Difference between revisions of "04-22-2015"

From New IAC Wiki
Jump to navigation Jump to search
 
(41 intermediate revisions by the same user not shown)
Line 1: Line 1:
==Cf source test==
+
===Pu-Be source test===
  
PMT detector #4 HV at -1000 Volts
+
PMT detector #4 HV at -1000 Volts
  
Pu-Be 18 gram source
+
Pu-Be 18 gram source# M1197
  
Run 4546 had a discriminator threshold set to 80 mV and the terminators were out. Cosmic rays test.
+
* Run 4546 had a discriminator threshold set to 80 mV and the terminators were out. Cosmic rays test.
  
 +
{| cellpadding="2" style="border: 1px solid darkgray;"
 +
|- border="0"
 +
| [[File:r4546_fake_stop.png|340px|Region 1]]
 +
| [[File:r4546_det4_up.png|340px|Region 2]]
 +
|- align="center"
 +
| Fake stop|| Second hit
 +
|}
  
Run 4547 had a discriminator threshold set to 80 mV and the terminators were out.  DAQ rate is 100 Hz. Pu-Be source in.
+
* Run 4547 had a discriminator threshold set to 80 mV and the terminators were out.  DAQ rate is 100 Hz. Pu-Be source in.
  
 +
{| cellpadding="2" style="border: 1px solid darkgray;"
 +
|- border="0"
 +
| [[File:r4547_fake_stop.png|340px|Region 1]]
 +
| [[File:r4547_det4_up.png|340px|Region 2]]
 +
|- align="center"
 +
| Fake stop|| Second hit
 +
|}
  
Run 4548 had a discriminator threshold set to 80 mV and the terminators were in.  DAQ rate is 100 Hz. Pu-Be source in.
+
* Run 4548 had a discriminator threshold set to 80 mV and the terminators were in.  DAQ rate is 100 Hz. Pu-Be source in.
  
Run 4549 had a discriminator threshold set to 40 mV and the terminators were in. DAQ rate is 2800 Hz. Pu-B source in.
+
{| cellpadding="2" style="border: 1px solid darkgray;"
 +
|- border="0"
 +
| [[File:r4548_fake_stop.png|340px|Region 1]]
 +
| [[File:r4548_det4_up.png|340px|Region 2]]
 +
|- align="center"
 +
| Fake stop|| Second hit
 +
|}
  
Run 4550 had a discriminator threshold set to 40 mV and the terminators were out.  DAQ rate is 2800 Hz. Pu-B source in.
+
* Run 4549 had a discriminator threshold set to 40 mV and the terminators were in.  DAQ rate is 2800 Hz. Pu-Be source in.
  
Run 4551 had a discriminator threshold set to 40 mV and the terminators were out. DAQ rate is 500 Hz. Pu-B source in, moved away from the detector by 26".
+
{| cellpadding="2" style="border: 1px solid darkgray;"
 +
|- border="0"
 +
| [[File:r4549_fake_stop.png|340px|Region 1]]
 +
| [[File:r4549_det4_up.png|340px|Region 2]]
 +
|- align="center"
 +
| Fake stop|| Second hit
 +
|}
  
Run 4552 had a discriminator threshold set to 40 mV and the terminators were in.  DAQ rate is 500 Hz. Pu-B source in, moved away from the detector by 26".
+
* Run 4550 had a discriminator threshold set to 40 mV and the terminators were out.  DAQ rate is 2800 Hz. Pu-Be source in.
  
Run 4553 had a discriminator threshold set to 20 mV and the terminators were in. DAQ rate is 1850 Hz. Pu-Be source in, moved away from the detector by 26".
+
{| cellpadding="2" style="border: 1px solid darkgray;"
 +
|- border="0"
 +
| [[File:r4550_fake_stop.png|340px|Region 1]]
 +
| [[File:r4550_det4_up.png|340px|Region 2]]
 +
|- align="center"
 +
| Fake stop|| Second hit
 +
|}
  
Run 4554 had a discriminator threshold set to 20 mV and the terminators were in.  DAQ rate is 625 Hz. Pu-Be source in, moved away from the detector by 52".
+
* Run 4551 had a discriminator threshold set to 40 mV and the terminators were out.  DAQ rate is 500 Hz. Pu-Be source in, moved away from the detector by 26".
  
Run 4555 had a discriminator threshold set to 20 mV and the terminators were out. DAQ rate is 650 Hz. Pu-Be source in, moved away from the detector by 52".
+
{| cellpadding="2" style="border: 1px solid darkgray;"
 +
|- border="0"
 +
| [[File:r4551_fake_stop.png|340px|Region 1]]
 +
| [[File:r4551_det4_up.png|340px|Region 2]]
 +
|- align="center"
 +
| Fake stop|| Second hit
 +
|}
  
Run 4556 had a discriminator threshold set to 10 mV and the terminators were out.  DAQ rate is 1400 Hz. Pu-Be source in, moved away from the detector by 52".
+
* Run 4552 had a discriminator threshold set to 40 mV and the terminators were in.  DAQ rate is 500 Hz. Pu-Be source in, moved away from the detector by 26".
  
