=== compare previous measurements & adjust threshold & Width ===
 
=== compare previous measurements & adjust threshold & Width ===
 +
 +
  
 
{| border="1" cellpadding="20" cellspacing="0"
 
{| border="1" cellpadding="20" cellspacing="0"
Line 96: Line 98:
 
|390/808 = 1/2 ||10.5 MeV : 100 ns FWHM, 3.2 mA || UNshielded illyusha scintilltor, 6 mV threshold on CFD, 4 ns CFD output pulse width
 
|390/808 = 1/2 ||10.5 MeV : 100 ns FWHM, 3.2 mA || UNshielded illyusha scintilltor, 6 mV threshold on CFD, 4 ns CFD output pulse width
 
|}<br>
 
|}<br>
 +
 +
 +
beam current decreased from 5 to 3 mA during the measurements ; must measure beam current for each measurement.
  
 
===Add 3 Pb Bricks to accelerator side===
 
===Add 3 Pb Bricks to accelerator side===

Revision as of 18:44, 10 December 2012

RunPlan

BeamTune

Tune beam into zero line, max current, 10.5 MeV energy, 20 ns width, 300 Hz rep rate

For Reference, below is the the tune used on last day of July 2012 Positron run

https://wiki.iac.isu.edu/index.php/Beam_Test_Jul_27_2012#Tune

Measure Emittance

Ilyusha Background measurements

Previous Background tests during 2n correlation run

https://wiki.iac.isu.edu/index.php/2n_BKG_data

HV=1000 CFD = -12 mV


Previous run during neutron correlation experiment found 1 gamma/15 electron beam pulses

Tune beam into zero line, max current, 10.5 MeV energy, 20 ns width, 300 Hz rep rate

1.) Take a run with no shielding around Ilyusha

2.) Move Pb bricks around the scintillator to determine direction of background

3.) add shielding to accelerator

compare previous measurements & adjust threshold & Width

Rate (gamma/pulse) Beam COnfig (E,pulse width,peak current) Configuration
213/3746 = 1 /17 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 12 mV threshold on CFD, 255 ns CFD output pulse width (this width was sued in the 2n correlation exp)
213/2343 = 1/11 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 12 mV threshold on CFD, 4 ns CFD output pulse width
220/2374 = 1/11 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 12 mV threshold on CFD, 100 ns CFD output pulse width
53202/616 = nonsense 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 0 mV threshold on CFD, 4 ns CFD output pulse width
794/862 = 1/1 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 2 mV threshold on CFD, 4 ns CFD output pulse width
574/1213 = 1/2 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 4 mV threshold on CFD, 4 ns CFD output pulse width
247/906 = 1/4 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 6 mV threshold on CFD, 4 ns CFD output pulse width
662/3914 = 1/6 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 8 mV threshold on CFD, 4 ns CFD output pulse width
201/1328 = 1/6.6 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 10 mV threshold on CFD, 4 ns CFD output pulse width
220/1580 = 1/7.2 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 14 mV threshold on CFD, 4 ns CFD output pulse width
250/1176 = 1/5 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 12 mV threshold on CFD, 4 ns CFD output pulse width
252/937 = 1/4 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 10 mV threshold on CFD, 4 ns CFD output pulse width
221/917 = 1/4 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 12 mV threshold on CFD, 4 ns CFD output pulse width
239/634 = 1/3 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 8 mV threshold on CFD, 4 ns CFD output pulse width
241/718 = 1/3 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 10 mV threshold on CFD, 4 ns CFD output pulse width
222/819 = 1/4 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 12 mV threshold on CFD, 4 ns CFD output pulse width
241/1011 = 1/4 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 14 mV threshold on CFD, 4 ns CFD output pulse width
246/884 = 1/4 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 12 mV threshold on CFD, 4 ns CFD output pulse width
224/713 = 1/3 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 10 mV threshold on CFD, 4 ns CFD output pulse width
242/561 = 1/2 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 8 mV threshold on CFD, 4 ns CFD output pulse width
269/550 = 1/2 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 6 mV threshold on CFD, 4 ns CFD output pulse width
254/666 = 1/3 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 8 mV threshold on CFD, 4 ns CFD output pulse width
250/755 = 1/3 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 10 mV threshold on CFD, 4 ns CFD output pulse width
233/775 = 1/3 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 12 mV threshold on CFD, 4 ns CFD output pulse width
205/754 = 1/3 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 14 mV threshold on CFD, 4 ns CFD output pulse width
204/718 = 1/3 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 12 mV threshold on CFD, 4 ns CFD output pulse width
214/760 = 1/3 10.5 MeV : 100 ns FWHM, 5.4 mA UNshielded illyusha scintilltor, 10 mV threshold on CFD, 4 ns CFD output pulse width
390/808 = 1/2 10.5 MeV : 100 ns FWHM, 3.2 mA UNshielded illyusha scintilltor, 6 mV threshold on CFD, 4 ns CFD output pulse width



beam current decreased from 5 to 3 mA during the measurements ; must measure beam current for each measurement.

Add 3 Pb Bricks to accelerator side

Rate (gamma/pulse) Beam COnfig (E,pulse width,peak current) Configuration
10.5 MeV : 100 ns FWHM, 3.2 mA UNshielded illyusha scintilltor, 10 mV threshold on CFD, 4 ns CFD output pulse width


Electron/photon ratio

Compare magnet in and out using a scintillator at the exit window of the 90 degree line.