A observable decrease in the energy resolution of a large HPGe was first seen after the irradiation of 5*10^7 n/cm^2<ref>P. H. Stelson, J. K. Dickens, S. Raman, and R. C. Trammell, “Deterioration of Large Ge(Li) Diodes Caused by Fast Neutrons,” Nuclear Instruments and Methods 98,481 (1972).</ref>, so 5*10^6 n/cm^2 is a conservative number to stay under.

The maximum neutron flux occurs right at the center of the detector, where the expression for flux is simply: [math]n_{rate}\frac{1}{4\pi d^2}[/math].

The number of days it would take to reach an integral flux of 5*10^6 n/cm^2, as a function of the distance from source to HPGe face is shown below. The graph assumes the activity of the Cf-252 source as of 01/2017, which is 19,066 n/s.

MaxHPGeCF252Time.png

Formula used in the graph above :