Run 4557 had a discriminator threshold set to 10 mV and the terminators were in. DAQ rate is 1400 Hz. Pu-B source in, moved away from the detector by 52".
+
{| cellpadding="2" style="border: 1px solid darkgray;"
 +
|- border="0"
 +
| [[File:r4552_fake_stop.png|340px|Region 1]]
 +
| [[File:r4552_det4_up.png|340px|Region 2]]
 +
|- align="center"
 +
| Fake stop|| Second hit
 +
|}
  
PMT detector #4 HV at -1400 Volts
+
* Run 4553 had a discriminator threshold set to 20 mV and the terminators were in.  DAQ rate is 1850 Hz. Pu-Be source in, moved away from the detector by 26".
  
Run 4558 had a discriminator threshold set to 10 mV and the terminators were in. DAQ rate is 4670 Hz. Pu-Be source in, moved away from the detector by 52".
+
{| cellpadding="2" style="border: 1px solid darkgray;"
 +
|- border="0"
 +
| [[File:r4553_fake_stop.png|340px|Region 1]]
 +
| [[File:r4553_det4_up.png|340px|Region 2]]
 +
|- align="center"
 +
| Fake stop|| Second hit
 +
|}
  
Run 4559 had a discriminator threshold set to 10 mV and the terminators were in.  DAQ rate is  Hz. Pu-Be source in, moved away from the detector by 91".
+
* Run 4554 had a discriminator threshold set to 20 mV and the terminators were in.  DAQ rate is 625 Hz. Pu-Be source in, moved away from the detector by 52".
 +
 
 +
{| cellpadding="2" style="border: 1px solid darkgray;"
 +
|- border="0"
 +
| [[File:r4554_fake_stop.png|340px|Region 1]]
 +
| [[File:r4554_det4_up.png|340px|Region 2]]
 +
|- align="center"
 +
| Fake stop|| Second hit
 +
|}
 +
 
 +
* Run 4555 had a discriminator threshold set to 20 mV and the terminators were out.  DAQ rate is 650 Hz. Pu-Be source in, moved away from the detector by 52".
 +
 
 +
{| cellpadding="2" style="border: 1px solid darkgray;"
 +
|- border="0"
 +
| [[File:r4555_fake_stop.png|340px|Region 1]]
 +
| [[File:r4555_det4_up.png|340px|Region 2]]
 +
|- align="center"
 +
| Fake stop|| Second hit
 +
|}
 +
 
 +
* Run 4556 had a discriminator threshold set to 10 mV and the terminators were out.  DAQ rate is 1400 Hz. Pu-Be source in, moved away from the detector by 52".
 +
 
 +
{| cellpadding="2" style="border: 1px solid darkgray;"
 +
|- border="0"
 +
| [[File:r4556_fake_stop.png|340px|Region 1]]
 +
| [[File:r4556_det4_up.png|340px|Region 2]]
 +
|- align="center"
 +
| Fake stop|| Second hit
 +
|}
 +
 
 +
* Run 4557 had a discriminator threshold set to 10 mV and the terminators were in.  DAQ rate is 1400 Hz. Pu-B source in, moved away from the detector by 52".
 +
 
 +
{| cellpadding="2" style="border: 1px solid darkgray;"
 +
|- border="0"
 +
| [[File:r4557_fake_stop.png|340px|Region 1]]
 +
| [[File:r4557_det4_up.png|340px|Region 2]]
 +
|- align="center"
 +
| Fake stop|| Second hit
 +
|}
 +
 
 +
 
 +
PMT detector #4 HV at -1400 Volts
 +
 
 +
* Run 4558 had a discriminator threshold set to 10 mV and the terminators were in.  DAQ rate is 4670 Hz. Pu-Be source in, moved away from the detector by 52".
 +
 
 +
{| cellpadding="2" style="border: 1px solid darkgray;"
 +
|- border="0"
 +
| [[File:r4558_fake_stop.png|340px|Region 1]]
 +
| [[File:r4558_det4_up.png|340px|Region 2]]
 +
|- align="center"
 +
| Fake stop|| Second hit
 +
|}
 +
 
 +
* Run 4559 had a discriminator threshold set to 10 mV and the terminators were in.  DAQ rate is 2700 Hz. Pu-Be source in, moved away from the detector by 91".
 +
 
 +
{| cellpadding="2" style="border: 1px solid darkgray;"
 +
|- border="0"
 +
| [[File:r4559_fake_stop.png|340px|Region 1]]
 +
| [[File:r4559_det4_up.png|340px|Region 2]]
 +
|- align="center"
 +
| Fake stop|| Second hit
 +
|}
 +
 
