Difference between revisions of "GEM Cosmic V792 Measurements"

From New IAC Wiki
Jump to navigation Jump to search
Line 31: Line 31:
  
 
=12-11-08=
 
=12-11-08=
 +
 +
 +
 +
[[Image:TrigOut_electronics_setup_12-11-08.png|250px]][[Image:StripPulse_electronics_setup_12-11-08.png|250px]]<br>
  
 
==Electronics Setup==
 
==Electronics Setup==
  
[[Image:TrigOut_electronics_setup_12-11-08.png|250px]][[Image:StripPulse_electronics_setup_12-11-08.png|250px]]<br>
+
'''1).''' Threshold Values: Threshold=1.90, ULD=9.65  && LLD=2.10 <br>
  
'''1).''' Threshold Values: <br>
+
Run Number is r613<br>
  
 
[http://www.iac.isu.edu/mediawiki/index.php/GEMS Go Back]
 
[http://www.iac.isu.edu/mediawiki/index.php/GEMS Go Back]

Revision as of 00:21, 12 December 2008

11-07-08

The pulse coming from the strips is delayed and amplified using the Timing Filter Amplifier Model_474(X6). The amplified pulse is connected to the V792 QDC.
Run 550

Start Time: Nov 6 19:58:03; End Time: Nov 7 09:45:57

ADC measurement GEMDetector HV3800V ADC1 11-07-08.gifADC measurement GEMDetector HV3800V ADC1 11-07-08 Landau fit.gif

12-10-08

The signal below represents a measurement of the Trig out signal from a 10 cm x 10 cm ionization chamber with 3 GEM preamplifiers. The chamber gas a 90/10 % mixture of Arcon and CO2. The Drift HV was set to 3200 Volts and the GEM preamplifier HV distribution network was set to 2900 Volts. The V792 CAEN ADC channels have a gain of 100 fC/channel. The main signal peak appears at channel number 759 and represents an accumulation of [math]75.9 \pm 4[/math] pC of charge.


TrigOutSignal R607.gif

Electronics Settings:

TrigOut goes to Timing Filter Amplifier(Coarse Gain=X6, Fine GAin=7.5ns && Integrate=20ns) and after amplification to ADC
StripOut is used as a trigger and gate. The signal from the strip is amplified by the Timing Filter Amplifier (Coarse Gain=X20, Fine Gain=12.5ns && Integrate=500ns)
The Threshold on DIFF. CFD is set to 0.70 (ULD=9.9~10 && LLD=0.5)

12-10(11)-08

HV Settings: [math]V_{Drift}=3200[/math] Volts and [math]V_{GEM}=2900[/math] Volts.

Using again only TrigOut and the signal from the Strip.

Electronics Setup: done (might change CFD threshold)


12-11-08

TrigOut electronics setup 12-11-08.pngStripPulse electronics setup 12-11-08.png

Electronics Setup

1). Threshold Values: Threshold=1.90, ULD=9.65 && LLD=2.10

Run Number is r613

Go Back