Difference between revisions of "05/12/2011"

From New IAC Wiki
Jump to navigation Jump to search
Line 45: Line 45:
  
 
Let's do the '''2n correlation test''' with CF252 source. The voltages and thresholds are the same.
 
Let's do the '''2n correlation test''' with CF252 source. The voltages and thresholds are the same.
 
 
First the set up is like:
 
First the set up is like:
 
:PS1 and PS2 are parallel to each other and are faced to the source.
 
:PS1 and PS2 are parallel to each other and are faced to the source.

Revision as of 20:54, 12 May 2011

Go Back


The powers are:

PS1 and PS2:  -1000 V (the old 2-channel TENNELEC TC952)
n's: -1100 V (the new 2-channel C.A.E.N N471)

The thresholds are:

PS1 and PS2:  -80 mV
n's: -200 mV

The discriminator widths are:

PS1: -20 ns
PS2: -10 ns
n's: -20 ns


Let's do the random coincidence cosmic ray test. PS1 and PS2 is about 30 cm away from each other. And let's look on the coincidence signal between PS1 and PS2:

total counts time counts/min counts/sec
PS1 [math]10247\pm101.2[/math] 70 min [math]146.4\pm1.44[/math] [math]2.44\pm0.02[/math]
PS1 [math]12969\pm113.9[/math] 70 min [math]185.3\pm1.63[/math] [math]3.09\pm0.03[/math]
PS1+PS2 [math]3\pm1.7[/math] 70 min [math]0.043\pm0.024[/math] [math]0.0007\pm0.0004[/math]


And below my scope pictures of this rear event:

Rnd coinc cosmic.png


Let's do the 2n correlation test with CF252 source. The voltages and thresholds are the same. First the set up is like:

PS1 and PS2 are parallel to each other and are faced to the source.
the CF252 source in the middle
the distance between PS1(PS2) and source is 6 cm

The total number of events (2n fission events + random coincident) are:

total counts time counts/min counts/sec
PS1 [math]255977\pm505.9[/math] 10 min [math]25597.7\pm50.6[/math] [math]426.63\pm0.84[/math]
PS1 [math]421917\pm649.5[/math] 10 min [math]42191.7\pm64.9[/math] [math]703.19\pm1.08[/math]
PS1+PS2 [math]1224\pm34.9[/math] 10 min [math]122.4\pm3.5[/math] [math]2.04\pm0.06[/math]


Second the set up is like:

PS1 and PS2 are perpendicular to each other and are faced to the source.
the CF252 source in the middle
the distance between PS1(PS2) and source is 6 cm

The total number of events (2n fission events + random coincident) are:

total counts time counts/min counts/sec
PS1 [math]250116\pm500.1[/math] 10 min [math]25011.6\pm50.0[/math] [math]416.86\pm0.83[/math]
PS1 [math]414774\pm644.0[/math] 10 min [math]64.4\pm64.9[/math] [math]691.29\pm1.07[/math]
PS1+PS2 [math]1359\pm36.9[/math] 10 min [math]135.9\pm3.7[/math] [math]2.26\pm0.06[/math]