Difference between revisions of "10312013 TestRun"

From New IAC Wiki
Jump to navigation Jump to search
Line 36: Line 36:
 
| 6584 || same as run 6580 but using Mn-54  source # 129804, 12 min 58 sec,
 
| 6584 || same as run 6580 but using Mn-54  source # 129804, 12 min 58 sec,
 
|-
 
|-
| 6585 || same as run 6580 but using Na-22  source # 129741, min sec,
+
| 6585 || same as run 6580 but using Na-22  source # 129741, 3 min 30 sec,
 
|-
 
|-
 
|}
 
|}

Revision as of 19:40, 7 November 2013

Goals

Use NaI to measure Y-88 coincidence lines of 1836 and 898 keV

Runlist

Run Description
6559 Coincidence run with Na-22 1 uCi source # 129742, ADC7 = HpGe, ADC3 = NaI, TDC16 = HpGe, TDC19 = NaI, Na-22's 1.2 keV photon is in channel 3600 for the HpGe detector and channel 500 for the NaI; need to decrease gain on HpGe so Y-88's 1.8 keV photon is in the ADC range
6560 Changed HpGe spec amp integration from 0.5 usec shaping time to 0.25 and ran in singles mode
6561 Running in coincidence mode, rate changed from about 70 Hz to about 2 Hz
6562 Running in OR mode with activated Y-88 sample #130100-010 (highest intensity line is at 1460 keV)
6575 same conditions as run 6571 but with activated Y-88 sample #130100-010
6576 same conditions as run 6575 Coincidence run started Thu 31 Oct 2013 11:37:34 AM MDT (1383241054), ran for 598790 seconds = 6.93 days
6577 same conditions as run 6571 started on 11/7/13 OR mode
6578 In OR mode, decreased threshold from 40 mV to 10 mV
6579 same as run 6578 but only triggering on HpGe detector B , readback says its at 4520 volts, spec amp course gain = x20 , fine gain = 54, shaping time = .25 usec
6580 same as run 6579 but set pulser to 10 Hz to record run time in scaler channel 5
6581 same as run 6580
6582 same as run 6580 but using Cs-137 source # 129792, 19 min 36 sec,
6583 same as run 6580 but using Co-60 source # 124838, ,
6584 same as run 6580 but using Mn-54 source # 129804, 12 min 58 sec,
6585 same as run 6580 but using Na-22 source # 129741, 3 min 30 sec,

Energy Calibration

calibrate with Na-22 (511,1200), Co-60 (1.2, 1.3), Mn-54 (), Cs-137 (662 keV)

Calibration runs

Cd=109 has coincidence lines at 203 & 250 keV

Run Description
6567 Singles run with only NaI detector and Co-60 source
6568 singles runs with only HpGe detector and Co-60 source,October 31, 2013 4:54:03 PM MDT,1383238443-1383238497=54 seconds
6569 OR mode, both detectors, 3k DAQ trigger rate. Co-60 1 uCi source # 129739
6570 Running in OR mode with Mn-54 source # 129806 700 Hz,October 31, 2013 4:58:07 PM MDT,1383238687-1383239009=322 sec
6571 Running in OR mode with only NaI in trigger and Mn-54 source, increased threshold from 10 mV to 40 mV to count more high energy photons
6572 same as run 6571 but using Na-22 1 uCi source # 129742, October 31, 2013 5:19:47 PM MDT, 1383239987-1383240130= 143 second
6573 same as run 6571 but using Co-60 1 uCi source # 129739
6574 same as run 6571 but using and used Cs-137 1 uCi source #129793


Source Energy HpGe ch # (ADC 7) NaI ch # (ADC3)
Na-22 511 821.76 +/- 2.6 567.735 +/- 19.9
Cs-137 661.657 1025.7 +/- 2.1 707.185 +/- 20.5
Mn-54 834.848 1260.46 +/- 2.1 866.607 +/- 21.9
Co-60 1173.228 1719.56 +/- 2.4 1173.85 +/- 35
Na-22 1274.537 1856.88 +/- 2.2 1265.59 +/- 36
Co-60 1332.492 1935.41 +/- 2.4 1318.99 +/- 34


HpGe Fit

A=-94.978 +/- 0.719711 B=0.737543 +/- 0.000480387

ntuple->Draw("ADC7*0.737543 - 94.978 >>r6573(2200,0,4400)")

3rd order

-95.4193 +- 3.43312 0.73824 +- 0.0053233 -2.49548e-07 +- 1.89781e-06

Nov 7 2013

Source Energy HpGe ch # (ADC 7) NaI ch # (ADC3)
Na-22 511
Cs-137 661.657 1025.68 +/- 1.6
Mn-54 834.848 1260.62 +/- 2.1
Co-60 1173.228 1718.59 +/- 1.8
Na-22 1274.537
Co-60 1332.492 1934.27 +/- 1.9 1

Activity

The photon spectrum measured on 10/31/13 from the Y-88 sample #130100-010 and a 1uCi Co-60 1 uCi source # 129739

R6568N 11-4-2013.pngR6572N 11-4-2013.png


The NaI-22 measured rate was 366.8 events/sec while the Y-88 rate was about 30.6 events/sec.

A 1uCi Mn-54 source # 129806 had a rate of 229.6 events/sec.

R6570N 11-4-2013.png


A rough estimate is that the Y-88 activity is less than 0.1 uCi.

Efficiency Calibration

button source 1.5" from face of Det B

Run Source Energy HpGe Det B Rate (ADC 7)
Na-22 511
Cs-137 661.657
Mn-54 834.848
Co-60 1173.228
Na-22 1274.537
Co-60 1332.492 1


PAA_CoincidenceCounting#NaI_.26_HpGe_Coinc