Difference between revisions of "DF He3 Tube Plateau Study"

From New IAC Wiki
Jump to navigation Jump to search
Line 25: Line 25:
  
 
[[File:DAF_He3_Tube_PlateauNoSourceLogarithmic.png|800px]]
 
[[File:DAF_He3_Tube_PlateauNoSourceLogarithmic.png|800px]]
 +
=Main Page=
 +
[[User:Friadavi]]

Revision as of 22:19, 19 June 2018

Hardware Settings

  • Ortec Bias Supply
    • Bias Signal Voltage = 1244V
  • Ortec Spec Amp
    • Fine Gain = 1
    • Coarse Gain = 10
    • Uni Shaping = Gaussian
    • Shaping Time = 2μs
    • PZ = auto
    • BLR = auto
    • Input = Norm -
  • Ortec Timer Counter
    • Base Coefficent = 6
    • Power = 3
    • Time Base = 0.1s
    • Dwell = max
  • CAEN LED
    • Threshold = variable
    • Pulse Width = 255 ns
  • CAEN Dual Timer
    • Coarse Pulse Width = 100 μs
    • Fine Pulse Width = 0.0

Graphs

DAF He3 Tube PlateauNoSourceLinear.png

DAF He3 Tube PlateauNoSourceLogarithmic.png

Main Page

User:Friadavi