LB DetLimits Thesis

From New IAC Wiki
Jump to navigation Jump to search

50% Excluded

For this analysis, begin by using the first measurement of the Se-75 line and using the standard exponential decay equation to correct it and its error back to the time of beam off. Below is a table of the Se-75 (120d) corrected measurements as well as the front nickel foils and the Mn-54 data for each measurement. Note the Mn-54 analysis was weighted by the mass of the soil.

Sample Activated (total beam time) Counted Fully Thesis Corrected Rate @ 1st Measurement (Hz) [math] A_{Beam Off} [/math]
10% Se/Soil Mixture 5/23/17 and 5/24/17 (83 min) 07/18/17 (16:25:44) 3109.68 [math] \pm [/math] 95.39 4280.08 [math] \pm [/math] 131.29
10% Se/Soil Pure Witness selenium 5/23/17 and 5/24/17 (83 min) 8/1/17 (16:05:51) 4161.29 [math] \pm [/math] 127.63 6210.50 [math] \pm [/math] 190.48
10% Se/Soil Front Inner Ni Foil 5/23/17 (1h) 5/23/17 (15:52:21) 385955 [math] \pm [/math] 23488.8 425064.36 [math] \pm [/math] 25868.95
10% Se/Soil Mixture Mn-54 5/23/17 (1h) 07/18/17 (16:25:44) 15.63 [math] \pm [/math] 0.56 17.67 [math] \pm [/math] 0.63
1% Sample Se/Soil Mixture 5/23/17 (1h) 6/23/17 (16:39:03) 2233.33 [math] \pm [/math] 68.52 2675.24 [math] \pm [/math] 82.08
1% Sample Se/Soil Pure Witness selenium 5/23/17 (1h) 7/20/17 (16:39:46) 3840.86 [math] \pm [/math] 117.81 5379.43 [math] \pm [/math] 165.00
1% Sample Se/Soil Front Inner Ni Foil 5/23/17 (1h) 5/23/17 (17:59:39) 261461 [math] \pm [/math] 16045.7 293170.78 [math] \pm [/math] 17991.71
1% Sample Se/Soil Mixture Mn-54 5/23/17 (1h) 6/23/17 (16:39:03) 9.38 [math] \pm [/math] 0.39 10.05 [math] \pm [/math] 0.42
0.1% Sample Se/Soil Mixture 5/23/17 (30 min) 6/19/17 (18:25:18) 849.46 [math] \pm [/math] 26.07 993.59 [math] \pm [/math] 30.49
0.1% Sample Se/Soil Pure Witness Selenium 5/23/17 (30 min) 07/25/17 (16:32:52) 834.41 [math] \pm [/math] 25.61 1201.58 [math] \pm [/math] 36.88
0.1% Sample Se/Soil Front Inner Ni Foil 5/23/17 (30 min) 5/24/17 (14:25:39 118412 [math] \pm [/math] 7280.06 197670.18 [math] \pm [/math] 12152.98
0.1% Sample Se/Soil Mn-54 5/23/17 (30 min) 6/19/17 (18:25:18) 9.38 [math] \pm [/math] 0.33 9.96 [math] \pm [/math] 0.35

Mn-54 Efficiency

A calibrated Mn-54 source was used to find the efficiency of an 834 keV line. The source was serial #J4-348, which had an activity of 9.882 [math] \mu Ci [/math] on 8/01/12, so the activity on 4/18/17 was

[math] A(4/18/17) = 9.882 \times e^{-2.57 \times 10^{-8} * 148694400} = 0.22 \mu Ci [/math]

Now converting to Hz gives

[math]\left (0.999760 \right )\left ( 0.22 \times 10^{-6} \mbox{Ci} \right) \left (\frac{ (3.7 \times 10^{10} \mbox{Hz}}{\mbox{Ci}} \right)= (8138.05 \pm 3\%) Hz [/math]

Source Energy (keV) Position Expected Rate (Hz) ROOT Window (keV) HpGe Rate (Hz) HpGe Detector A Efficiency (%)
Mn-54 (J4-348) 834 10cm (Det A) 8138.05 [math] \pm [/math] 3\% [829,839] 25.87 [math] \pm [/math] 0.28 0.32 [math] \pm [/math] 0.01

Activity Ratio Plots

Below are plots of the activity ratio of the Se/Soil mixture and specific activity of the pure selenium pellet as a reference material. The plots also have no restrictions on the equation of the line.

A Mix Div a Pure NormCorrected Se75 NoFitRestrict.png


Following the analysis in Nate's thesis, the initial concentration should be the intercept on the x-axis, which is

[math] \frac{0.054558 \pm 0.004056}{0.122961 \pm 0.006568} = (0.44 \pm 0.04)\% [/math]

This is much more physical than previous answers, but I was able to detect Se-75 at a 0.1% level. Let's try some other reference materials

Below is a plot where the activity ratio was taken using Ni-57 as a reference material. This was the front inner Nickel foil

A Mix Div Ni NormCorrected Se75 NoFitRestrict.png

This is not very physical because the graph implies that the initial concentration in the soil was 19.18%, but there were no Se-75 lines observed in a pure soil sample, so this cannot be true

Finally let's try Mn-54 as a reference material as it was internal to the sample.

A Mix Div a Mn NormCorrected Se75 NoFitRestrict.png

Now find the initial concentration by finding the x-intercept.

[math] \frac{15.320237 \pm 0.742753}{16.456772 \pm 0.694779} = (0.93 \pm 0.06)\% [/math]

This is still better than the nickel, but still not physical because of the argument presented in the section about the pure selenium ratio.

Let's focus on the ratio between the mixture and the pure selenium as well as the Mn-54 ratio. I was instructed to force the fit to have an x-intercept of less than 0.1%, so

[math] y = Bx + A [/math] when y = 0, then [math] x = \frac{-A}{B} \gt -0.1 [/math]