750 3.6 3.7 Humidity rose from 25 to 40%.
06/13/13 11:00 536 1500 -750 4.5 4.6 Humidity is 40 +/- 5 %.
06/14/13 08:56 536 1500 -750 3.5 3.6
06/17/13 07:58 536 1500 -750 3.6 3.7

Why is the current increasing?

is the current going to ground ?

Try raising the sense to 1750 and lowering the field to 500 (same voltage difference), and see if the sense current still equals the field current - want to know if the current is inside the chamber from wire to wire or if it's a current to ground, for example ...

If current is from field to sense:

if it is current from field to sense the current should increase exponentially with voltage, doubling with 100V increase; i.e. 2250 -> 2350. You could check to see if this is the case. If so, you could then try to locate where the problem is: which wire(s) are responsible?

Increasing Sense 10 volts at a time

Date Time (24h) Guard (V) Sense (V) Field (V) Sense ([math]\mu[/math]A) Field ([math]\mu[/math]A) comments
06/17/13 12:20 536 1510 -750 8 8.1
06/17/13 12:16 536 1520 -750 10.7 10.8
06/17/13 12:23 536 1530 -750 14.3 14.4
06/17/13 12:24 536 1530 -750 20.1 21.4 field power supply tripped