Difference between revisions of "MiniPrototype CAEN V792 Calibration"
Line 30: | Line 30: | ||
[[Image:ADC_V792_Calib_20mV_Ped_12-11-07_Histo.gif | 200 px]] | [[Image:ADC_V792_Calib_20mV_Ped_12-11-07_Histo.gif | 200 px]] | ||
[[Image:ADC_V792_Calib_20mV_S_12-11-07_Histo.gif | 200 px]] | [[Image:ADC_V792_Calib_20mV_S_12-11-07_Histo.gif | 200 px]] | ||
+ | |||
+ | |||
+ | == 100 mV signal == | ||
+ | |||
+ | Reducing the attenuation, a 100 mV signal was sent into the ADC. | ||
+ | |||
+ | [[Image:ADC_V792_Calib_S100mV_12-11-07.jpg | 200 px]] | ||
+ | |||
+ | The data was then collected using coda and recorded as run #190. A histogram of channel #1 is shown below. | ||
+ | |||
+ | [[Image:ADC_V792_Calib_100mV_Output_12-11-07_Histo.gif | 200 px]] | ||
+ | [[Image:ADC_V792_Calib_100mV_Ped_12-11-07_Histo.gif | 200 px]] | ||
+ | [[Image:ADC_V792_Calib_100mV_S_12-11-07_Histo.gif | 200 px]] |
Revision as of 17:27, 11 December 2007
I injected a signal into the Drift chamber preamps and read it out using the CAEN V792 ADC.
12/11/07
7 mV Signal
Here is a scope picture of the signal coming out of the preamp and going into the ADC (Blue) as well as the integration gate used. The signal is about 7 mv as shown below. There is a 2 mv peak to peak noise level which has structure.
The data was then collected using coda and recorded as run #189. A histogram of channel #1 is shown below.
20 mV signal
Reducing the attenuation, a 20 mV signal was sent into the ADC.
The data was then collected using coda and recorded as run #190. A histogram of channel #1 is shown below.
100 mV signal
Reducing the attenuation, a 100 mV signal was sent into the ADC.