Started testing detector 4 again.

Fake stop time

Discriminator width was decreased from 40 ns (255 on discriminator display) to 6.3 ns (10 on discriminator display)

Discriminator threshold is set to 100 mv

The delay for the fake stop is set to 800 ns


FakeStop 4-20-2015 A.png


CODA/CODAreader/ROOT_V5.30/R1DC/evio2nt -fr4516.dat > /dev/null


R1DC->Draw("(evt.TDC1190[1][6]-evt.TDC1190[2][6])/10>>(4096,-4096,4096)");

Run 4536


Below is a picture of the typical 800 ns fake stop pulse that should appear as a second hit in the Multihit TDC.


R4536FakeStop C.png



Now decrease delay of fake stop to 400 ns


FakeStop 4-20-2015 B.png

R4537FakeStop C.png

Terminator test

Change discriminator threshold with terminators in

PMT detector #4 HV at -1000 Volts


Run 4537 had a discriminator threshold set to 100 mv and the terminators were in.

Now change threshold to 30 mv

Run 4538

R4538 Scope.png

R4538 PMT4 TDCmultihits.png


2nCor_44#Run_Plan