04-20-2015

From New IAC Wiki
Jump to navigation Jump to search

Started testing detector 4 again.

Fake stop time

Discriminator width was decreased from 40 ns (255 on discriminator display) to 6.3 ns (10 on discriminator display)

Discriminator threshold is set to 100 mv

The delay for the fake stop is set to 800 ns


FakeStop 4-20-2015 A.png

Below is a picture of the typical 400 ns fake stop pulse that should appear as a second hit in the Multihit TDC.


FakeStop 4-20-2015 B.png


CODA/CODAreader/ROOT_V5.30/R1DC/evio2nt -fr4516.dat > /dev/null


R1DC->Draw("(evt.TDC1190[1][6]-evt.TDC1190[2][6])/10>>(4096,-4096,4096)");

Run 4534

R4534FakeStop C.png

Now increase delay of fake stop to 800 ns

200 px

Terminator test

r4534

Terminators in

Discriminator at 20 mv

PMT detector #4 HV at -1000 Volts


2nCor_44#Run_Plan