 +
* Run 4560 had a discriminator threshold set to 10 mV and the terminators were out. DAQ rate is 2700 Hz. Pu-Be source in, moved away from the detector by 91".
 +
 
 +
{| cellpadding="2" style="border: 1px solid darkgray;"
 +
|- border="0"
 +
| [[File:r4560_fake_stop.png|340px|Region 1]]
 +
| [[File:r4560_det4_up.png|340px|Region 2]]
 +
|- align="center"
 +
| Fake stop|| Second hit
 +
|}
 +
 
 +
* Run 4561 had a discriminator threshold set to 10 mV and the terminators were out.  DAQ rate is 200 Hz. Cosmic rays test.
 +
 
 +
{| cellpadding="2" style="border: 1px solid darkgray;"
 +
|- border="0"
 +
| [[File:r4561_fake_stop.png|340px|Region 1]]
 +
| [[File:r4561_det4_up.png|340px|Region 2]]
 +
|- align="center"
 +
| Fake stop|| Second hit
 +
|}
 +
 
 +
;1.8 g Pu-Be source should be used next time

Latest revision as of 23:53, 25 April 2015

Pu-Be source test

PMT detector #4 HV at -1000 Volts

Pu-Be 18 gram source# M1197

  • Run 4546 had a discriminator threshold set to 80 mV and the terminators were out. Cosmic rays test.
Region 1 Region 2
Fake stop Second hit
  • Run 4547 had a discriminator threshold set to 80 mV and the terminators were out. DAQ rate is 100 Hz. Pu-Be source in.
Region 1 Region 2
Fake stop Second hit
  • Run 4548 had a discriminator threshold set to 80 mV and the terminators were in. DAQ rate is 100 Hz. Pu-Be source in.
Region 1 Region 2
Fake stop Second hit
  • Run 4549 had a discriminator threshold set to 40 mV and the terminators were in. DAQ rate is 2800 Hz. Pu-Be source in.
Region 1 Region 2
Fake stop Second hit
  • Run 4550 had a discriminator threshold set to 40 mV and the terminators were out. DAQ rate is 2800 Hz. Pu-Be source in.
Region 1 Region 2
Fake stop Second hit
  • Run 4551 had a discriminator threshold set to 40 mV and the terminators were out. DAQ rate is 500 Hz. Pu-Be source in, moved away from the detector by 26".
Region 1 Region 2
Fake stop Second hit
  • Run 4552 had a discriminator threshold set to 40 mV and the terminators were in. DAQ rate is 500 Hz. Pu-Be source in, moved away from the detector by 26".
Region 1 Region 2
Fake stop Second hit
  • Run 4553 had a discriminator threshold set to 20 mV and the terminators were in. DAQ rate is 1850 Hz. Pu-Be source in, moved away from the detector by 26".
Region 1 Region 2
Fake stop Second hit
  • Run 4554 had a discriminator threshold set to 20 mV and the terminators were in. DAQ rate is 625 Hz. Pu-Be source in, moved away from the detector by 52".
Region 1 Region 2
Fake stop Second hit
  • Run 4555 had a discriminator threshold set to 20 mV and the terminators were out. DAQ rate is 650 Hz. Pu-Be source in, moved away from the detector by 52".
Region 1 Region 2
Fake stop Second hit
  • Run 4556 had a discriminator threshold set to 10 mV and the terminators were out. DAQ rate is 1400 Hz. Pu-Be source in, moved away from the detector by 52".
Region 1 Region 2
Fake stop Second hit
  • Run 4557 had a discriminator threshold set to 10 mV and the terminators were in. DAQ rate is 1400 Hz. Pu-B source in, moved away from the detector by 52".
Region 1 Region 2
Fake stop Second hit


PMT detector #4 HV at -1400 Volts
  • Run 4558 had a discriminator threshold set to 10 mV and the terminators were in. DAQ rate is 4670 Hz. Pu-Be source in, moved away from the detector by 52".
Region 1 Region 2
Fake stop Second hit
  • Run 4559 had a discriminator threshold set to 10 mV and the terminators were in. DAQ rate is 2700 Hz. Pu-Be source in, moved away from the detector by 91".
Region 1 Region 2
Fake stop Second hit
  • Run 4560 had a discriminator threshold set to 10 mV and the terminators were out. DAQ rate is 2700 Hz. Pu-Be source in, moved away from the detector by 91".
Region 1 Region 2
Fake stop Second hit
  • Run 4561 had a discriminator threshold set to 10 mV and the terminators were out. DAQ rate is 200 Hz. Cosmic rays test.
Region 1 Region 2
Fake stop Second hit
1.8 g Pu-Be source should be used next time