Difference between revisions of "Detector Description"

From New IAC Wiki
Jump to navigation Jump to search
Line 1: Line 1:
detector consists of :
+
The GEM based neutron detector consists of :
  
 +
1- A cathode; a square copper plate has a dimension of 10x10cm and has a distance of 1.6 cm from the top of the fist GEM card. The cathode design allows to raise the potential on its surface up to 5 kV (in air) without any discharge effect.
  
2- U-233 is used as a Neutron fissionable material which is coated on the top of metal plate in circular disk that has a diameter of 3cm. U-233 think film is fixed as a part of the detector cathode.
+
2- A neutron sensitive material; in our case U-233 have been used as a sensitive material and has relatively a high fission cross in the desired energy range, U-233 is coated on the top of a metal plate in circular disk that has a diameter of 3cm and fixed as a part of the detector cathode's surface.
 +
 
 +
3- GEM cards; the detector has three GEM cards in square plastic frames, they have  dimension of 10x10cm and are placed parallel to the cathode as shown in the figure below.
  
3- Cathode , square copper plate has a dimension of 10x10cm and has a ditance of 1.6 cm from the top of the fist GEM card.
 
  
3- GEM cards: the detector has three GEM in square plastic of frame and has dimension of 10x10cm, it is  placed parallel to the cathode.
 
  
  

Revision as of 12:16, 29 June 2013

The GEM based neutron detector consists of :

1- A cathode; a square copper plate has a dimension of 10x10cm and has a distance of 1.6 cm from the top of the fist GEM card. The cathode design allows to raise the potential on its surface up to 5 kV (in air) without any discharge effect.

2- A neutron sensitive material; in our case U-233 have been used as a sensitive material and has relatively a high fission cross in the desired energy range, U-233 is coated on the top of a metal plate in circular disk that has a diameter of 3cm and fixed as a part of the detector cathode's surface.

3- GEM cards; the detector has three GEM cards in square plastic frames, they have dimension of 10x10cm and are placed parallel to the cathode as shown in the figure below.



All the previous components exist in a plastic chamber, the chamber has been sealed to insulate the chamber has from the atmospheric gas components.

The chamber has a square window of kipton of dimension of 13x13cm.

The detector operates by providing the high voltage circuit by two channel high voltage power supply up to 3.1 kV for the cathode, and up to 2.8 kV for the GEM cards.

The circuit diagram is shown in the figure below:


Abdel GEM HV distrib 6-19-09.png


The circuit will make the potential difference between the top and the bottom of each card 300-350 V.