R1S5 SL1 HVB1 BurnIn
Jump to navigation
Jump to search
Date | Time (24h) | Guard (V) | Sense (V) | Field (V) | Sense ( | A)Field ( | A)comments |
06/25/13 | 15:05 | 30 | 40 | -50 | 0.0 | 0.025 | Guard current is at 15.9 microamps. Have never observed it above 0.000 before. Turned off voltage to check cabling setup. |
06/25/13 | 15:19 | 30 | 40 | -50 | 0.0 | 0.025 | Found a grounding strap attached to the HVB side of Board 1. Removed and torqued down the screw it was on. Guard current is normal now. Field/sense current stabilized within 30 seconds to this reading. |
06/25/13 | 15:22 | 536 | 1500 | -750 | 4.6 | 4.7 | Current appears stable +/- 0.2 microamps for at least 30 seconds after finishing voltage rise. |
06/25/13 | 15:29 | 536 | 1500 | -750 | 4.1 | 4.2 | |
06/26/13 | 07:14 | 536 | 1500 | -750 | 1.18 | 1.2 | |
06/26/13 | 17:29 | 536 | 1500 | -750 | 1.2 | 1.1 | Shut off HV to switch back to the Kraken on SL2 and check current levels w/ grounding strap now removed. |
07/19/13 | 22:40 | 536 | - | returned to burning in SL1 using the kraken after finishing SL2 | |||
07/19/13 | 14:51 | 536 | 680 | -680 | 4.0 | 4.2 | |
07/22/13 | 08:20 | 536 | 770 | -750 | 8.3 | 9.5 | guard current 1 uA |
07/22/13 | 10:47 | 536 | 990 | -750 | 7.7 | 8.7 | guard current 0.89 uA |
07/22/13 | 11:08 | 536 | 1040 | -750 | 6.6 | 6.7 | |
07/24/13 | 07:52 | 536 | 1030 | -750 | 3.3 | 3.4 | after replacing a tank.
|