Set up like that:

Detector1.png


The voltages, thresholds and widths are:

PMT voltage, V threshold, mV discr. width, ns
PS1 -1000 -80 50
PS1 -1000 -80 50
n's -1500 -200 10



Let's do the timing test. The distances are between PS1+PS2 and n's PMT. The time is time difference between PS1(PS2) and n's detector signals

Tbl timing total.png


As the distance becomes 10 cm and less the coincidence rate becomes very low. I have:

▫ 11 coincidence for 98 min (10 cm)
▫ 0 coincidence for 10 min


Below is the plot of measured time as function of distances:

▫ plot for first 10 events (measurements on May 13, 2001)
▫ plot for last 20 events (measurements on May 16, 2001)
▫ combined plot for total 30 events
Timing 1 10.png Timing 11 30.png
Timing 1 30.png