Initial Fits

These fits are listed dose (from Landauer) vs measured dose (from OSL reader).

This fit was calculated using only the OSLs provided by Landauer for calibrating the OSL reader. [math] {\chi}^2 = 359.191[/math] [math] \text{# of Degrees of Freedom} = 13[/math]
This fit was calculated using the OSLs provided by Landauer and 6 Nanodot OSLs dosed by using the 137Cs source. [math] {\chi}^2 = 686.664[/math] [math] \text{# of Degrees of Freedom} = 19[/math]

Calculated Dose vs PMT Counts

These fits were created using OSLs I irradiated using the 20Ci [math] ^{137}Cs [/math] Source. I calculated the dose the OSLs should have on them using the dose calculations on the previous page. The PMT counts came straight from the OSL reader.

Number of degrees of freedom is data points - fit parameters

[math] {\chi}^2 = 0.31801[/math], [math] \text{# of Degrees of Freedom} = 2[/math]













Reproducibility with 137Cs