== 100 mV signal ==
 +
 +
Reducing the attenuation, a 100 mV signal was sent into the ADC.
 +
 +
[[Image:ADC_V792_Calib_S100mV_12-11-07.jpg | 200 px]]
 +
 +
The data was then collected using coda and recorded as run #190.  A histogram of channel #1 is shown below.
 +
 +
[[Image:ADC_V792_Calib_100mV_Output_12-11-07_Histo.gif | 200 px]]
 +
[[Image:ADC_V792_Calib_100mV_Ped_12-11-07_Histo.gif | 200 px]]
 +
[[Image:ADC_V792_Calib_100mV_S_12-11-07_Histo.gif | 200 px]]

Revision as of 17:27, 11 December 2007

I injected a signal into the Drift chamber preamps and read it out using the CAEN V792 ADC.

12/11/07

7 mV Signal

Here is a scope picture of the signal coming out of the preamp and going into the ADC (Blue) as well as the integration gate used. The signal is about 7 mv as shown below. There is a 2 mv peak to peak noise level which has structure.

ADC V792 Calib a 12-11-07.jpg ADC V792 Calib a noise 12-11-07.jpg ADC V792 Calib a noise 2 12-11-07.jpg

The data was then collected using coda and recorded as run #189. A histogram of channel #1 is shown below.


ADC V792 Calib 7mV Output 12-11-07 Histo.gif ADC V792 Calib 7mV Ped 12-11-07 Histo.gif ADC V792 Calib 7mV S 12-11-07 Histo.gif

20 mV signal

Reducing the attenuation, a 20 mV signal was sent into the ADC.

ADC V792 Calib S20mV 12-11-07.jpg

The data was then collected using coda and recorded as run #190. A histogram of channel #1 is shown below.

ADC V792 Calib 20mV Output 12-11-07 Histo.gif ADC V792 Calib 20mV Ped 12-11-07 Histo.gif ADC V792 Calib 20mV S 12-11-07 Histo.gif


100 mV signal

Reducing the attenuation, a 100 mV signal was sent into the ADC.

ADC V792 Calib S100mV 12-11-07.jpg

The data was then collected using coda and recorded as run #190. A histogram of channel #1 is shown below.

ADC V792 Calib 100mV Output 12-11-07 Histo.gif ADC V792 Calib 100mV Ped 12-11-07 Histo.gif ADC V792 Calib 100mV S 12-11-07 Histo.gif